K073675 · Gaymar Industries, Inc. · ILO · May 8, 2008 · Physical Medicine
Device Facts
Record ID
K073675
Device Name
T-PUMP LOCALIZED TEMPERATURE THERAPY PUMP
Applicant
Gaymar Industries, Inc.
Product Code
ILO · Physical Medicine
Decision Date
May 8, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5720
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The T-Pump Localized Temperature Therapy Pump is intended for use in supplying warm or cold water at controlled temperatures via water circulating pad for the application of localized therapy in situations where a physician determines that temperature therapy is necessary or desirable. Localized temperature therapy is of particular benefit in treating the following; orthopedic conditions such as acute injuries, chronic pain, lower back pain, muscle spasm and strains; skin trauma such as abscesses, boils, bruises, burns and contusions; other medical conditions such as chronic arthritis, neuritis, phlebitis, tendonitis and I.V. infiltration: and symptoms such as infection and localized pain.
Device Story
T-Pump is an electronically controlled water heater/circulator for localized temperature therapy. Device uses heating element to warm water to user-selected set points via digital interface; cooling achieved by adding ice water to reservoir. Dual microprocessor circuit and thermostat manage temperature range (50°F to 107°F) and prevent overheating. Water circulates through a connector hose to a patient-applied pad. Used in clinical settings; operated by healthcare providers. Provides controlled thermal therapy to treat orthopedic injuries, chronic pain, skin trauma, and inflammatory conditions. Benefits include localized symptom relief and management of specific medical conditions.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use.
Technological Characteristics
Water circulating pump; ABS plastic and polycarbonate housing; 11.5 x 8 x 8 inches; <10 lbs; 2750ml reservoir capacity. Heating element; dual microprocessor control circuit; thermostat; digital display. Flow rate 9 gph minimum. Leakage current 100 microamperes max. Temperature range 50°F-107°F.
Indications for Use
Indicated for patients requiring localized temperature therapy for orthopedic conditions (acute injuries, chronic pain, back pain, spasms, strains), skin trauma (abscesses, boils, bruises, burns, contusions), or medical conditions (arthritis, neuritis, phlebitis, tendonitis, I.V. infiltration) and associated symptoms like infection or localized pain.
Regulatory Classification
Identification
A water circulating hot or cold pack is a device intended for medical purposes that operates by pumping heated or chilled water through a plastic bag and that provides hot or cold therapy for body surfaces.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 890.9.
Predicate Devices
Gaymar Industries, Inc. T-Pump Temperature Therapy Pump (K760804)
Adroit Medical Systems, Inc. HTP-1500 Localized Heat Therapy Pump (K970197)
Submission Summary (Full Text)
{0}------------------------------------------------
Ko73675
MAY - 8 2006
# 510(k) SUMMARY
### DATE PREPARED
December 27, 2007
### INTRODUCTION
According to the requirements of 21 CFR 807.92, the following provides sufficient detail to understand the basis for a determination of substantial equivalence.
### CONTACT INFORMATION
Submitter: Gaymar Industries, Inc. 10 Centre Drive Orchard Park, NY 14127 716.662.8636 Contact Person: Greg Pepe Application Correspondent: Hodgson Russ LLP 140 Pearl Street, Suite 100 Buffalo, NY 14202 716.848.1554 Contact Person: Bethany Gilbert, Esq.
#### DEVICE NAME
| Proprietary Name: | T-Pump Localized Temperature Therapy Pump |
|----------------------|-------------------------------------------|
| Catalog Numbers: | TP650, TP650C, TP700, TP700C |
| Common Name: | Temperature Therapy Pump |
| Classification Name: | Pack, Hot or Cold Pack, Water Circulating |
### PREDICATE DEVICE
Gaymar Industries, Inc. ('Gaymar'') claims substantial equivalence to:
- 1. Gaymar Industries, Inc. T-Pump Temperature Therapy Pump (K760804)
- 2. Adroit Medical Systems, Inc. HTP-1500 Localized Heat Therapy Pump (K970197)
### DEVICE DESCRIPTION
The T-Pump Localized Temperature Therapy Pump is a small electronically controlled water heater that supplies warm or cold water at controlled temperatures to a water circulating pad for the application of localized temperature therapy. The control unit is an cloctrical device that uses a heating element to increase the temperature of water to controlled temperature set points,
{1}------------------------------------------------
which the user selects from the digital display face. The temperature is controlled by a dual microprocessor control circuit and a thermostat to prevent overheating. To utilize the cold therapy option, the user adds ice water to the device, which will heat the water to a set point of 50 degrees Fahrenheit. The temperature range of the device ranges from 50 degrees Fahrenheit (for the cooling option) to 107 degrees Fahrenheit (the highest heating set point).
