← Product Code [KTT](/productcode/KTT) · K073527

# VARIAX ELBOW SYSTEM (K073527)

_Howmedica Osteonics Corp. · KTT · Feb 14, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K073527

## Device Facts

- **Applicant:** Howmedica Osteonics Corp.
- **Product Code:** [KTT](/productcode/KTT.md)
- **Decision Date:** Feb 14, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The VariAx™ Elbow System is intended for fracture fixation of long bones. Indications include: Distal Humerus Proximal Ulna .

## Device Story

VariAx Elbow System is a line extension of the Stryker Plating System; consists of metallic bone plates and screws for fracture fixation. Used by surgeons in clinical settings for stabilization of distal humerus and proximal ulna fractures. Operates as a mechanical fixation system; provides structural support to bone segments to facilitate healing. No electronic, software, or algorithmic components involved.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Metallic bone fixation appliances (plates and screws). Design based on existing Stryker plating systems. Mechanical fixation principle. No software, energy source, or connectivity.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Stryker Foot ([K063875](/device/K063875.md))
- Profyle ([K062498](/device/K062498.md))
- Stryker Locked Plating ([K050512](/device/K050512.md))
- Universal Distal Radius Systems ([K040022](/device/K040022.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

073527 #1/2

| VariAx       TM Elbow System   |                                                                                                                                                                    |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name:              | VariAxTM Elbow System                                                                                                                                              |
| Common Name:                   | Bone plates and screws                                                                                                                                             |
| Classification Name/Reference: | Single/multiple component metallic bone fixation<br>appliances and accessories, 21 CFR §888.3030                                                                   |
| Device Product Code:           | 87 KTT                                                                                                                                                             |
| Proposed Regulatory Class:     | Class II                                                                                                                                                           |
| For Information contact:       | Francisco Haro, Regulatory Affairs Specialist<br>Howmedica Osteonics Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-5493 Fax: (201) 831-6038 |
| Date Summary Prepared:         | December 14, 2007                                                                                                                                                  |

# 510(k) Summary of Safety and Effectiveness

## Description:

This submission is a line extension to the Stryker® Plating System. The components of the VariAx™ Elbow System are intended to add different types of plates and screws to the Stryker® Plating System portfolio.

#### Indications:

The VariAx™ Elbow System is intended for fracture fixation of long bones. Indications include:

- . Distal Humerus
- . Proximal Ulna

Substantial Equivalence:

The VariAx™ Elbow System is substantially equivalent to the Stryker® Foot, Profyle® Stryker® Locked Plating, and Universal Distal Radius Systems in regards to intended use, design, materials, and operational principles as a fracture plating system.

{1}------------------------------------------------

073527 #212

# Executive Summary:

# Device Description:

This submission is a line extension to the Stryker® Plating System. The components of the VariAx™ Elbow System will be based on the component designs of the Stryker® Foot, Profyle®, Stryker® Locked Plating, and Universal Distal Radius Systems which were determined substantially equivalent in the following 510(k)s K063875, K062498. K050512, and K040022. Refer to Table 1 in Section 11 for a description of each predicate component, subject component, and modifications.

## Indications:

The VariAx™ Elbow System is intended for fracture fixation of long bones. Indications include:

- . Distal Humerus
- Proximal Ulna .

Substantial Equivalence:

The VariAx™ Elbow System is substantially equivalent to the Stryker® Foot, Profyle® Stryker® Locked Plating, and Universal Distal Radius Systems in regards to intended use, design, materials, and operational principles as a fracture plating system.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The logo is black and white.

#### Public Health Service

FEB 1 4 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Howmedica Osteonics Corp. % Mr. Francisco Haro 325 Corporate Drive Mahwah, NJ 07430

> K073527 Trade/Device Name: VariAx™ Elbow System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: December 14, 2007 Received: December 17, 2007

Dear Mr. Haro:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

## Page 2 - Francisco Haro

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): _

Device Name: VariAx™ Elbow System

Indications for Use:

The VariAx™ Elbow System is intended for fracture fixation of long bones. Indications include:

- Distal Humerus ●
- Proximal Ulna .

Prescription Use Over-The-Counter Use x AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) 

������������������������������������������������������������������������������������������������������������������������������������������������������������������������������ Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Yaubare Buemut
Division Sign

Division of General, Resporative, and Neurological D. ... es

**510(k)** K073527

---

**Source:** [https://fda-staging.innolitics.com/device/K073527](https://fda-staging.innolitics.com/device/K073527)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com