K080093 · Gyrus Acmi, Inc. · GEI · May 2, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080093
Device Name
PKS PLASMA MORCELLATOR, MODELS 962000PK, 3620PK
Applicant
Gyrus Acmi, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
May 2, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Gyrus ACMI PKS Plasma Morcellator is intended for cutting, coring, and extracting tissue in various operative laparoscopy procedures, including laparoscopic general surgical procedures, laparoscopic urological procedures, and laparoscopic gynecologic procedures.
Device Story
The Gyrus ACMI PKS Plasma Morcellator is a laparoscopic surgical instrument used for tissue resection. It consists of a pistol-grip handle, a hollow cutting shaft, and an internal grasper. The device connects to a Gyrus Medical electrosurgical generator to deliver bipolar energy for tissue cutting. During surgery, the surgeon inserts the device into the abdomen; the grasper is passed through the hollow shaft to secure tissue, which is then retracted toward the cutting tip. The device facilitates the extraction of tissue during laparoscopic general, urological, and gynecological procedures. It is operated by a surgeon in an operating room setting. The device combines the mechanical morcellation function of traditional morcellators with the bipolar electrosurgical cutting technology used in Gyrus PlasmaCision instruments, allowing for efficient tissue removal.
Clinical Evidence
Bench testing only. Performance was compared against the predicate Gynecare Laparoscopic Morcellator and the bipolar cutting technology of the Gyrus PlasmaCision Laparoscopic Spatula. Testing demonstrated that the device met performance requirements and exhibited comparable characteristics to the predicate devices.
Technological Characteristics
Pistol-grip electrosurgical morcellator with hollow cutting shaft and integrated grasper. Utilizes bipolar electrosurgical energy for tissue cutting. Connects to Gyrus Medical electrosurgical generators. Patient-contacting materials are consistent with previously cleared Gyrus ACMI devices. Dimensions are similar to the Ethicon Gynecare Laparoscopic Morcellator.
Indications for Use
Indicated for patients undergoing operative laparoscopic procedures requiring tissue cutting, coring, and extraction, including general surgical, urological, and gynecological procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Gyrus Medical Inc. PlasmaCision Laparoscopic Spatula (K041633)
Ethicon Inc. Gynecare Laparoscopic Morcellator (K993801)
Gyrus General Purpose Electrosurgical System (K050550)
Submission Summary (Full Text)
{0}------------------------------------------------
MAY - 2 2008
Gyrus ACMI PKS Plasma Morcellator Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
Traditional 510(k) Notification 510(k) Summary January 11, 2008
K080093
510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc. Gyrus ACMI PKS Plasma Morcellator
Page 1 of 3
#### General Information
Manufacturer:
Gyrus Medical Ltd. Fortran Road, St. Mellons Cardiff UNITED KINGDOM CF3 0LT
Establishment Registration Number:
510(k) Submitter:
9617070
3003790304
Terrence E. Sullivan
January 11, 2008
Gyrus ACMI, Inc. 136 Turnpike Rd. Southborough, MA 01772-2104
Vice President, Regulatory Affairs
Establishment Registration Number:
Contact Person:
Date Prepared:
#### Device Description
Classification Name:
Laparoscope, Gynecologic and accessories (21 CFR 884.1720), Class II, Obstetrics/Gynecology Panel; Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400), Class II General and Plastic Surgery Panel
Gyrus Medical Solid Organ Resection Device (SORD)
Gyrus ACMI PKS Plasma Morcellator
Generic/Common Name:
Electrosurgical Instruments
#### Predicate Devices
Project Name:
Trade Name:
Gyrus Medical Inc. PlasmaCision Laparoscopic Spatula K041633 Ethicon Inc. Gynecare Laparoscopic Morcellator K993801
{1}------------------------------------------------
Traditional 510{k) Notification 510(k) Summary January 11, 2008
## Intended Uses
The Gyrus ACMI PKS Plasma Morcellator is intended for cutting, coring, and extracting tissue in various operative laparoscopy procedures, including laparoscopic general surgical procedures, laparoscopic urological procedures, and laparoscopic gynecologic procedures.
