BIOMET SPORTS MEDICINE ANCHOR DEVICES AND ZIPLOOP CONSTRUCTS
K080088 · Biomet Manufacturing, Inc. · HWC · Jun 11, 2008 · Orthopedic
Device Facts
Record ID
K080088
Device Name
BIOMET SPORTS MEDICINE ANCHOR DEVICES AND ZIPLOOP CONSTRUCTS
Applicant
Biomet Manufacturing, Inc.
Product Code
HWC · Orthopedic
Decision Date
Jun 11, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Biomet Sports Medicine™ Anchors and ZipLoop™ Constructs are intended for use in soft tissue reattachment procedures in the shoulder, foot / ankle, elbow, knee, hand / wrist, and hip.
Device Story
Biomet Sports Medicine™ suture anchors and ZipLoop™ constructs are internal fixation devices for arthroscopic and orthopedic soft tissue reattachment. Devices are either preloaded with suture or a ZipLoop™ construct. ZipLoop™ constructs provide a flexible, adjustable means of soft tissue fixation, eliminating the need for manual knot-tying. Used by surgeons in clinical settings (OR) for procedures involving shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. The device secures soft tissue to bone, facilitating healing and reconstruction. Benefit includes simplified, adjustable fixation during complex orthopedic surgeries.
Clinical Evidence
No clinical data provided as a basis for substantial equivalence. Substantial equivalence is supported by non-clinical laboratory testing.
Technological Characteristics
Internal fixation devices including suture anchors and ZipLoop™ constructs. Materials include titanium, PEEK, and resorbable polymers. ZipLoop™ constructs utilize UHMWPE strands. Devices are designed for arthroscopic/orthopedic soft tissue fixation. No software or electronic components.
Indications for Use
Indicated for soft tissue reattachment in shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. Specific procedures include rotator cuff repair, Bankart lesion repair, biceps tenodesis, Achilles tendon repair, ligament reconstruction, and patellar realignment. Contraindications not specified.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K080088
JUN 1 1 2008
Image /page/0/Picture/2 description: The image shows the logo for BIOMET MANUFACTURING CORP. The word "BIOMET" is in a bold, sans-serif font, with each letter connected to the next. Below the logo, the words "MANUFACTURING CORP." are printed in a smaller, sans-serif font.
# 510(k) Summary
January 11, 2008 Preparation Date:
Biomet Sports Medicine, Inc.(formerly known as Arthrotek, Inc.) Applicant/Sponsor:
Contact Person: Robert R. Friddle
Biomet Sports Medicine™ Anchor devices and ZipLoop™ Constructs Proprietary Name:
Soft tissue anchor Common Name:
Fastener, fixation, nondegradable, soft tissue (21 CFR 888.3040); Screw, Classification Name: fixation, bone (21 CFR 888.3040); Suture, nonabsorbable, synthetic, polyester (21CFR878.5000); Suture, nonabsorbable, synthetic, polyethylene (21CFR878.5000); Clip, implantable (21 CFR 878.4300); Marker, radiographic, implantable (21 CFR 878.4300); Staple, fixation, bone (21 CFR 888.3030); Fastener, fixation, biodegradable, soft tissue (21 CFR888.3030),
## Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
Multitak Suture System™ - K973015 (Bonutti Research, Inc.), Cruciate Ligament Button - K813581 (Biomet, Inc.), BioRaptor 2.9mm Suture Anchors - K053344 (Smith & Nephew, Inc.), Ti-Screw - K012503 (Biomet, Inc.), Hitch™ L-15 - K061657 (Biomet, Inc.), ALLthread™ Tltanium - K042460 (Biomet, Inc.), ALLthread™ PEEK K060693 and K071569(Biomet, Inc.), Hitch™ PEEK - K060693(Biomet, Inc.), ALLthread™ Lactosorb® - K061389 (Biomet, Inc.) LactoScrew® L-15 - K012872, K033355 and K061801 (Biomet, Inc.) MicroMax™ - K040475 (Biomet, Inc.), Hitch™ L-15 - K061657(Biomet, Inc.),
Device Description: The Blomet Sports Medicine™ suture anchor devices (Ti-Screw, ALLthread™ Titanium, ALLthread™ PEEK, Hitch™ PEEK, ALLthread™ L-15, LactoScrew® L-15, MicroMax™ and Hitch™ L-15) Suture Anchors are internal fixation devices intended to aid in arthroscopic and orthopedic reconstructive procedures requiring soft tissue fixation due to injury or degenerative disease. The devices are available preloaded with suture and / or a ZipLoop™ Construct.
