ARCHITECT HOMOCYSTEINE REAGENTS, CALIBRATORS AND CONTROLS
Applicant
Axis-Shield Diagnostics, Ltd.
Product Code
LPS · Clinical Chemistry
Decision Date
Apr 24, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1377
Device Class
Class 2
Indications for Use
The ARCHITECT Homocysteine assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of total L-homocysteine in human serum or plasma on the ARCHITECT i System. Homocysteine values can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria. The ARCHITECT Homocysteine Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of total l-homocysteine in human serum or plasma. The ARCHITECT Homocysteine Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System (reagents, calibrators and instrument), when used for the quantitative determination of total l-homocysteine in human serum or plasma. For in vitro diagnostic use.
Device Story
ARCHITECT Homocysteine is a chemiluminescent microparticle immunoassay (CMIA) for quantitative determination of total L-homocysteine in human serum or plasma. Operated on ARCHITECT i System; utilizes Chemiflex protocols. Process: bound/dimerized homocysteine reduced by dithiothreitol (DTT) to free homocysteine; converted to S-adenosyl homocysteine (SAH) by recombinant S-adenosyl homocysteine hydrolase (rSAHHase) with excess adenosine. SAH competes with acridinium-labeled S-adenosyl cysteine for particle-bound monoclonal antibody. Magnetic separation and wash performed; trigger solutions added; chemiluminescence measured as relative light units (RLUs). Inverse relationship between homocysteine concentration and RLUs. Used in clinical laboratory settings by trained personnel. Output assists clinicians in diagnosing and managing hyperhomocysteinemia and homocystinuria.
Clinical Evidence
Method comparison study performed on 456 plasma samples (range 3.70 to 49.94 umol/L). Passing-Bablok linear regression analysis showed slope of 0.98 (95% CI: 0.97 to 1.00), intercept of -0.74 (95% CI: -0.99 to -0.54), and r-value of 0.98 (95% CI: 0.98 to 0.99) compared to AxSYM Homocysteine assay.
Technological Characteristics
Chemiluminescent microparticle immunoassay (CMIA). Reagents include recombinant S-adenosyl homocysteine hydrolase (rSAHHase), dithiothreitol (DTT), and acridinium-labeled S-adenosyl cysteine. Automated system (ARCHITECT i System) with magnetic separation and chemiluminescence detection. In vitro diagnostic use.
Indications for Use
Indicated for patients suspected of having hyperhomocysteinemia and homocystinuria. Contraindicated for patients on S-adenosyl-L-methionine therapy due to potential for falsely elevated results; caution advised for patients taking methotrexate, carbamazepine, phenytoin, nitrous oxide, or 6-azuridine triacetate.
Regulatory Classification
Identification
A urinary homocystine (nonquantitative) test system is a device intended to identify homocystine (an analogue of the amino acid cystine) in urine. The identification of urinary homocystine is used in the diagnosis and treatment of homocystinuria (homosystine in urine), a heritable metabolic disorder which may cause mental retardation.
Predicate Devices
AxSYM Homocysteine Assay (k992858)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k073640
B. Purpose for Submission:
New device
C. Measurand:
Homocysteine
D. Type of Test:
Quantitative immunoassay
E. Applicant:
Axis-Shield Diagnostics, Ltd.
F. Proprietary and Established Names:
ARCHITECT Homocysteine Reagents
ARCHITECT Homocysteine Calibrators
ARCHITECT Homocysteine Controls
G. Regulatory Information:
1. Regulation section:
21 CFR §862.1377, Urinary homocystine (non-quantitative test system)
21 CFR §862.1150, Calibrator
21 CFR §862.1660, Quality control material (assayed and unassayed)
2. Classification:
Class II for assay reagents and calibrator
Class I, reserved for control
3. Product code:
LPS, JIT, JJX
{1}
4. Panel: Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
Reagents:
The ARCHITECT Homocysteine assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of total L-homocysteine in human serum or plasma on the ARCHITECT i System. Homocysteine values can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.
Calibrators:
The ARCHITECT Homocysteine Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of total l-homocysteine in human serum or plasma.
Controls:
The ARCHITECT Homocysteine Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System (reagents, calibrators and instrument), when used for the quantitative determination of total l-homocysteine in human serum or plasma.
