The ICI Series P and S IR Cameras, which provides capture of skin surface temperature of any part of the body, and the IR Flash Software version 1.0, which provides visualization and reporting functionalities, are intended for use as an adjunct to other clinical diagnostic procedures in the diagnosis, quantifying, and screening of relative skin surface temperature. Environment of use: hospitals, sub-acute healthcare settings, public areas, i.e., airport.
Device Story
ICI P and S Series IR Cameras capture skin surface temperature; IR Flash Software v1.0 provides visualization and reporting. System uses uncooled microbolometer (vanadium oxide) to detect infrared radiation (8-14 um). Data output via USB 2.0 to a PC running Windows XP/Vista. Used in hospitals, sub-acute settings, and public areas (e.g., airports) by clinicians or trained operators. Output displays thermal patterns; assists healthcare providers in screening or quantifying relative temperature changes as an adjunct to other diagnostic procedures. Benefits include non-invasive, real-time thermal monitoring.
Clinical Evidence
No clinical data. Bench testing only, including compliance with IEC-60601-1 and IEC 60601-1-2, and software validation.
Technological Characteristics
FPA uncooled microbolometer; Vanadium Oxide material; 8-14 um spectral response; 320x240 array; 1.13 mrad IFOV; 38 mK NETD; USB 2.0 connectivity; IP54 encapsulation; manual focus; 5 VDC power via USB; Windows XP/Vista compatible software.
Indications for Use
Indicated for use as an adjunct to clinical diagnostic procedures for the diagnosis, quantification, and screening of relative skin surface temperature in any part of the body. Suitable for use in hospitals, sub-acute healthcare settings, and public areas (e.g., airports).
Regulatory Classification
Identification
A telethermographic system for adjunctive diagnostic screening for detection of breast cancer or other uses is an electrically powered device with a detector that is intended to measure, without touching the patient's skin, the self-emanating infrared radiation that reveals the temperature variations of the surface of the body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
#### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
510(k) number:
#### 1. Submitter's Identification:
JUL 11 2008
Texas Infrared 2105 West Cardinal Drive Beaumont, TX 77705 Contact: Gary Strahan President/CEO Phone: 409-861-0788 FAX: 409-866-7229 Date Summary Prepared: 7-27-2007
# 2. Name of the Device:
Common Name: Telethermographic System (Adjunctive Use) Trade or Proprietary Name: ICI P and S Series IR Camera(s) and the IR Flash Software version 1.0 Device Class: 1 Product Code: LHQ Regulation Number: 884.2980
### 4. Predicate Device Information:
Trade Name: A20M FLIR Systems, Inc. 27700A SW Parkway Avenue Wilsonville, OR 97070 USA Phone: +1 800 322.3731 Fax: +1 503.498.3904 510(k) number: K033967
# 5. Device Description:
The ICI Series P and S IR Cameras, which provides capture of skin surface temperature of any part of the body, and the IR Flash Software version 1.0, which provides visualization and reporting functionalities, are intended for use as an adjunct to other clinical diagnostic procedures in the diagnosis, quantifying, and screening of relative skin surface temperature.
Environment of use: hospitals, sub-acute healthcare settings, public areas, i.e., airport.
{1}------------------------------------------------
| ICI P and S Series Cameras | |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Base IR Camera | ICI P and S series Cameras |
| Intended Use | The ICI Series P and S IR Cameras, which provides capture of skin<br>surface temperature of any part of the body, and the IR Flash<br>Software version 1.0, which provides visualization and reporting<br>functionalities, are intended for use as an adjunct to other clinical<br>diagnostic procedures in the diagnosis, quantifying, and screening of<br>relative skin surface temperature. |
| | Environment of use: hospitals, sub-acute healthcare settings, public<br>areas, i.e., airport. |
| Technology | FPA uncooled Microbolometer |
| Material | Vanadium Oxide |
| Spectral Response | 8 - 14 um |
| Contrast/Brightness | Software Controlled |
| Spatial Resolution IFOV | 1.13 mrad |
| Data Output | Digital USB 2.0 |
| Thermal Time Constant | 14 ms |
| Thermal Sensitivity | 0.038C @ 25C |
| Accuracy | +2C or 2% |
| Emissivity Correction | Computer Controlled |
| Performance | 38 mK NETD |
| Frame Rate | S Series Camera 50-60fps<br>P Series Camera < 9 fps |
| Pitch Size | 25 um |
| Encapsulation | IP54 |
| Optics | 25mm with 22° FOV |
| Vibration | 3G's |
| Shock | 30G's |
| Array size | 320 x 240 array |
| Weight | 5.2oz (148g) w/lens |
| Tripod Mount | 1/4" - 20 female thread |
| Operating Temperature | -20C to +50C |
| Storage Temperature | -40C to +70C |
| Dimensions | 2.1"x3.2"x0.5" |
| Focus | Manual |
| Special Computer Hardware | USB 2.0 Compatible running Windows XP or Windows Vista. 64-bit<br>operating systems are not supported at this time. |
| Power Supply | 5 VDC @ 500ma max draw from USB. Systems with motorized focus have an additional 12 VDC supply @ 1A max. |
:
{2}------------------------------------------------
#### Components of the system include:
ICI P or S Series IR Camera USB Cable Tripp-Lite model IS250HG isolation transformer
### 6. Intended Use:
The ICI Series P and S IR Cameras, which provides capture of skin surface temperature of any part of the body, and the IR Flash Software version 1.0, which provides visualization and reporting functionalities, are intended for use as an adjunct to other clinical diagnostic procedures in the diagnosis, quantifying, and screening of relative skin surface temperature.