The Biomet Sports Medicine™ ZipLoop™ Constructs provide a flexible, adjustable means of soft tissue The blomer in a manner that does not require the surgeon to the a knot in sutures. The ZipLoop™ Constructs are available in three configurations and two strand materials.
Intended Use: Biomet Sports Medicine™ Anchors and ZipLoop™ Constructs are intended for use In soft Intended DSC. District oper of it the shoulder, foot / ankle, elbow, knee, hand / wrist, and hip. The Biomet
Mailing Address: P.O. Box 587 Warsaw IN 46581 0587 Tel: Fren' 200 348 9500 Official 474 257 6939 Main Fax: 574 287 8137 25:47 Granet.com
Shipping Address: 56 East Bell Dove Warsen, IN 46582
5 - Page 1 of 5
{1}------------------------------------------------
suture anchor devices have similar indications for use with one Indication for Use statement for nonresorbable anchor devices and another for resorbable anchor devices. ZipLoop™ Constructs may be (esoraged separately (one Indication for Use Statement covers a ZipLoop™ Construct packaged separately) or preloaded on suture anchors In place of sutures (an Indication for Use statement for non-resorbable anchor device with ZipLoop™ Construct and another for a resorbable anchor device with a ZipLoop™ Construct). Specific indications for use are as follows:
### Anchor devices without ZipLoop Constructs:
Device Name: Ti-Screw, ALLthread™ Titanium, ALLthread™ PEEK, Hitch™ PEEK Anchors
Ti-Screw, ALLthread™ Titanium, ALLthread™ PEEK and Hitch™ PEEK Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. Specific Indications are as follows:
Shoulder: Acromio-Clavicular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Biceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs
Foot and Ankle: Achilles Tendon Repair/Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot/Forefoot Reconstruction/Repairs
Elbow: Blceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral/Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
Knee: ACL/PCL (only for ALLthread™ Titanlum and ALLthread™ PEEK anchors), Illiotibial Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Reallgnment / Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb), Scapholunate Ligament Reconstruction, Tendon Transfers In Phalanx, Ulnar/Radial Collateral Ligament Reconstruction Acconstraction) Tenash Titanium and ALLthread™ PEEK), Volar Plate Reconstruction
Hip: Labral
Device Name: ALLthread™ L-15, LactoScrew® L-15, MicroMax™ and Hitch™ L-15 Suture Anchors
ALLthread™ L-15, LactoScrew® L-15, MicroMax™ and Hltch™ L-15 Suture Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot and ankle, elbow, knee, hand and wrist, and hip. Specific indications as follows:
Shoulder: Acromio-Clavicular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Biceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs
Foot and Ankle: Achilles Tendon Repair / Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot / Forefoot Reconstruction / Repairs
Elbow: Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral / Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
{2}------------------------------------------------
Knee: ACL/PCL (only for ALLthread™ L-15), Illiotibial Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment / Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb) (not Including Hitch™ L-15), Scapholunate Ligament Reconstruction, Tendon Transfers In Phalanx (not Including Hitch™ L-15), Ulnar/Radial Collateral Ligament Reconstruction (only for ALLthread™ L-15 and LactoScrew® L-15}, Volar Plate Reconstruction (not including Hitch™ L-15)
Hip: Labral
#### ZipLoop™ Constructs packaged separately:
Device Name: Full, Bowtie and Half Zip-Loop™ Constructs
Full, Bowtle and Half ZipLoop™ Constructs are indicated for use in soft tissue reattachment procedures in the shoulder, wrist/hand, ankle/foot, elbow, knee, and hip.
