The Audicor® System, when used with AUDICOR Sensors on the chest wall, is intended for use in acquiring, analyzing and reporting ECG and heart sound data and to provide interpretation of the data in an integrated ACG (acoustic cardiograph) report for consideration by physicians. Audicor systems allow detection, reporting and interpretation of standard ECG data as well as advanced parameters such as EMAT, LVST, S3 strength, S4 strength, and SDI. Data may be reported in a single snapshot report and in a trended report format where multiple data points are trended over time. The Audicor system also accepts and analyzes downloads of up to 48 hours of patient data from an optional ambulatory ECG/heart sounds collection device. ECG and heart sound data offered by the device are only significant when used in conjunction with physician over read as well as consideration of other relevant patient data. The device is intended for use only under the direct supervision of a physician and is for use on adults (≥ 18 years).
Device Story
Audicor system captures ECG and heart sound data via chest-wall sensors; processes inputs to calculate advanced cardiac parameters including EMAT, LVST, S3/S4 strength, and Systolic Dysfunction Index (SDI). System operates as a standalone unit for 10-second snapshots or with an optional ambulatory harness for up to 48-hour data collection. Data is analyzed and presented in snapshot or trended reports. Used in clinical settings under physician supervision to monitor patient changes during therapeutic treatment or CRT studies. Output assists physicians in evaluating systolic dysfunction and cardiac performance; provides objective data to support clinical decision-making regarding heart failure or acute coronary syndrome management.
Clinical Evidence
Bench testing only. The device was tested for compliance with ANSI/AAMI EC-11, EN 60601-1, EN 60601-1-2, IEC 60601-2-25, and IEC 60601-2-51. The ambulatory harness was tested for compliance with ANSI/AAMI EC38:1998.
Technological Characteristics
Wearable, battery-powered ambulatory monitor. Uses proprietary chest-wall sensors for simultaneous ECG and heart sound acquisition. Data stored on non-volatile memory for download to analysis module. Complies with AAMI EC-11, IEC 60601-2-25, and IEC 60601-2-51 standards.
Indications for Use
Indicated for adults (≥ 18 years) suspected of heart failure or acute coronary syndrome. Used for acquisition, analysis, and reporting of ECG and heart sound data to assist physicians in clinical assessment.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Predicate Devices
GE Medical Systems SEER Light Holter Recorder (K021470)
Submission Summary (Full Text)
{0}------------------------------------------------
:
·
:
SEP 0 9 2008
| Summary of Safety and Effectiveness | | SEP 09 2008 |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| As required by 21 CFR, part 807.92 | | |
| Submitted By: | Inovise Medical, Inc.<br>10565 SW Nimbus Ave, Suite 100<br>Portland, OR 97233-4311<br>Phone 503-431-3849<br>Fax 503-431-3801 | |
| Contact: | Kendra Rathkey<br>Manager, Quality and Regulatory | |
| Date Prepared: | July 2, 2008 | |
| Proprietary<br>Name: | Modification to AUDICOR 200 | |
| Common/ Usual<br>Name: | Electrocardiograph/Acoustic Cardiograph | |
| Classification: | 870.2340, DPS class II, Electrocardiograph<br>870.1875, DQD, class II, Electronic Stethoscope<br>870.2800, MLO, class II, Electrocardiograph, Ambulatory (with analysis program) | |
| Performance<br>Standards: | AAMI EC 11 | |
| Intended Use: | The Audicor® System, when used with AUDICOR Sensors on the chest wall, is<br>intended for use in acquiring, analyzing and reporting ECG and heart sound data<br>and to provide interpretation of the data in an integrated ACG (acoustic<br>cardiograph) report for consideration by physicians. Audicor systems allow<br>detection, reporting and interpretation of standard ECG data as well as advanced<br>parameters such as EMAT, LVST, S3 strength, S4 strength, and SDI. | |
| | Data may be reported in a single snapshot report and in a trended report format<br>where multiple data points are trended over time. The Audicor system also accepts<br>and analyzes downloads of up to 48 hours of patient data from an optional<br>ambulatory ECG/heart sounds collection device. | |
| | ECG and heart sound data offered by the device are only significant when used in<br>conjunction with physician over read as well as consideration of other relevant<br>patient data. | |
| | The device is intended for use only under the direct supervision of a physician and<br>is for use on adults (≥ 18 years). | |
| Device<br>Description: | The Audicor 200 is a stand-alone device that can be used to capture 10-second<br>snapshots of ECG and heart sounds in patients suspected of heart failure or acute<br>coronary syndrome. | |
| | The Audicor TS system, a combination of the Audicor 200 with an Audicor-enabled<br>laptop computer, is used to display and analyze patient data over time in a trended<br>format. The trending system can be used in monitoring patient changes during<br>therapeutic treatment or during CRT studies. | |
| | The Audicor system, when used with either a 10 wire or 4 wire patient cable<br>analyzes and reports the following advanced parameters and a systolic<br>dysfunction index (SDI) calculated from advanced parameters: | |
| | LVST ( Left Ventricular Systolic Time )<br>LVDT (Left Ventricular Diastolic Time) | |
{1}------------------------------------------------
- PADT (Pre-atrial Diastolic Filling Time) �
- AAFT (Accelerated Atrial Filling Time ) .
