SONOTRAX ULTRASONIC FETAL AND VASCULAR POCKET DOPPLER, MODELS LITE, BASIC, PRO, II, II PRO AND VASCULAR
K080087 · Edan Instruments, Inc. · KNG · Mar 11, 2008 · Obstetrics/Gynecology
Device Facts
Record ID
K080087
Device Name
SONOTRAX ULTRASONIC FETAL AND VASCULAR POCKET DOPPLER, MODELS LITE, BASIC, PRO, II, II PRO AND VASCULAR
Applicant
Edan Instruments, Inc.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
Mar 11, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Indications for Use
The 2 mHz and/ or 3 mHz obstetrical probes are indicated for the detection of fetal life from early gestation thru delivery. They can also be used to verify fetal heart viability. The 4 mHz and/ or 8 mHz vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease. The Sonotrax series of pocket dopplers are to be used by health care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physicians assistants, by prescription from licensed physicians in hospitals, clinics and private offices.
Device Story
Handheld pocket Doppler device; uses continuous wave (CW) ultrasound to detect fetal heart rate and peripheral blood flow. Input: ultrasound signals via 2/3 MHz (fetal) or 4/8 MHz (vascular) split-D piezoelectric transducers. Operation: high-frequency oscillator drives transmitter; reflected waves from moving blood cells/fetal heart received by second transducer; signal amplified, filtered, and converted to baseband audio Doppler signal; output via loudspeaker/headphone jack. Fetal heart rate calculated and displayed on LCD. Used in hospitals, clinics, and private offices by healthcare professionals. Provides audible feedback and heart rate data to assist clinicians in assessing fetal well-being and vascular health.
Clinical Evidence
Bench testing only. Device subjected to performance testing and validation under design control procedures to ensure functional requirements and safety standards compliance. No clinical data presented.
Technological Characteristics
Handheld pocket Doppler; split-D piezoelectric transducers (2, 3, 4, 8 MHz); continuous wave (CW) ultrasound; battery-powered; LCD display (except Lite model); audio output via speaker/headphone jack. No complex software/AI mentioned; signal processing is analog/baseband filtering.
Indications for Use
Indicated for detection of fetal life from early gestation through delivery, verification of fetal heart viability, and detection of blood flow in veins and arteries to assist in peripheral vascular disease diagnosis. For use by healthcare professionals (nurses, midwives, technicians, PAs) via physician prescription.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{0}------------------------------------------------
X 510K SUBMISSION
K080087
pg. 1 of 4
510k Submission , SONOT Ultrasonic Fetal And Vascular Pocket Doppler Edan Instruments,Inc.
# SUMMARY
MAR 1 1 2008
This summary of 510k safety and effectiveness information is being submitted in Accordance with 21CFD part 807.92
- Submitters name, address, phone number, contact person and preparation 1. date:
- Name: Edan Instruments.Inc 3/F-B ,Nanshan Medical Equipments Park, Nanhai Road #1019 Shekou, Nanshen, Shenzhen 518067 P.R. China Tel: 86 755 2689 9197 Fax:86 755 2688 2223 Responsible person: Liu Yongying
Official Correspondent:
William Stern
Multigon Industries, Inc.
1 Odell Plaza
Yonkers, N.Y. 10701
Phone: 914 376 5200 x27 Fax: 914376 6111
Date of Preparation: 1/9/08
- 2. Device:
Proprietary Name: Sontotrax Ultrasonic Fetal and Vascular Pocket Doppler Models: Sonotrax Lite, Sonotrax Basic,
> Sonotrax Basic A, Sonotrax Pro, Sonotrax II, Sonotrax II Pro,
Sonotrax Vascular
{1}------------------------------------------------
SONOTRAX 510K SUBMISSION
K080087
pg. 2 of 4 6
Common Name: Handheld Fetal and Peripheral Vascular Doppler
Classification name: Fetal Ultrasonic Monitor and Accessories Ultrasonic Bloodflow Monitor Diagnostic ultrasonic transducer
Product Code:
KNG HEP
Manufactured By: Edan Instruments, Inc., China
Predicate Devices: 3.