3. Special conditions for use statement(s):
For prescription use
The labeling contains a prominent black-box warning:
Patients who are on drug therapy involving S-adenosyl-L-methionine may show falsely elevated levels of homocysteine. Specimens from patients taking methotrexate, carbamazepine, phenytoin, nitrous oxide, anticonvulsants and 6-azuridine triacetate may have elevated levels of homocysteine due to their effect on the metabolic pathway.
4. Special instrument requirements:
For use on the ARCHITECT i System
{2}
I. Device Description:
Each Reagent Kit contains enough reagents for 100 tests and consists of 1 bottle of each of the following components:
MICROPARTICLES: 1 Bottle (6.5 mL) Anti-S-adenosyl-L-homocysteine (mouse, monoclonal) coated microparticles in Bis-Tris buffer with surfactants. Minimum Concentration: 0.1% solids. Preservatives: sodium azide and other antimicrobial agents.
CONJUGATE: 1 Bottle (5.7 mL) S-adenosyl-L-cysteine (SAC) acridinium-labeled conjugate in citrate buffer with surfactants. Minimum Concentration: 1 ng/mL. Preservative: ProClin 300.
ENZYME: 1 Bottle (8.6 mL) Recombinant S-adenosyl-L-homocysteine hydrolase (SAHHase) in 4-(2-hydroxyethyl) piperazine-1-propane sulfonic acid (EPPS) buffer. Preservative: sodium azide.
REDUCTANT: 1 Bottle (21.5 mL) Dithiothreitol (DTT) in citrate buffer.
The calibrator and control kits are sold separately. Each Calibrator kit consists of 6 Bottles (3.6 mL each) of ARCHITECT Homocysteine Calibrators consisting of phosphate buffer and S-adenosyl-L-homocysteine. The target values are 0, 2.5, 5, 10, 20 and 50 μmol/L. Each Control kit consists of one bottle (7.7mL each) of each of Low, Medium and High ARCHITECT Homocysteine Controls containing L-homocystine in processed human serum and phosphate buffer.
The human source material was tested using FDA approved methods and found to be nonreactive for HBsAg, HIV-1 RNA or HIV-1 Ag, anti-HIV-1/HIV-2 and anti-HCV antibodies.
J. Substantial Equivalence Information:
1. Predicate device name(s):
AxSYM Homocysteine Assay
2. Predicate K number(s):
k992858
3. Comparison with predicate:
{3}
Similarities
| Parameter | Submission device ARCHITECT Homocysteine | Predicate device AxSYM Homocysteine |
| --- | --- | --- |
| Intended use | Quantitative measurement of total L-homocysteine in human serum or plasma. Homocysteine values can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria. | Same |
| Specimen type | EDTA plasma, Lithium Heparin plasma, Serum, Serum Separator Tubes | Same |
| Capture antibody | Anti-S-adenosyl Homocysteine mouse monoclonal | Same |
| Enzyme conversion | Recombinant S-adenosyl Homocysteine hydrolase | Same |
| Storage conditions | Reagent Pack, Calibrator Pack and Control Pack must be stored at 2-8°C | Same |
| Calibration | Quantitative assay using six gravimetrically prepared S-adenosyl Homocysteine calibrators at 0, 2.5, 5, 10, 20 and 50 μmol/L | Same |
| Assay range with dilution | up to 500 μmol/L | Same |
| Interference | No interference (< 10%) from bilirubin, triglycerides and protein <10% interference from hemoglobin at 1000mg/dL. | Same |
| Specificity | No % cross-reactivity (< 10%) from L-Cysteine, L-Cystathionine, Adenosine, Glutathione and DL-homocysteine thiolactone | Same |
{4}
Differences
| Parameter | Submission device ARCHITECT Homocysteine | Predicate device AxSYM Homocysteine |
| --- | --- | --- |
| Assay Technology | CMIA Chemiluminescent Microparticle Immunoassay | FPIA Fluorescence Polarization Immunoassay |
| Substrate / Signal Generation | Acridinium tracer | Fluorescein tracer |
| Imprecision | Within run %CV from 1.2% to 4.0%, total %CV from 2.1% to 6.3% for samples from 4.71μmol/L to 41.84μmol/L homocysteine | Within run %CV from 1.4% to 4.5%, total %CV from 2.0% to 5.1% for samples from 7.29μmol/L to 28.17μmol/L homocysteine |
| Assay range | 1.00 to 50.00 μmol/L | 0.8 to 50.00 μmol/L |
| Limit of detection | ≤1.0 μmol/L | ≤0.8 μmol/L |
| Specificity | % cross-reactivity: S-adenosyl-L-methionine 11.78% at 0.5mM | % cross-reactivity: S-adenosyl-L-methionine 1.28% at 0.5mM |
# K. Standard/Guidance Document Referenced (if applicable):
- Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline (CLSI EP17-A).