When a ZipLoop™ Construct is packaged separately, the specific Indications for use are as follows:
Shoulder: Acromio-Clavicular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Biceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs
Foot and Ankle: Achilles Tendon Repair / Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs / Reconstruction, Midfoot/Forefoot Reconstruction / Repairs
Elbow: Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral / Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
Knee: ACL/PCL (only for 2-0 or larger UHMWPE Full or Bowtie ZipLoop™ Construct), Illiotibial Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment / Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb), Scapholunate Ligament Reconstruction, Tendon Transfers in Phalanx, Ulnar/Radial Collateral Ligament Reconstruction (only for 2-0 or larger UHMWPE Full and Bowtie ZipLoop™ Construct), Volar Plate Reconstruction
Hip: Labral
When a ZipLoop™ Construct Is packaged with a Blomet Sports Medicine™ internal fixation device, please refer to the package insert included with the surgeon's choice of internal fixation device for specific indications.
### Anchor devices packaged with ZlpLoop™ Constructs:
Device Name: Ti-Screw, ALLthread™ Titanlum, ALLthread™ PEEK and Hitch™ PEEK Anchors with ZipLoop Constructs
Ti-Screw, ALLthread™ Titanlum, ALLthread™ PEEK and Hitch™ PEEK Anchors with ZipLoop Constructs are Indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. Specific indications are as follows:
{3}------------------------------------------------
Shoulder: Acromio-Clavicular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Biceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs
Foot and Ankle: Achilles Tendon Repair/Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot/Forefoot Reconstruction/Repairs
Elbow: Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral/Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
Knee: ACL/PCL (only for ALLthread™ Titanium and ALLthread™ PEEK anchors in conjunction with 2-0 or larger UHMWPE Full or Bowtie ZipLoop™ Construct), Illiotibial Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment / Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb), Scapholunate Ligament Reconstruction, Tendon Transfers in Phalanx, Ulnar/Radial Collateral Ligament Reconstruction (only for Tl-Screw, ALLthread™ Titanium and ALLthread™ PEEK In conjunction with 2-0 or larger UHMWPE Full or Bowtle ZipLoop™ Construct); Volar Plate Reconstruction
Hip: Labral
ALLthread™ L-15, LactoScrew® L-15, MicroMax™ and Hitch™ L-15 Suture Anchors with Device Name: ZipLoop Constructs
ALLthread™ L-15, LactoScrew® L-15, MicroMax™ and Hitch™ L-15 Suture Anchors with ZipLoop Constructs are indicated for soft tissue reattachment procedures in the shoulder, foot and ankle, elbow, knee, hand and wrist, and hip. Specific indications as follows:
Shoulder: Acromio-Clavicular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Biceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Leslon Repairs
Foot and Ankle: Achilles Tendon Repair / Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot / Forefoot Reconstruction / Repairs
Elbow: Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral / Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
Knee: ACL/PCL (only for ALLthread™ L-15 in conjunction with 2-0 or larger UHMWPE Full or Bowtle ZipLoop™ Construct), Illiotibial Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment / Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb) (not Including Hitch™ L-15), Scapholunate Ligament Reconstruction, Tendon Transfers In Phalanx (not including Hitch™ L-15), Unar/Radial Collateral Ligament Reconstruction (only for ALLthread™ L-15 and LactoScrew® L-15 in conjunction with 2-0 or larger UHMWPE Full or Bowtie ZipLoop™ Construct), Volar Plate Reconstruction (not Including Hitch™ L-15)
Hip: Labrai
{4}------------------------------------------------
Summary of Technologies: The technological characteristics (material, design and sizing) of the current suture anchor devices are not changed and are similar or identical to the predicate devices. The technological characteristics of the new ZipLoop™ Constructs (material, design, sizing and Indications) are similar or identical to the suture constructs provided pre-loaded with predicate devices.
