K073456 · Medx, Inc. · IYX · Dec 20, 2007 · Radiology
Device Facts
Record ID
K073456
Device Name
T-QUEST SYSTEM
Applicant
Medx, Inc.
Product Code
IYX · Radiology
Decision Date
Dec 20, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1100
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The intended uses of the T-Quest™ System are the acquisition, processing, display and analysis of planar images of the thyroid gland and other small organs using approved pharmaceuticals.
Device Story
T-Quest System is a mobile scintillation (gamma) camera for planar imaging of thyroid and small organs. Input: gamma rays from radioisotopes (e.g., Technetium 99m, Iodine 123) administered to patients. Operation: collimator focuses gamma rays onto a scintillation crystal; crystal converts rays to light; photomultiplier tubes (19) convert light to electrical signals. System includes acquisition board, notebook PC, and NuQuest (IM512P) software. Output: 2D planar images displayed on PC screen. Used in clinical settings by healthcare professionals. Software processes, analyzes, and documents isotope studies. Physician interprets output for clinical decision-making regarding organ size and isotope uptake. Benefits: diagnostic information for thyroid and small organ function.
Clinical Evidence
Bench testing only. Performance characteristics (linearity, resolution, uniformity) and thyroid gland phantom images were evaluated. Results deemed excellent, demonstrating performance equivalent to similar devices in the marketplace.
Technological Characteristics
Mobile scintillation camera; 19 photomultiplier tubes; collimator-based gamma ray detection. External notebook PC interface; Windows XP OS. Connectivity: DICOM data transfer. Storage: CD/DVD. Standards: UL 60601-1, CAN/CSA C22.2 No. 601-1, EMI standards, NEMA NU 1-2001 (uniformity, linearity, resolution). Software: NuQuest (IM512P) for acquisition/processing.
Indications for Use
Indicated for acquisition, processing, display, and analysis of planar images of the thyroid gland and other small organs in patients using approved pharmaceuticals.
Regulatory Classification
Identification
A scintillation (gamma) camera is a device intended to image the distribution of radionuclides in the body by means of a photon radiation detector. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
Inter Medical Gamma Camera System Model CX 250 C Plus (K052473)
{0}------------------------------------------------
# 173456
### 510(k) SUMMARY
DEC 2 0 2007
Submitted by: MEDX, Incorporated 3456 N. Ridge Ave., #100 Arlington Heights, IL 60004 Telephone: (847)-463-2020 FAX: (847)-463-2019 Contact Person Floyd R. Rowan, Executive Vice President Date Summary Prepared: November 9, 2007
Trade Name of Device
Common Name
Classification Name:
Classification:
Substantial Equivalence:
T-Quest™ System
Scintillation (Gamma) Camera System.
Scintillation (Gamma) Camera
This Device is Classified as a Class I Device. Ref. 21 CFR 892.1100. Product Code IYX.
The T-Quest™ System is Substantially Equivalent to the Inter Medical Gamma Camera System Model CX 250 C Plus. 510(k) Number K052473.
Description of Device: The T-Quest System consists of a small mobile scintillation camera, accessories such as optional phantoms, and an optional patient bed. This product includes a notebook computer, an acquisition board, and IM512P Data and Image processor software marketed by MEDX, Inc. under the trade name of NuQuest™ Software. This system is primarily used for the acquisition, processing, and analysis of planar images of the thyroid gland and other small organs.
{1}------------------------------------------------
Scientific Concepts: Diagnostic Nuclear Medicine began in early 1950's with the availability of short half-life radioisotopes. Isotopes such I 131 were administered to the patient and were selectively taken up by organ systems such as the thyroid. Measurement of the resulting radioactivity in the organ provided information on both the size of the organ and the relative amount of the isotope taken up.
Nuclear Medicine cameras work on the principle similar to television cameras. A collimator (lens) "focuses" gamma ravs on a scintillation crystal. The scintillation crystal converts the gamma rays into light. Photomultiplier tubes are then used to convert the light into an electrical signal proportional to the energy of the detected gamma ray. Early instruments used a single hole lead collimator and detector moved in a raster pattern forming a 2-D image of the organ of interest. In the late 1950's methods were developed for directly obtaining a 2-D image by using a large crystal with multiple photomultiplier tubes and electronically calculating the position and energy of the gamma event. As computers became available the resulting images were portrayed on computer screens.
Indication for Use: The intended uses of the T-Quest™ System are the acquisition, processing, display and analysis of planar images of the thyroid gland and other small organs using approved pharmaceuticals.
Note: The predicate device in its indication for use does not include the word "analysis", this refers to the analysis of energy levels. It the physician who interprets the data and makes the clinical decisions, not the gamma camera system.
| Safety Information: | The T-Quest™ System<br>Meets UL 60601-1 and CAN/CSA<br>C22.2 No. 601-1 Standards and<br>is In Compliance with EMI<br>Standards. |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Performance Standard: | This Device Meets NEMA NU 1-<br>2001 Standard for Uniformity,<br>Linearity, and Resolution. |
Non-Clinical Data: The T-Quest™ System performance characteristics with respect to Linearity, Resolution, Uniformity, and of the thyroid gland phantom
{2}------------------------------------------------
images were deemed to be excellent. The conclusion drawn from this data is that the T-Quest™ System provides effective results as expected from similar devices in the marketplace..