K040480 Sonotrax Ultrasonic Pocket Doppler, manufactured by Edan Instruments, Inc.
K024197 LifeDop Doppler Ultrasound System , manufactured by Summit Doppler Systems,
5350 Vivian St., Suite A, Arvada, CO 80002
4. Classification Names :
> Class II as per 21CFR 884-2660, Fetal Ultrasonic Monitor and accessories.(KNG)
Class II as per 21CFR 884-2660, Ultrasonic Bloodflow Monitor (HEP)
5. Description:
> The Sonotrax fetal Dopplers enumerated in their various models in 2. above use the tried and true principle of Dopplershift of an ultrasound signal to detect the fetal heartbeat using the 2 mHz and pr 3 mHz probes and for blood flow detection in veins and artefies using the 4 mHz and 8 mHz probes.
> The Sonotrax Fetal Dopplers uses a split D piezoelectric transducer. A high frequency oscillator supplies a continuous high frequency voltage to one half of the split D transmitter transducer. The high frequency voltage is converted to an ultrasound acoustic wave by the transducer and is transmitted to biophysical objects thru an applied coupling water based medium and moves thru biophysical objects. The acoustic ultrasound is
Image /page/1/Picture/17 description: The image contains handwritten text that appears to be notes or calculations. The text includes numbers such as 2, 3, 4, and 8, followed by the symbols 'mHz' and 'H2'. Additionally, the word 'Heartbeat' is written in a cursive style. The overall impression is that the image captures a snapshot of someone's notes related to measurements or observations, possibly in a scientific or medical context.
{2}------------------------------------------------
### SONOTRAX 510K SUBMISSION
reflected by moving blood cells and moving objects such as the fetal heart. The reflected ultrasound is received by the second split D receiver transducer and is converted via the piezoelectric effect into a high frequency electronic signal. The received electronic signal is amplified and detected. The result is a base band audio Doppler shifted signal which is filtered , and converted to audio via a loud speaker. At the same time the fetal heart rate is applied to and displayed on a liquid crystal counter display.
The Sonotrax product includes four interchangeable probes (2mHz and 3 mHz Obstetrical probes and 4mHz and 8mHz vascular probes).
The user interface includes an on/off button, a play /record button (only in the Sonotrax Pro model), a loud speaker, headphone jack, LCD display for heartrate (all models except Sonotrax lite), and a battery.
#### Indications for use: 6.
The 2 mHz and/ or 3 mHz obstetrical probes are indicated for the detection of fetal life .from early gestation thru delivery and as a general indication of fetal well being. They can also be used to verify fetal heart viability following patient trauma.
The 4 mHz and/ or 8 mHz vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease.
The Sonotrax series of pocket dopplers are to be used by health care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physicians assistants, by prescription from licensed physicians in hospitals, clinics and private offices.
- 7. Contraindications: None known at this time.
Comparison to Predicate Device.
The Sonotrax Ultrasonic Fetal and Vascular Pocket Doppler models in 2. above including the Lite, Basic, Pro, II, and Vascular models have the same device characteristics as all the predicate approved devices in item 3 above. All of these above models use the same technology and circuitry as the already approved Sonotrax Doppler cleared under K040480. In this application we have added a 3 mHz probe frequency for fetal use, and a 4 mHz and 8 mHz vascular probes for vascular use. These Sonotrax models are also similar and comparable to the Lifedop Doppler cleared under K024197. Hence the Sonotrax Fetal and Vascular Pocket Doppler models in paragraph 2 above are substantially equivalent to the predicate devices cited.
{3}------------------------------------------------
### SONOTRAX 510K SUBMISSION
2080087
pg. 4 of 48
#### Test Data: 9.
The Sonotrax Ultrasonic Fetal and Vascular Doppler devices have been designed under design control procedures in the company's QSR system as enumerated in the Cerctificate of Conformance and subjected to extensive safety, performance testing and validations before release. Final testing of the Sonotrax Ultrasonic Fetal and Vascular Dopplers includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications. Safety tests have further been performed to ensure the device complies with applicable industry and safety standards.