Non-Clinical Testing: Non-clinical laboratory testing was performed to determine substantial equivalence. The results indicated that the devices were functional within their intended use.
Clinical Testing: None provided as a basis for substantial equivalence.
All trademarks are property of Blomet, Inc. unless otherwise specificed. Multitak Suture System™ is a trademark of Bonutti Research Corporation. PEEK-OPTIMA® is a registered trademark of Inviblo, Inc. BloRaptor™ Is a trademark of Smith&Nephew.
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with healthcare, to the right. To the left of the caduceus, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Biomet Manufacturing Corp. % Mr. Robert R. Friddle Regulatory Specialist P.O. Box 587 Warsaw, Indiana 46581-0587
JUN 1 1 2008
K080088 Trade/Device Name: Expanded indications for Biomet Sports Medicine™ Anchor Devices and ZipLoop™ Constructs Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC, MBI, JDR, MAI Dated: May 16, 2008 Received: May 19, 2008
Dear Mr. Friddle:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. Robert R. Friddle
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Mellema
Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Ko80088 510(k) Number (if known):
Device Name: Ti-Screw, ALLthread™ Titanium, ALLthread™ PEEK, Hitch™ PEEK Anchors
Indications for Use:
Ti-Screw, ALLthread™ Titanium, ALLthread™ PEEK and Hitch™ PEEK Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. Specific indications are as follows:
Shoulder: Acromio-Clavicular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Biceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs
Foot and Ankle: Achilles Tendon Repair/Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot/Forefoot Reconstruction/Repairs
Elbow: Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral/Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
Knee: ACL/PCL (only for ALLthread™ Titanium and ALLthread™ PEEK anchors), Illiotibial Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment / Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb), Scapholunate Ligament Reconstruction, Tendon Transfers in Phalanx, Ulnar/Radial Collateral Ligament Reconstruction (only for TI-Screw, ALLthread™ Titanium and ALLthread™ PEEK), Volar Plate Reconstruction
Hip: Labral
YES Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
Neike Ohe frank
and Neurological Devices
**510(k) Number** K080088
{8}------------------------------------------------
510(k) Number (if known): K080088
Device Name: ALLthread™ L-15, LactoScrew® L-15, MicroMax™ and Hitch™ L-15 Suture Anchors
Indications for Use:
ALLthread™ L-15, LactoScrew® L-15, MicroMax™ and Hitch™ L-15 Suture Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot and ankle, elbow, knee, hand and wrist, and hip. Specific indications as follows:
Shoulder: Acromio-Clavicular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Biceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs
Foot and Ankle: Achilles Tendon Repair / Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot / Forefoot Reconstruction / Repairs
Elbow: Blceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral / Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
Knee: ACL/PCL (only for ALLthread™ L-15), Illiotibial Band Tenodesis, John Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment / Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb) (not Including Hitch™ L-15), Scapholunate Ligament Reconstruction, Tendon Transfers In Phalanx (not including Hitch™ L-15), Ulnar/Radial Collateral Ligament Reconstruction (only for ALLthread™ L-15 and LactoScrew® L-15}, Volar Plate Reconstruction (not Including Hitch™ L-15)
Hip: Labral
Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page Dof 1
Nail R.E. Ogil for mkm
Division of General, Restorative, and Neurological Devices
**510(k) Number** K080087
{9}------------------------------------------------
510(k) Number (if known): k080044
Device Name: Full, Bowtie and Half Zip-Loop™ Constructs
Indications for Use:
Full, Bowtie and Half ZipLoop™ Constructs are indicated for use in soft tissue reattachment procedures in the shoulder, wrist/hand, ankle/foot, elbow, knee, and hip.
