# DEPUY DELTA XTEND REVERSE SHOULDER HIGH MOBILITY CUP (K073676)

_DePuy Orthopaedics, Inc. · PHX · Jan 29, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K073676

## Device Facts

- **Applicant:** DePuy Orthopaedics, Inc.
- **Product Code:** [PHX](/productcode/PHX.md)
- **Decision Date:** Jan 29, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3660
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Delta Xtend Reverse Shoulder prosthesis is intended for use in total or hemi shoulder arthroplasty procedures in patients with non-functional rotator cuffs, with or without bone cement. HA coated components are for cementless use only.

## Device Story

Modular shoulder prosthesis for total or hemi-arthroplasty; designed for patients with non-functional rotator cuffs. System includes humeral stem, epiphysis, and metaglene components; some components feature hydroxyapatite (HA) coating for cementless fixation, while others require bone cement. Device implanted by orthopedic surgeons in clinical/surgical settings to replace damaged shoulder joint structures. Provides mechanical stability and joint function in patients lacking rotator cuff integrity; relies on functional deltoid muscle for operation. Clinical benefit includes restoration of shoulder joint function and pain relief in patients with severe arthropathy or failed prior replacements.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and technological similarities to previously cleared predicate devices.

## Technological Characteristics

Modular shoulder prosthesis; components include humeral stem, epiphysis, and metaglene. Materials include metal and polymer; specific components feature hydroxyapatite (HA) coating for cementless fixation. Designed for semi-constrained use. Sterilization and packaging methods are consistent with predicate devices.

## Regulatory Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Predicate Devices

- DePuy Delta Xtend Reverse Shoulder System ([K062250](/device/K062250.md))
- DePuy Delta Xtend Reverse Shoulder System ([K071379](/device/K071379.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The caduceus is depicted in a simple, black line drawing, and the text is also in black, creating a clear and recognizable emblem.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 2, 2016

DePuy Orthopaedics, Incorporated Ms. Rhonda Myer Regulatory Affairs Associate 700 Orthopaedic Drive Warsaw, Indiana 46581

Re: K073676

Trade/Device Name: DePuv Delta Xtend Reverse Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, HSD Dated: December 21, 2007 Received: January 2, 2008

Dear Ms. Myer:

This letter corrects our substantially equivalent letter of January 29, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# K073676 (pg1/1)

## Indications for Use Statement

510 (k) Number (if known):

Device Name: DePuy Delta Xtend Reverse Shoulder System

Indications for Use:

DePuy

The Delta Xtend Reverse Shoulder prosthesis is indicated for use in a grossly deficient rotator cuff joint with severe arthropathy or a previous failed joint replacement with a grossly deficient rotator cuff joint.

The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.

In cases of bone defects in the proximal humerus, the monobloc implant should be used and then only in cases where the residual bone permits firm fixation of this implant.

Delta Xtend hemi-shoulder replacement is also indicated for hemi-arthroplasty if the glenoid is fractured intraoperatively or for revision of a previously failed Delta Xtend Reverse Shoulder.

The metaglene component is HA coated and is intended for cementless use with the addition of screws for fixation.

The modular humeral stem and epiphysis components are HA coated and intended for cementless use.

All other components are for cemented use only.

| Prescription Use | X (Part 21 CFR 801 Subpart D) |
|------------------|-------------------------------|
|------------------|-------------------------------|

AND/OR

| Over-The-Counter Use | (21 CFR 807 Subpart C) |
|----------------------|------------------------|
|----------------------|------------------------|

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

(Posted November 13, 2003)

Division of General, Restorative,

and Neurological Devices

| Orthopaedics, Inc. | Delta Xtend Reverse Shoulder High Mobility Cup |
|--------------------|------------------------------------------------|
| 510(k) Number      | K073676                                        |
|                    | Page 4 of 54                                   |

{3}------------------------------------------------

# 1,073676 (pg 1/2)

## 510 (k) Summary

(As required by 21 CFR 807.92 and 21 CFR 807.93)

| NAME OF SPONSOR:                    | DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>Warsaw, Indiana 46582<br>Establishment Registration Number: 1818910<br>JAN 29 2008        |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| MANUFACTURER:                       | DePuy France SAS<br>7 Allée Irène Joliot Curie<br>69801 Saint Priest Cedex France<br>Establishment Registration: 3003895575                    |
| 510(K) CONTACT:                     | Rhonda Myer<br>Regulatory Affairs Associate<br>Telephone: (574) 371-4927<br>Facsimile: (574) 371-4987<br>Electronic Mail: Rmyer7@dpyus.jnj.com |
| DATE PREPARED:                      | December 19, 2007                                                                                                                              |
| PROPRIETARY NAME:                   | DePuy Delta Xtend Reverse Shoulder High<br>Mobility Cup                                                                                        |
| COMMON NAME:                        | Shoulder Prosthesis, Polyethylene Cup                                                                                                          |
| CLASSIFICATION:                     | Class II per 21 CFR 888.3660: Prosthesis,<br>Shoulder, Semi-Constrained, Metal/Polymer<br>Cemented                                             |
| DEVICE PRODUCT CODE:                | 87 KWS<br>87 HSD                                                                                                                               |
| SUBSTANTIALLY EQUIVALENT<br>DEVICE: | DePuy Delta Xtend Reverse Shoulder System,<br>K062250 and K071379                                                                              |

### DEVICE DESCRIPTION:

The Delta Xtend Reverse Shoulder System is a modular shoulder prosthesis designed for use in patients with non-functional rotator cuffs.

## INDICATIONS AND INTENDED USE:

### Indications:

The Delta Xtend Reverse Shoulder prosthesis is indicated for use in a grossly deficient rotator cuff joint with severe arthropathy or a previous failed joint replacement with a grossly deficient rotator cuff joint.

{4}------------------------------------------------

K073676 (pg 2/2)

The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.

In cases of bone defects in the proximal humerus, the monobloc implant should be used and then only in cases where the residual bone permits firm fixation of this implant.

Delta Xtend hemi-shoulder replacement is also indicated for hemi-arthroplasty if the glenoid is fractured intraoperatively or for revision of a previously failed Delta Xtend Reverse Shoulder.

The metaglene component is HA coated and is intended for cementless use with the addition of screws for fixation.

The modular humeral stem and epiphysis components are HA coated and intended for cementless use.

All other components are for cemented use only.

### Intended Use:

The Delta Xtend Reverse Shoulder prosthesis is intended for use in total or hemi shoulder arthroplasty procedures in patients with non-functional rotator cuffs, with or without bone cement. HA coated components are for cementless use only.

#### BASIS OF SUBSTANTIAL EQUIVALENCE:

Based on the similarities in intended use, indications for use, materials, design, method of manufacture, sterilization and packaging methods, DePuy believes the subject Delta Xtend Reverse Shoulder High Mobility Cup is substantially equivalent to the previously cleared Delta Xtend Reverse Shoulder System, K062250 and K071379.

---

**Source:** [https://fda-staging.innolitics.com/device/K073676](https://fda-staging.innolitics.com/device/K073676)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).