## Product Description
The Gyrus ACMI PKS Plasma Morcellator consists of two main elements: a Morcellator that includes a cutting shaft and handle with an integrated electrical cable that connects to a Gyrus Medical generator; and a Grasper that passes through the hollow shaft of the Morcellator, to grasp tissue and retract it toward the cutting element on the Morcellator shaft tip. An obturator may also be used to assist in the introduction of the device in laparoscopic procedures, and like the grasper, is positioned through the shaft of the Morcellator tube when being utilized
## Technological Characteristics and Substantial Equivalence
The Gyrus ACMI PKS Plasma Morcellator utilizes features incorporated into the following legally marketed predicate devices:
The Gyrus ACMI PKS Plasma Morcellator utilizes the same bipolar electrosurgical energy to cut tissue as that used in the predicate Gyrus Medical Inc. PlasmaCision Laparoscopic Spatula.
The Gyrus ACMI PKS Plasma Morcellator connects to the same electrosurgical generators, the currently marketed Gyrus PlasmaKinetic SuperPulse System and Gyrus General Purpose Electrosurgical System generators, as the predicate Gyrus Medical Inc. PlasmaCision Laparoscopic Spatula. These generators were cleared. under K031085 and K050550, respectively.
The Gyrus ACMI PKS Plasma Morcellator is dimensionally similar to the predicate Ethicon Gynecare Laparoscopic Morcellator, having similar Morcellator shaft diameters.
Like the predicate Ethicon Gynecare Laparoscopic Morcellator, the Gyrus ACMI PKS Plasma Morcellator utilizes a pistol grip Morcellator with a hollow shaft through which a grasper is passed to grasp tissue to be resected.
{2}------------------------------------------------
Gyrus ACMI PKS Plasma Morcellator Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
Traditional 510(k) Notification 510(k) Summarv January 11, 2008
Page 3 of 3
Like the predicate Ethicon Gynecare Laparoscopic Morcellator, the Gyrus ACMI PKS Plasma Morcellator is indicated for cutting, coring, and extracting tissue in various operative laparoscopy procedures, including laparoscopic general surgical procedures, laparoscopic urological procedures, and laparoscopic gynecologic procedures.
KOQ0093
The Gyrus ACMI PKS Plasma Morcellator uses patient-contacting materials that are utilized in the predicate devices, as well as other legally marketed devices manufactured by Gyrus ACMI.
During design verification, the performance of the PKS Plasma Morcellator was compared against the known performance characteristics of the predicate Gynecare Laparoscopic Morcellator using the bipolar tissue cutting technology of the predicate Gyrus PlasmaCision Laparoscopic Spatula. Testing demonstrated that the performance requirements were met, and that the PKS Plasma Morcellator exhibited comparable performance characteristics to both the predicate care Laparoscopic Morcellator and predicate Gyrus PlasmaCision Laparoscopic Spatula.
In summary, the Gyrus ACMI PKS Plasma Morcellator is substantially equivalent to the predicate devices and presents no new questions of safety or efficacy.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 2 2008
Gyrus ACMI, Inc. % Mr. Terrence E. Sullivan VP, Regulatory Affairs 136 Turnpike Road Southborough, Massachusetts 01772
Re: K080093
Trade/Device Name: Gyrus ACMI PKS Plasma Morcellator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI, HET Dated: April 17, 2008 Received: April 18, 2008
Dear Mr. Sullivan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Terrence E. Sullivan
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Gyrus ACMI PKS Plasma Morcellator Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
Traditional 510(k) Notification Statement of Intended Use January 11, 2008
Device Name: Gyrus ACMI PKS Plasma Morcellator
510(k) Number: K080093
## Statement of Intended use:
The Gyrus ACMI PKS Plasma Morcellator is intended for cutting, coring, and extracting tissue in various operative laparoscopy procedures, including laparoscopic general surgical procedures, laparoscopic urological procedures, and laparoscopic gynecologic procedures.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: X
OR Over-the-Counter Use:
(Per 21 CFR 801.109)
Mark A. Mellman
Ayiston 2 . Restorative, ivision of Gener ad Neurological Devices
100 Number
K080093
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.