- Interference Testing in Clinical Chemistry; Approved Guideline, Second Edition (CLSI EP7-A2)
# L. Test Principle:
The ARCHITECT Homocysteine assay is a one-step immunoassay for the quantitative determination of total L-homocysteine in human serum or plasma using CMIA (Chemiluminescent microparticle immunoassay) technology. Bound or dimerised homocysteine (oxidized form) is reduced by dithiothreitol (DTT) to free homocysteine, which is then converted to S-adenosyl homocysteine (SAH) by the action of the recombinant enzyme S-adenosyl homocysteine hydrolase (rSAHHase) in the presence of excess adenosine. The SAH then competes with acridinium-labeled S-adenosyl cysteine for particle-bound monoclonal antibody. Following a wash stage and magnetic separation, pre-trigger and trigger solutions are added to the reaction mixture and the resulting chemiluminescence is measured as relative light units
{5}
(RLUs). An indirect relationship exists between the amount of homocysteine in the sample and the RLUs detected by the ARCHITECT $i$ System optics.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
The sponsor states that precision was evaluated following the Clinical and Laboratory Standards Institute (CLSI) Protocol EP5-A2.
Three ARCHITECT Homocysteine Controls Low, Medium, and High as well as five human EDTA plasmas with homocysteine concentrations ranging from 4.71 to $41.84\mu \mathrm{mol} / \mathrm{L}$ were tested in duplicate, two runs per day, using two lots of reagents on two analyzers for 20 days $(n = 80)$ .
The results are summarized in the table below:
| Sample | Instrument | Reagent Lot | Mean (μmol/L) | Within Run | | Between Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Low Control | 1 | 1 | 7.40 | 0.26 | 3.5 | 0.27 | 3.7 | 0.22 | 3.0 | 0.43 | 5.9 |
| | 2 | 2 | 7.58 | 0.18 | 2.4 | 0.18 | 2.3 | 0.00 | 0.0 | 0.25 | 3.3 |
| Medium Control | 1 | 1 | 13.21 | 0.26 | 2.0 | 0.37 | 2.8 | 0.45 | 3.4 | 0.64 | 4.8 |
| | 2 | 2 | 13.37 | 0.24 | 1.8 | 0.25 | 1.9 | 0.21 | 1.6 | 0.41 | 3.0 |
| High Control | 1 | 1 | 26.77 | 0.63 | 2.3 | 0.72 | 2.7 | 0.52 | 1.9 | 1.09 | 4.1 |
| | 2 | 2 | 25.73 | 0.47 | 1.8 | 0.43 | 1.7 | 0.37 | 1.5 | 0.73 | 2.9 |
| Panel 1 | 1 | 1 | 4.78 | 0.19 | 4.0 | 0.24 | 4.9 | 0.00 | 0.0 | 0.30 | 6.3 |
| | 2 | 2 | 4.71 | 0.13 | 2.8 | 0.01 | 0.3 | 0.12 | 2.6 | 0.18 | 3.8 |
| Panel 2 | 1 | 1 | 11.03 | 0.22 | 2.0 | 0.32 | 2.9 | 0.28 | 2.5 | 0.48 | 4.3 |
| | 2 | 2 | 10.89 | 0.13 | 1.2 | 0.17 | 1.6 | 0.08 | 0.7 | 0.23 | 2.1 |
| Panel 3 | 1 | 1 | 17.60 | 0.43 | 2.4 | 0.47 | 2.7 | 0.43 | 2.5 | 0.77 | 4.4 |
| | 2 | 2 | 17.29 | 0.27 | 1.6 | 0.21 | 1.2 | 0.30 | 1.7 | 0.46 | 2.6 |
| Panel 4 | 1 | 1 | 35.40 | 0.79 | 2.2 | 0.78 | 2.2 | 0.44 | 1.2 | 1.19 | 3.4 |
| | 2 | 2 | 34.83 | 0.71 | 2.0 | 0.32 | 0.9 | 0.39 | 1.1 | 0.87 | 2.5 |
| Panel 5 | 1 | 1 | 41.84 | 0.71 | 1.7 | 1.15 | 2.8 | 0.00 | 0.0 | 1.35 | 3.2 |
| | 2 | 2 | 41.46 | 0.68 | 1.6 | 0.73 | 1.8 | 0.50 | 1.2 | 1.12 | 2.7 |
# b. Linearity/assay reportable range:
The reportable range is stated to be $1 - 50\mu \mathrm{mol} / \mathrm{L}$ . The claim is supported by data from the sponsor's dilution, method comparison study (see section 2.a. below) and limit of detection study (see section 1.d below). Three EDTA plasma sample were spiked with L-homocysteine to give the following
{6}
concentrations: 41.97, 44.23 and 45.79 µmol/L. Each was diluted with ARCHITECT wash buffer in 10% increments, ranging from 9:1 to 1:9. All samples and dilutions were evaluated with the ARCHITECT Homocysteine test. The observed results were compared to the expected results and summarized in the table below.