The T-Pump Localized Temperature Therapy Purnp attaches to a connector hose that connects to the water-circulating pad; the temperature-controlled water flows from the pump to the pad. The pads are applied to the part of the body requiring temperature therapy, thereby providing the interface for the therapy.
The T-Pump Localized Temperature Therapy Pump measures, in inches, 11.5 x 8 x 8. The device weighs under 10 pounds with the unit filled with water reservoir has a maximum capacity of 2750ml. The housing is composed of ABS Plastic and Polycarbonate.
The leakage current is 100 microamperes maximum, and the flow rate is 9 gph (34 lph) minimum, with the water circulating pad attached.
### INTENDED USE
The T-Pump Localized Temperature Therapy Pump is intended for use in supplying warm or cold water at controlled temperatures via water circulating pad for the application of localized therapy in situations where a physician determines that temperature therapy is necessary or desirable.
Localized temperature therapy is of particular benefit in treating the following; orthopedic conditions such as acute injuries, chronic pain, lower back pain, muscle spasm and strains; skin trauma such as abscesses, boils, bruises, burns and contusions; other medical conditions such as chronic arthritis, neuritis, phlebitis, tendonitis and I.V. infiltration: and symptoms such as infection and localized pain.
### SUBSTANTIAL EQUIVALENCE
Substantial equivalence is based on two predicate devices.
First, the T-Pump Localized Temperature Therapy Pump is substantially equivalent to the predicate T-Pump Temperature Therapy Pump. Both devices supply warm or cold water to a water-circulating pad for the application of localized heat or cold therapy. None of the device modifications pose risk to safety or effectiveness different from those posed by the predicate T-Pump Temperature Therapy Pump.
Second, the T-Pump Localized Temperature Therapy Pump is substantially equivalent to the predicate Adroit HTP 1500 Localized Heat Therapy Pump. Both devices use a digital interface and software to control the temperature of the circulating water. None of the device
{2}------------------------------------------------
modifications pose a risk to safety or effectiveness different from those posed by the predicate Adroit HTP 1500 Localized Heat Therapy Pump.
.
:
:
. . . . . . . . . .
:
:
1. September 19. 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 199
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Gaymar Industries, Inc. % Hodgson Russ LLP Ms. Bethany Gilbert Associate Attorney 140 Pearl St., Suite 100 Buffalo, New York 14202
MAY -8 2008
K073675 Re:
> Trade Name: T-Pump Localized Temperature Therapy Pump Regulation Number: 21 CFR 890.5720 Regulation Name: Water Circulating Hot, Cold Pack Regulatory Class: Class II Product Code: ILO Dated: March 21, 2008 Received: March 24, 2008
Dear Ms. Gilbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
### Page 2 - Ms. Bethany Gilbert
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Mark N. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name:
T-Pump Localized Temperature Therapy Pump
Indications For Use:
The T-Pump Localized Temperature Therapy Pump is intended for use in supplying warm or cold water at controlled temperatures via water circulating pad for the application of localized therapy in situations where a physician determines that temperature therapy is necessary or desirable.
Localized temperature therapy is of particular benefit in treating the following: orthonedic conditions such as acute injuries, chronic pain, lower back pain, muscle spasm and strains; skin trauma such as abscesses, boils, bruises, burns and contusions; other medical conditions such as chronic arthritis, neuritis, phlebitis, tendonitis and I.V. infiltration; and symptoms such as infection and localized pain.
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Neil RP Oyler for MMM
| Sign-Off | |
|--------------------------------------------------------|---------|
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
| Division of General, Restorative, | |
| Neurological Devices | |
| (k) Number | K073675 |
001978/00406 GBDOCS 910698v2
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.