The Sonotrax Ultrasonic Fetal and Vascular Doppler device labeling includes instructions for safe and effective use, warnings cautions and guidance for use.
- 10. Literature Review:
A review of the literature pertaining to the safety of Doppler Blood flowmeters has been conducted and appropriate safeguards have been incorporated in the design of the Sonotrax Ultrasonic Fetal and Vascular Fetal Dopplers.
10. Conclusions:
> The conclusion drawn from these tests is that the Sonotrax Ultrasonic Fetal and Vascular Dopplers described herein are equivalent in safety and efficacy to the predicate devices.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 1 2008
Edan Instruments, Inc. c/o Mr. William Stern Official Correspondent Multigon Industries, Inc. 1 Odell Plaza YONKERS NY 10701
Re: K080087
Trade/Device Name: Sonotrax Ultrasonic Fetal and Vascular Pocket Doppler Regulation Number: 21 CFR §884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: January 9, 2008 Received: January 23, 2008
Dear Mr. Stern:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Sonotrax Ultrasonic Fetal and Vascular Pocket Doppler, as described in your premarket notification:
### Transducer Model Number
| 8 mHz CW Vascular Probe | 4 mHz CW Vascular Probe |
|-------------------------|-------------------------|
| 3 mHz CW Fetal Probe | 2 mHz CW Fetal Probe |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. William Stern
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Gamal Akabani, Ph.D. at (240) 276-3666.
Sincerely yours,
Arnthethly
NANCY C. BENDER
ancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{6}------------------------------------------------
## Indication for Use
510(k) Number (if known):
Device Name:
Sonotrax Ultrasonic Fetal and Vascular Pocket Doppler
Indication for Use:
The 2 mHz and/ or 3 mHz obstetrical probes are indicated for the detection of fetal life from early gestation thru delivery. They can also be used to verify fetal heart viability.
The 4 mHz and/ or 8 mHz vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease.
The Sonotrax series of pocket dopplers are to be used by health care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physicians assistants, by prescription from licensed physicians in hospitals, clinics and private offices.
The model includes Sonotrax Lite, Sonotrax Basic. Sonotrax Pro, Sonotrax Basic A, Sonotrax II, Sonotrax II Pro, Sonotrax Vascular.
Prescription Use _ Yes (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use No (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Aonhnh
ive. Abdominal ision of Reproduc Radiological Devic 510(k) Number
{7}------------------------------------------------
SONOTRAX
Appendix F
08008
510(k) Number
### Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer.
> FETAL - PROBE - - MODEL : - SONOTRAX
Diagnostic ultrasound imaging or fluid flow analysis of the human bod
| | | Mode of Operation | | | | | | | | |
|-------------------------------|---|-------------------|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: The above is a 3 mHz CW transducer for FETAL HEART RATE DETECTION MODEL CD3.0-MS3
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
F-3
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
K080087
{8}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer, 8 mHz CW VASCLR PROBE - MODEL - SONOTRAX
Internet CW VASCLR PROBE - MODEL - SONOTRAX
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the hu
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: The above is a 8 mHz CW transducer for
detection of blood flow in veins and arteries
MODEL CD8.0-MS3
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
lom m why
(Division Sign-Off)
Division of Reproductive, Abdominal, F-3 and Radiological Devi 510(k) Number.
{9}------------------------------------------------
SONOTRAX
Appendix F
### Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer, 4 mHz CW VASC one for for facil and assembly system and each transfucer,
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: The above is a 4 mHz CW transducer for
DETECTION OF BLOOD FLOW IN VEINS AND ARTERIES
MODEL CD4.0-MS3
MODEL CD4.0-MS3
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
escription Use (Per 21 CFR 801.109)___________________________________________________________________________________________________________________________________________
Tony M. White
(Division Sign-Off)
E-3 Division of Reproductive, Abdominal, and Radiological Device.
510(k) Number K080087
{10}------------------------------------------------
SONOTRAX
Appendix F
### continious wave Diagnostic Ultrasound Indications for Use
#
tended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | P | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: The above is a 2 mHz CW transducer for fetal heart rate detection. MODEL CD2.0-MS3
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) --
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devi 510(k) Number.
F-3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.