| Sample | Dilution factor | Expected value (µmol/L) | Rep 1 | Rep 2 | %CV | Mean Observed Value (µmol/L) | % Recovery |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | Undiluted | 41.97 | 42.74 | 41.21 | 2.58 | 41.97 | 100.0 |
| | 9:1 | 37.78 | 38.28 | 37.13 | 2.17 | 37.71 | 99.8 |
| | 8:2 | 33.58 | 34.15 | 34.07 | 0.17 | 34.11 | 101.6 |
| | 7:3 | 29.38 | 29.87 | 29.63 | 0.56 | 29.75 | 101.3 |
| | 6:4 | 25.18 | 26.29 | 25.87 | 1.13 | 26.08 | 103.6 |
| | 5:5 | 20.99 | 21.55 | 21.99 | 1.44 | 21.77 | 103.7 |
| | 4:6 | 16.79 | 17.70 | 17.73 | 0.12 | 17.72 | 105.5 |
| | 3:7 | 12.59 | 12.85 | 13.76 | 4.83 | 13.31 | 105.7 |
| | 2:8 | 8.39 | 9.04 | 9.10 | 0.50 | 9.07 | 108.1 |
| | 1:9 | 4.20 | 4.60 | 4.63 | 0.51 | 4.62 | 110.0 |
| | diluent | 0.00 | 0.05 | -0.28 | N/A | -0.11 | N/A |
| | | | | | | Mean | 103.9 |
| 2 | Undiluted | 44.23 | 44.02 | 44.44 | 0.68 | 44.23 | 100.0 |
| | 9:1 | 39.81 | 40.71 | 39.53 | 2.07 | 40.12 | 100.8 |
| | 8:2 | 35.38 | 36.22 | 36.39 | 0.33 | 36.31 | 102.6 |
| | 7:3 | 30.96 | 32.43 | 32.18 | 0.54 | 32.30 | 104.3 |
| | 6:4 | 26.54 | 28.44 | 27.92 | 1.31 | 28.18 | 106.2 |
| | 5:5 | 22.11 | 22.72 | 22.90 | 0.56 | 22.81 | 103.1 |
| | 4:6 | 17.69 | 19.05 | 19.19 | 0.52 | 19.12 | 108.1 |
| | 3:7 | 13.27 | 14.33 | 14.17 | 0.83 | 14.25 | 107.4 |
| | 2:8 | 8.85 | 9.98 | 9.97 | 0.07 | 9.98 | 112.8 |
| | 1:9 | 4.42 | 5.05 | 5.26 | 2.86 | 5.15 | 116.5 |
| | diluent | 0.00 | -0.29 | -0.18 | N/A | -0.24 | N/A |
| | | | | | | Mean | 106.2 |
| 3 | Undiluted | 45.79 | 45.81 | 45.77 | 0.07 | 45.79 | 100.0 |
| | 9:1 | 41.21 | 41.97 | 40.95 | 1.74 | 41.46 | 100.6 |
| | 8:2 | 36.63 | 36.22 | 36.89 | 1.31 | 36.55 | 99.8 |
| | 7:3 | 32.05 | 33.47 | 32.32 | 2.48 | 32.90 | 102.6 |
| | 6:4 | 27.48 | 27.31 | 27.28 | 0.09 | 27.29 | 99.3 |
| | 5:5 | 22.90 | 23.18 | 23.94 | 2.28 | 23.56 | 102.9 |
| | 4:6 | 18.32 | 19.25 | 19.68 | 1.54 | 19.46 | 106.3 |
| | 3:7 | 13.74 | 14.81 | 14.95 | 0.63 | 14.88 | 108.3 |
| | 2:8 | 9.16 | 10.49 | 10.22 | 1.89 | 10.35 | 113.1 |
| | 1:9 | 4.58 | 5.13 | 4.74 | 5.57 | 4.94 | 107.8 |
| | diluent | 0.00 | -0.32 | -0.19 | N/A | -0.26 | N/A |
| | | | | | | Mean | 104.1 |
{7}
Specimens with homocysteine values exceeding 50 μmol/L are flagged by the analyzer. They may be automatically diluted by the analyzer 1:10 with ARCHITECT Wash Buffer or manually using ARCHITECT Multi-Assay Manual Diluent (which is identical in formulation as the Wash Buffer). The sponsor performed a study to demonstrate that auto-diluted results are within 10% of results if samples are diluted manually.