- EMAT (QS1, Electromechanical Activation Time) .
- QS2 .
- R-R Interval .
- S3 Strength .
- S4 Strenth �
- SDI (Systolic Dysfunction Index)
The Audicor-enabled laptop computer also accepts, analyzes and displays trended data from an optional ambulatory data collection harness for trend and event analysis.
Test Summary & The Audicor System has been tested to the applicable requirements of the Conclusion: following standards, and shown to comply.
- . ANSI/AAMI EC-11 Standard for Diagnostic Electrocardiographic Devices
- EN 60601-1 (UL 60601-1) Standard for Medical Electrical Equipment: General . Requirements for Safety
- . EN 60601-1-2 Standard for Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
- IEC 60601-2-25 Medical Electrical Equipment Part 2-25: Particular . Requirements for the Safety of Electrocardiographs
- IEC 606601-2-51 Medical Electrical Equipment Part 2. Particular . Requirements for Safety, Including Essential Performance, of Recording and Analyzing Single Channel and Multi-channel Electrocardiographs
The optional ambulatory ECG/heart sounds data collection harness (Hemo) is designed to comply with
ANSI/AAMI EC38:1998 Ambulatory Electrocardiographs .
Substantial Equivalence:
.
The added Hemo accessory wireless ambulatory ECG/heart sounds data collection harness is found to be substantially equivalent to the following cleared devices:
- GE Medical Systems SEER Light Holter Recorder (K021470)
The Audicor System Hemo wireless ECG/heart sounds data collection harness Technological and the predicate GE SEER Light Holter Recorder are technologically equivalent in Characteristics: that both devices:
- Are small, wearable, battery-powered devices .
- Acquire ECG data from patients using limited ECG leads .
- Use non-volatile memory to store patient data for download to the analysis . module of the system
The Audicor Ambulatory system is different from the predicate in the following ways:
- It requires use of proprietary Audicor sensors to capture ECG and heart . sounds data
- The GE SEER Light does not perform the analysis on recorded data while the . Audicor system does.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## SEP 0 9 2008
Inovise Medical, Inc. c/o Ms. Kendra Rathkey Manager, Quality and Regulatory 10565 SW Nimbus Avenue Suite 100 Portland, OR 97223-4311
Re: K073545
Trade Name: AUDICOR Hemo Ambulatory Monitor Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II (two) Product Code: MWJ Dated: August 6, 2008 Received: August 11, 2008
Dear Ms. Rathkey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Kendra Rathkey
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Frank Lacy at (240) 276-4095.
Sincerely yours,
er R. V. Annes
A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K073545
Device Name: Audicor Hemo Amonalastory,
Indications For Use:
The Audicor® System, when used with AUDICOR Sensors on the chest wall, is intended for use in acquiring, analyzing and reporting ECG and heart sound data and to provide interpretation of the data in an integrated ACG (acoustic cardiograph) report for consideration by physicians. Audicor systems allow detection, reporting and interpretation of standard ECG data as well as advanced parameters such as EMAT, LVST, S3 strength, S4 strength, and SDI.
Data may be reported in a single snapshot report and in a trended report format where multiple data points are trended over time. The Audicor system also accepts and analyzes downloads of up to 48 hours of patient data from an optional ambulatory ECG/heart sounds collection device,
ECG and heart sound data offered by the device are only significant when used in conjunction with physician over read as well as consideration of other relevant patient data.
The device is intended for use only under the direct supervision of a physician and is for use on adults (≥ 18 years).
Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Duna D. Schunes
(Division Sian-Off Division of Cardiovascular Devices
510(k) Number_K0+5545
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.