TINNITUS SOUND GENERATOR MODULE

K073636 · Gn Resound A/S · KLW · Mar 13, 2008 · Ear, Nose, Throat

Device Facts

Record IDK073636
Device NameTINNITUS SOUND GENERATOR MODULE
ApplicantGn Resound A/S
Product CodeKLW · Ear, Nose, Throat
Decision DateMar 13, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3400
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Tinnitus Sound Generator Module is a tool to generate sounds to be used in a tinnitus management program to relieve patients suffering from tinnitus. The target population is primarily the adult population over 18 years of age. This product may also be used with children 5 years of age or older. The Tinnitus Sound Generator module is targeted for healthcare professionals, which are treating patients suffering from tinnitus, as well as conventional hearing disorders. The fitting of the Tinnitus Sound generator module must be done by hearing professional participating in a tinnitus management program.

Device Story

TSG Module is a software-based sound generator embedded within digital hearing instruments; designed for all-day wear in various environments. Healthcare professionals (audiologists/physicians) fit the module to individual patient needs as part of a tinnitus management program. Device generates sounds to shift patient focus away from perceived tinnitus in quiet environments; does not aim to mask tinnitus completely or impact speech perception thresholds. Optional volume control allows output level adjustment; for pediatric or challenged users, volume control can be configured to only decrease output to prevent unintended usage. Children and physically/mentally challenged users require guardian supervision and training for device insertion/removal. Benefits include tinnitus relief through sound therapy and counseling.

Clinical Evidence

Bench testing only. No clinical data provided. Functionality verified by the device's ability to provide sound stimuli to shift tinnitus emphasis in quiet environments.

Technological Characteristics

Software module embedded in digital hearing instruments. Provides sound generation for tinnitus relief. Features optional user-adjustable volume control. Configurable by healthcare professionals. Operates as part of a hearing aid platform.

Indications for Use

Indicated for patients suffering from tinnitus, including adults over 18 and children 5 years or older. Requires fitting by a hearing professional participating in a tinnitus management program.

Regulatory Classification

Identification

A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.

Special Controls

The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # # K073636 510(k) Notification for TSG Module 10 December, 2007 Rev. 31 January 2008 Rev. 12 February 2008 ## 510(k) SUMMARY | SUBMITTER: | GN Resound A/S<br>Lars Hagander | MAR 13 2008 | |-----------------------|----------------------------------|-------------| | DATE PREPARED: | 10 December, 2007 | | | DEVICE NAME: | TSG Module | | | CLASSIFICATION NAMES: | Tinnitus masker | | | CLASSIFICATION CODE: | KLW | | | PREDICATE DEVICES: | Quiescence and Siemens TCI-Combi | | #### Device Description: The Tinnitus Sound Generator provides a means for healthcare professionals to create a hearing instrument solution that provide relief for tinnitus patients. This software solution is embedded into a digital hearing instrument platform, so that the end-user can wear this device all day in all environments. The fitting of the mistrations portions, so that the sound generator module is fitted by a healthcare professional to fit the exact needs of the tinnitus patient. ### Predicate Devices: Quiescence K040330 Siemens TCI-Combi K003558 ### Intended Use: The Tinnitus Sound Generator Module is a tool to generate sounds to be used in a tinnitus management ITIC Thinds Sound Generator Nieders is a tool ... The target population is primarily the adult population program to reneve pantis product may also be used with children 5 years of age or older. ## Summary of Non-Clinical Tests: The tinnitus sound generator is providing means to help tinnitus patients under the Tinnitus Retraining The timmus sound generator is provides neals to help not to provide a high masking sound of the timitus I herceived by the end-user, thereby causing implication of the performance of the end-user speech perception thresholds, but to provide them means to shift emphasis from the timitus dominantly perceived perception thresnows, out to provide in quiet environments. It the assumption that the timitus patients are not bothered significantly by their perceived timitus, when exposed to other sound stimuli originating from the environment. The functionality of the tinnitus sound generator is fulfilling its purpose, as providing means of shifting tinnitus emphasis, when in quiet environment. fitcals of sinting thinked emplaasis with counselling done by the healthcare professional. Page 44 of 49 {1}------------------------------------------------ 510(k) Notification for TSC Module 10 December, 2007 Rev. 31 January 2008 Rey. 12 February 2008 ### Other: Pedriatic use precautions The target population is primarily the adult population over 18 years of age. This product may also be used with children 5 years of age or older. However, children and physically or mentally challenged users will require training by a doctor, audiologists, hearing care practitioner or the guardian for the insertion and removal of the device containing the TSG module. Children and physically or mentally challenged users will require guardian supervision while wearing the device. The volume control is an optional feature in the TSG module used for adjusting the sound generator output level. To prevent unintended usage by pediatric or physically or mentally challenged users, the volume control must, if enabled, be configured to only provide a decrease of the sound generator output level, #### Conclusions: Testing performed on the TSG Module indicates that they are safe, effective, and perform as well as the predicate devices, when used in accordance with the instructions for use. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized depiction of an eagle or bird-like figure, represented by three curved lines that suggest wings and a tail. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 GN ReSound A/S c/o Lars Hagander Lautrupbjerg 9 DK-2750 Ballerup Denmark MAR 1 3 2 2008 Re: K073636 Trade/Device Name: Tinnitus Sound Generator Module (TSG Module) Regulation Number: 21 CFR 874.3400 Regulation Name: Tinnitus masker Regulatory Class: Class II Product Code: KLW Dated: December 20, 2007 Received: December 26, 2007 Dear Mr. Hagander: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, M. B. Egerlton, mD Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Notification for TSG Module 10 December, 2007 Rev. 31 January 2008 KO1363 # ndications for Us 510(k) Number (if known): K07 3636 Device Name: TSG Module Indications for Use: The Tinnitus Sound Generator Module is a tool to generate sounds to be used in a tinnitus management program to relieve patients suffering from tinnitus. The target population is primarily the adult population over 18 years of age. This product may also be used with children 5 years of age or older. The Tinnitus Sound Generator module is targeted for healtheare professionals, which are treating patients suffering from tinnitus, as well as conventional hearing disorders. The fitting of the Tinnitus Sound generator module must be done by hearing professional participating in a tinnitus munagement program. CFR 801 Subpart D Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) Shullon Peng ision Sign sion Sign-Of Ophthalmic Ton of Ophthalmic 510(k) Numbe Page 49 of 49 Page . Prescription Use (Per 21 CFR 801.109)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...