| Sample | Auto diluted value μmol/L | Manual dilution value μmol/L | % recovery |
| --- | --- | --- | --- |
| 1 | 401.9 | 440.6 | 91.2 |
| 2 | 411 | 448.5 | 91.6 |
| 3 | 405.6 | 444.6 | 91.2 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods): Calibrators and controls were previously cleared under k980812 and k992858.
## Traceability and Value Assignment:
Calibrators are prepared from in-house S-adenosyl-L-homocysteine by gravimetric addition of the compound to phosphate buffer. Calibrators are tested and compared to frozen internal reference calibrators. Kit calibrators A-F are assigned values 0, 2.5, 5, 10, 20 and 50 μmol/L, respectively.
Controls are prepared by gravimetric addition of L-homocysteine to buffered serum then are adjusted to within a specified concentration range around target means for controls low, medium and high respectively.
Shelf-life and in-use stability is supported by real time and accelerated stability studies. The protocols and acceptance criteria were reviewed and determined to be adequate.
d. Detection limit:
The sponsor states that determination of Limit of Blank (LoB) and Limit of Detection (LoD) followed guidance from CLSI document EP17-A.
A sample panel was prepared by diluting an EDTA plasma sample spiked with ~5.4 μmol/L homocysteine with ARCHITECT on-board wash buffer, to create a series of samples with very low concentrations of homocysteine. ARCHITECT Homocysteine Calibrators A to F (ranging in value from 0-50 μmol/L), Controls Low, Medium and High and the plasma sample panel were tested in a series of experiments over 20 days, using two ARCHITECT instruments and two lots of reagents. The data from a total of n=40 measurements for Calibrators B-F, controls, and plasma samples and n=120 measurement of Calibrator A were analyzed.
{8}
The limit of blank (LOB) was calculated using the following equation and was determined to be 0.30 μmol/L.
$$
LOB = Mean Cal A result + 1.645 \times SD
$$
The mean and Standard Deviation (SD) was calculated for the pooled data from each sample in the plasma panel. Then the following equation was applied:
$$
LOD = LOB + 1.645 \times pooled SD
$$
The Limit of Detection was determined to be 0.64 μmol/L and is below the claimed limit of detection of ≤ 1.00 μmol/L.