Environment of use: hospitals, sub-acute healthcare settings, public areas, i.e., airport.
# 7. Comparison to Predicate Devices:
# Technical characteristics of the device(s) compared to the predicate device:
| Base IR Camera | ICI P and S Series IR<br>Camera(s) and the IR Flash<br>Software version 1.0 | A20M<br>(K033967) |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The ICI Series P and S IR Cameras,<br>which provides capture of skin<br>surface temperature of any part of the<br>body, and the IR Flash Software<br>version 1.0, which<br>provides visualization and reporting<br>functionalities, are intended for use as<br>an adjunct to other clinical diagnostic<br>procedures in the diagnosis,<br>quantifying, and screening of relative<br>skin surface temperature.<br>Environment of use: hospitals, sub-<br>acute healthcare settings, public<br>areas, i.e., airport. | The Flir devices are intended for use<br>as an adjunct to other clinical<br>diagnostic procedures in the<br>diagnosis, quantifying, and screening<br>of differences in skin surface<br>temperature changes. It can<br>visualize, document temperature<br>patterns and changes. Environment<br>of use: hospital, sub-acute, public<br>areas, i.e., airports |
| Technology | FPAuncooled Microbolometer | FPA uncooled Microbolometer |
| Material | Vanadium Oxide | Amorphous Silica |
| Spectral Response | 8-14 um | 7.5-13um |
| Contrast/Brightness | Software Controlled | Manual or Software Controlled |
| Spatial Resolution IFOV | 1.13 mrad | 2.7 mrad |
| Data Output | Digital USB 2.0 | RS170 EIA/NTSC or CCIR/PAL<br>composite |
| Thermal Time Constant | 14 ms | Unknown |
| Thermal Sensitivity | 0.038C @ 25C | 0.120C @ 30C |
| Accuracy | +2C or 2% | +2C or 2% |
| Emissitivity Correction | Computer Controlled | Variable from 0.1 to 1.0 |
| Performance | 38 mK NETD | Under <80mK |
{3}------------------------------------------------
| Frame Rate | S Series Camera 50-60fps<br>Series Camera < 9fps | 60 fps |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Pitch Size | 25 um | Unknown |
| Encapsulation | IP54 | IP40 |
| Optics | 25mm with 22° FOV | 25mm with 19° FOV |
| Vibration | 3G's | 2G's |
| Shock | 30G's | 25G's |
| Array size | 320 x 240 array | 320 x 240 array |
| Weight | 5.2oz (148g) w/lens | 1.7lbs (0.8 kg) |
| Tripod Mount | 1/4" -20 female thread | 1/4" -20 female thread |
| Operating Temperature | -20C to +50C | -15C to +50C |
| Storage Temperature | -40C to +70C | -40C to +70C |
| Dimensions | 2.1"x3.2"x0.5" | 6.2"x2.9"x3.1" |
| Focus | Manual | Manual or Software Controlled |
| Special Computer<br>Hardware | USB 2.0 Compatible running<br>Windows XP or Windows Vista. 64-<br>bit operating systems are not<br>supported at this time. | Ethernet connection, Video capture<br>device to vonvert RS-170 video,<br>Firewire 8/16-bit monochrome and 8-<br>bit color |
| Power Supply | 5 VDC @ 500ma max draw from<br>USB. Systems with motorized focus<br>have an additional 12 VDC supply @<br>1 A max. | AC adaptor 110/220 vac, 50/60hz<br>input to 12/24vdc nominal, <6w<br>output |
### 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The devices have been tested and found to comply with IEC-60601-1 and IEC 60601-1-2. Software validation was performed.
### 9. Discussion of Clinical Tests Performed:
Not applicable
#### 10. Conclusions:
The subject device(s) has the same intended use and similar characteristics as the predicate device. Moreover, documentation supplied in this submission demonstrates that any difference in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the ICI P and S Series IR Camera(s) and the IR Flash Software version 1.0 are substantially equivalent to the predicate device.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 11 2008
Texas Infrared % Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services NA, Inc. 2307 E. Aurora Rd., Unit B7 TWINSBURG OH 44087
Re: K073581
Trade/Device Name: ICI P and S Series IR Camera(s) and the IR Flash Software version 1.0 Regulation Number: 21 CFR 884.2980 Regulation Name: Telethermographic System Regulatory Class: I Product Code: LHO Dated: June 27, 2008 Received: June 30, 2008
#### Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Aut or any Federal statutes and regulations administered by other Federal agencies. Ynoundus of comply with all the Act's requirements, including, but not limited to: registration and liasting (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the lectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permoits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Of Crice of Compliance a to one of the following numbers, based on the regulation number at the top of this Utter
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 241-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You max cobtain other general information on your responsibilities under the Division of Simil Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-241 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE
#### K073581 510(k) Number (if known):
Device Name : ICI P and S Series IR Camera(s) and the IR Flash Software version 1.0
#### Indications for Use:
The ICI Series P and S IR Cameras, which provides capture of skin surface temperature of any part of the body, and the IR Flash Software version 1.0, which provides visualization and reporting functionalities, are intended for use as an adjunct to other clinical diagnostic procedures in the diagnosis, quantifying, and screening of relative skin surface temperature. Environment of use: hospitals, sub-acute healthcare settings, public areas, i.e., airports.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
hozu Mnhany
(Divis on Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.