Device Technological Characteristics: The characteristics of the T-Quest™ Scintillation Camera when used with the NuQuest™ computer and NuQuest™ (IM512P) Software compare substantially with the predicate device, in both materials used, technology applied, and functional methodology. Differences of note do not affect safety and effectiveness of the device, intended use, or application methods. The device operates in a manner substantially equivalent to other cleared devices in this category, and performs as well as the predicate device.
| | T-Quest System | Inter Medical CX250C<br>510(k); K052473 |
|----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The intended uses are<br>the acquisition, proces-<br>sing, display and analysis<br>of planar images of the<br>thyroid gland and other<br>small organs using<br>approved pharmaceuti-<br>cals. | The CX250Plus Gamma<br>Camera is a mobile<br>system for acquisition<br>and processing nuclear<br>medicine diagnostic.<br>Note: this camera also<br>displays the planar<br>images and requires the<br>use of approved<br>pharmaceuticals. |
| Compact Mobile<br>Gamma camera for<br>acquisition, processing<br>representation and<br>documentation of<br>patient isotope<br>studies. | Yes, but only for small<br>organs such as the<br>thyroid. | Yes, for both small and<br>large organs. |
| Thyroid Collimator | Yes | Yes |
| Iodine 131 Studies | No | Yes |
| Iodine 123 Studies | Yes | Yes |
| Technetium 99m<br>Studies | Yes | Yes |
| DICOM Data Transfer<br>Capability | Yes | Yes |
| Adjustable Head | Yes | Yes |
| Pinhole Collimator | Yes | Yes |
| General Purpose | Yes | Yes |
| Collimator | | |
| Computer | External Notebook PC<br>Providing the Keyboard<br>and Viewing Screen. | Built-in Computer with<br>external Keyboard and<br>Viewing Screen. |
| Operating System | Windows XP | Windows XP |
| Display | Color or Grayscale | Color or Grayscale |
| Number of<br>Photomultiplier Tubes | 19 | 19 or 37 |
| Corrections | Linearity, Energy, and<br>Uniformity | SAME |
| Acquisition Matrixes | 64x64, 128x128,<br>256x256, 512x512, and<br>1024x1024. | 64x64, 128x12, 256x256.<br>and 512x512 |
| Frame Mode | Static, Dynamic, and<br>Gated. | Static, Dynamic, and<br>Gated. |
| Patient Archive | CD/DVD | CD/DVD, PLUS Magneto<br>Optical Drive is Available |
| Acquisition Board | Nuquest™ Acquisition<br>Board is Located Below<br>the Shelf that Holds the<br>Notebook PC. | Proprietary electronics<br>performing the same<br>function. |
| Software | Uses previously 510(k)<br>Cleared Software Called<br>IM512P which is<br>Marketed by MEDX, INC<br>under the trade name of<br>NuQuest™ Software.<br>This software was<br>cleared in the following<br>510(k)'s for FDA Code<br>IYX Products: K945792<br>and K961104 | Gamma XP Acquisition<br>and Processing Software.<br>Its previous history is<br>unknown to us. |
## COMPARISON OF T-Quest™ SYSTEM TO PREDICATE DEVICE
{3}------------------------------------------------
#### EXPLANATIONS PERTAINING TO 510(K) NUMBERS PROVIDED IN THE COMPARISSON TABLE ABOVE. s
K052473 is the FDA cleared 510(k) for the model CX250C Plus Gamma Camera System manufactured by Inter Medical. This is the predicate device to which we are comparing our T-Quest System.
{4}------------------------------------------------
K945792 is the FDA cleared 510(k) for IM512P software filed by Globus Industries which is the U.S. Agent for Alfanuclear S.A. I. y C. in Buenos Aires Argentina which designs and markets the IM512P software.
K961104 is the FDA cleared 510(k) for Notebook Imager a product of Digirad Corp. This device uses IM512P software cleared Sitco 510(k) K945792 which is now owned by MEDX, Inc.
The purpose of discussing the 510(k)'s (other than K052473) is to demonstrate that the Alfanuclear IM512P software has been in use for over 12 years, that it has been cleared by the FDA and that is not a new technology for acquisition and imaging of scintillation (gamma) camera images .
DISCUSSION OF THE DIFFERENCES LISTED IN THE COMPARISSON OF T-QUEST SYSTEM TO THE INTER MEDICAL PREDICATE DEVICE.
The T-Quest System is a smaller gamma camera system as is observed that it uses only 19 Photomultiplier tubes (PMT)s whereas the Inter Medical Device uses both 19 and 37 PMTs. For thyroid and other small organ imaging the 19 PMT detector heads are used by us and the predicate device.
As for acquisition matrixes, which define the resolution of the images, the 1024x1024 mode can be used but it is impractical for thyroid imaging because it takes a longer time to acquire the image and requires large amount of storage memory.
Patient Archives - we do not view the use of Magneto Optical Storage Devices as a detraction from our system which does not use them. The CD and DVD data storage methods are widely used and preferred by users of gamma camera systems
lodine 131 Studies the T-Quest System would require a greater level of lead shielding to handle this isotope. The Technetium 99m and lodine 123 provide very usable thyroid images.
Overall both systems are substantially equivalent in the acquisition, processing, display and analysis of planar images of the thyroid and other small organs using approved pharmaceuticals.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its head, representing the department's mission to promote health and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 2 0 2007
MEDX, Inc. % Mr. Jeff D. Rongero Senior Project Engineer, Medical Business Unit Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle Park, NC 27709
Re: K073456
Trade/Device Name: T-Quest™ System Regulation Number: 21 CFR 892.1100 Regulation Name: Scintillation (gamma) camera Regulatory Class: I Product Code: IYX Dated: November 30, 2007 Received: December 10, 2007
#### Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known):
073456/
Device Name: T-Quest™ System
Indications For Use:
The intended uses of the T-Quest™ System are the acquisition, processing, display and analysis of planar images of the thyroid gland and other small organs using approved pharmaceuticals.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arquihi Whay
(Division Sign-Off)
് Malon of Reproductive, Abdominal and Adjological Devices 510(k) Number __
Page 1 of _ l _ _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.