When a ZipLoop™ Construct is packaged separately, the specific indications for use are as follows:
Shoulder: Acromio-Clavicular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Biceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Leslon Repairs
Foot and Ankle: Achilles Tendon Repair / Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs / Reconstruction, Midfoot/Forefoot Reconstruction / Repairs
Elbow: Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral / Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
Knee: ACL/PCL (only for 2-0 or larger UHMWPE Full or Bowtle ZipLoop™ Construct), Illiotiblal Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment / Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb), Scapholunate Ligament Reconstruction, Tendon Transfers In Phalanx, Ulnar/Radial Collateral Ligament Reconstruction (only for 2-0 or larger UHMWPE Full and Bowtie ZipLoop™ Construct), Volar Plate Reconstruction
Hip: Labral
When a ZlpLoop™ Construct is packaged with a Blomet Sports Medicine™ Internal fixation device, please refer to the package insert included with the surgeon's choice of internal fixation device for specific Indications.
YES Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | K080088 |
|---------------|---------|
|---------------|---------|
Page 3 of 5
{10}------------------------------------------------
510(k) Number (if known): K010088
Device Name: TI-Screw, ALLthread™ Titanium, ALLthread™ PEEK and Hitch™ PEEK Anchors with ZipLoop Constructs
Indications for Use:
Ti-Screw, ALLthread™ Titanium, ALLthread™ PEEK and Hitch™ PEEK Anchors with ZipLoop Constructs are indicated for soft tissue reattachment procedures in the shoulder, footfankle, elbow, knee, hand/wrist, and hip. Specific indications are as follows:
Shoulder: Acromio-Clavlcular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Blceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs
Foot and Ankle: Achilles Tendon Repair/Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot/Forefoot Reconstruction/Repairs
Elbow: Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral/Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
Knee: ACL/PCL (only for ALLthread™ Titanium and ALLthread™ PEEK anchors in conjunction with 2-0 or larger UHMWPE Full or Bowtie ZipLoop™ Construct), Illiotiblai Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment / Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb), Scapholunate Ligament Reconstruction, Tendon Transfers in Phalanx, Ulnar/Radial Collateral Ligament Reconstruction (only for Tl-Screw, ALLthread™ Titanium and ALLthread™ PEEK In conjunction with 2-0 or larger UHMWPE Full or Bowtie ZipLoop™ Construct); Volar Plate Reconstruction
Hip: Labral
YES Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nilfre Igl fr mdn
Page of
Division of General, Restorative, and Neurological Devices
4-5
**510(k) Number** K080088
{11}------------------------------------------------
510(k) Number (if known): Koofoff
ALLthread™ L-15, LactoScrew® L-15, MicroMax™ and Hitch™ L-15 Suture Anchors with Device Name: ZipLoop Constructs
Indications for Use:
ALLthread™ L-15, LactoScrew® L-15, MicroMax™ and Hitch™ L-15 Suture Anchors with ZipLoop Constructs are indicated for soft tissue reattachment procedures in the shoulder, foot and ankle, elbow, knee, hand and wrist, and hip. Specific indications as follows:
Shoulder: Acromio-Clavicular Separation, Anterior Shoulder Instability Repair, Bankart Lesion Repairs, Biceps Tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs
Foot and Ankle: Achilles Tendon Repair / Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot / Forefoot Reconstruction / Repairs
Elbow: Bloeps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral / Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction
Knee: ACL/PCL (only for ALLthread™ L-15 in conjunction with 2-0 or larger UHMWPE Full or Bowtle ZipLoop™ Construct), Illiotiblal Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment/Repair, Posterior Oblique Ligament Repair, VMO Advancement
Hand and Wrist: Collateral Ligament Repair (Gamekeeper's Thumb) (not including Hitch™ L-15), Scapholunate Ligament Reconstruction, Tendon Transfers In Phalanx (not Including Hitch™ L-15), Uinar/Radial Collateral Ligament Reconstruction (only for ALLthread™ L-15 and LactoScrew® L-15 In conjunction with 2-0 or larger UHMWPE Full or Bowtie ZipLoop™ Construct), Volar Plate Reconstruction (not including Hitch™ L-15)
Hip: Labral
Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | Page 3 of 5 |
|------------------------------------------------------------|-------------|
| (Division Sign-Off) | |
| Division of General, Restorative, and Neurological Devices | |
| 510(k) Number | K080088 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.