The analyzer reports sample results ≥ 1 μmol/L.
e. Analytical specificity:
Interference studies were designed using CLSI EP7-A. Five (5) EDTA plasma samples with homocysteine concentrations ranging from 3.5 to 40.5 μmol/L were spiked with the bilirubin, hemoglobin, lipid or protein (bovine gamma globulin) and compared to the same sample without the added compound. No significant interference was defined as the observed value within ± 10% of the control value. The results obtained indicated that there was no significant interference by the following interferents:
- Hemoglobin up to 1000 mg/dL
- Bilirubin up to 20 mg/dL
- Triglycerides up to 6000 mg/dL
- Protein up to 12 g/dL
- Heparin up to 1 U/mL
The cross-reactivity of six structurally related compounds was evaluated using the ARCHITECT Homocysteine assay. Plasma samples with normal levels of homocysteine were spiked with each compound and these were compared to control sample spiked with an equivalent volume of saline. The results are summarized in the table below.
| Substance | Concentration of substance tested | % Cross- Reactivity |
| --- | --- | --- |
| S-adenosyl-L-methionine | 0.5 mM | 11.78 |
| L-Cysteine | 100 mM | 0.01 |
| L-Cystathionine | 0.5 mM | 0.30 |
{9}
| Adenosine | 5 mM | 0.72 |
| --- | --- | --- |
| Glutathione | 100 mM | 0.003 |
| DL-homocysteine thiolactone | 0.25 mM | 3.22 |
Under the “Limitations of the procedure” section of the package insert, the following statements have been provided for end-users:
"S-adenosyl-L-methionine is an antidepressant whose molecular form is similar to S-adenosyl-L-homocysteine. This drug may interfere with the ARCHITECT Homocysteine assay."
"The following drugs may elevate levels of homocysteine: methotrexate, carbamazepine, phenytoin, nitrous oxide, and azuridine triacetate. The mechanism of action of these drugs affects different parts of the metabolic pathway of homocysteine."
The package insert also contains language to alert users that specimens from patients containing human anti-mouse antibodies or heterophilic antibodies may interfere with the ARCHITECT Homocysteine assay and produce erroneous results.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The sponsor states that the method comparison study was based on guidance from CLSI protocol EP9-A2.
A total of 456 samples from apparently healthy male and female donors were obtained from a US sample bank. Of these samples 292 were spiked with L-homocysteine to obtain samples across the assay range.
Samples covered the range 3.18 to 49.39 μmol/L (ARCHITECT) and 3.70 to 49.94 μmol/L (AxSYM) in order to support the measuring range for the assay up to 50.00 μmol/L. Data from this study are summarized in the following table.
{10}
| ARCHITECT Homocysteine vs AxSYM Homocysteine | | | |
| --- | --- | --- | --- |
| Number of Observations | Slope (95% CI) | Intercept (95% CI) | Correlation Coefficient |
| 456 | 0.98 | -0.74 | 0.98 |
| | (0.97-1.00) | (-0.99 to -0.54) | (0.98-0.99) |
b. Matrix comparison:
Matrix comparison studies were performed by collecting fresh whole blood with each of the following collection tubes: Potassium EDTA plasma, Lithium Heparin plasma, Serum and Serum Separator Tube (SST). Samples were collected on ice. A total of forty (40) matched sample set were evaluated over two assay runs. Twenty (20) samples were natural and 20 were spiked with L-homocysteine to cover the range of the assay $(6 - 48\mu \mathrm{mol} / \mathrm{L})$ . The percent recovery for each tube-type was calculated by comparing the mean value of the EDTA tube type sample to the mean value for the corresponding Serum, SST, and Lithium Heparin tube type sample
When comparing homocysteine results for potassium EDTA plasma samples (control tube type) to the same samples collected in serum clot, serum separator and lithium heparin tubes, all but one sample recovered within $\pm 10\%$ of the control. One lithium heparin sample with homocysteine value approximately $13~\mu \mathrm{mol / L}$ , recovered within $112\%$ of the potassium EDTA control sample.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
{11}
Human EDTA plasma specimens from 300 apparently healthy individuals in the United States, 150 men and 150 women, were evaluated with the Architect Homocysteine assay.
The Expected Ranges / Reference Intervals were taken as the central 95% of the observations (2.5th percentile to 97.5th percentile) as detailed in the table below:
Expected Range / Reference Intervals
| Sex | N | Median (μmol/L) | Percentile | |
| --- | --- | --- | --- | --- |
| | | | 2.5% (μmol/L) | 97.5% (μmol/L) |
| Male | 150 | 9.05 | 5.46 | 16.20 |
| Female | 150 | 7.61 | 4.44 | 13.56 |
| Overall | 300 | 8.14 | 5.08 | 15.39 |
The package insert recommends that each laboratory establish its own expected range since homocysteine values can vary depending on geographical, patient, dietary, and environmental factors.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.