BIPORE ACCUFLEX PERCUTANEOUS SHEATH INTRODUCER

K073434 · Bipore, Inc. · DYB · Apr 1, 2008 · Cardiovascular

Device Facts

Record IDK073434
Device NameBIPORE ACCUFLEX PERCUTANEOUS SHEATH INTRODUCER
ApplicantBipore, Inc.
Product CodeDYB · Cardiovascular
Decision DateApr 1, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The Bipore Accuflex™ Percutaneous Sheath Introducer is recommended for percutaneous introduction of various devices into arteries and/or veins for diagnostic and therapeutic procedures.

Device Story

Bipore Accuflex™ Percutaneous Sheath Introducer is a flexible, kink-resistant tube used for percutaneous access to arteries and veins. Device facilitates introduction of diagnostic or therapeutic instruments. Construction includes nylon-based, fluoropolymer-lined material reinforced with stainless steel wire; proximal end features 24-karat gold radiopaque marker, integral side port/hemostasis valve (silicone rubber), and female luer connection for Touhy Borst adapter. Available in 5-8 French sizes and 5-110 cm lengths; some models include pre-bend (45° to 180°) for steerability via distal torque transfer. Side port allows fluid/contrast media administration. Used by clinicians in clinical settings to provide a conduit for catheters or dilators. Benefits include improved procedural access and kink resistance during navigation of patient anatomy.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Materials: nylon-based, fluoropolymer-lined tube, stainless steel wire reinforcement, non-DEHP PVC side port, silicone rubber hemostasis valve, 24k gold radiopaque marker. Dimensions: 5-8 French, 5-110 cm lengths. Features: kink-resistant coil, optional 45°-180° pre-bend for steerability. Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring percutaneous access to arteries or veins for diagnostic or therapeutic procedures.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) Summary K073434 #### Submitted bv: Keith Paluch Director of Technical and Regulatory Affairs Bipore Medical Devices, Inc. 31 Industrial Parkway Northvale, NJ 07647 #### Date Prepared: March 27, 2008 #### Proposed Device: Bipore Accuflex™ Percutaneous Sheath Introducer #### Product Classification: Class II C.F.R. 870.1340 Product code: DYB, Catheter Introducer #### FDA registration number: 2248069 #### Predicate Device: Bipore Accuflex™ Percutaneous Sheath Introducer K051513 #### Indications for Use: The Bipore Accuflex™ Percutaneous Sheath Introducer is recommended for percutaneous introduction of various devices into arteries and/or veins for diagnostic and therapeutic procedures. #### Device Description: The Bipore Accuflex™ Percutaneous Sheath Introducer is a flexible tube designed for the percutaneous introduction of various devices into arteries and/or veins for a variety of diagnostic and therapeutic procedures. The product consists of a tube fabricated out of nylon based, flouropolymer lined material, reinforced by a flexible stainless steel wire within the tube. Proximally located on the flexible tube is an integral side port/hemostasis valve and/or female luer connection for attachment to a Touhy Borst adaptor. The side port, which provides a means of introducing contrast media or other fluids, is fabricated from non-DEHP PVC. The hemostasis valve is made from silicone rubber. The Touhy Borst adapter is a separate or connected component purchased from B.Braun. This purchased component is used in a manner consistent with B.Braun specification and premarket clearance. The proximal end contains a 24 karat gold radiopaque marker. The stainless steel coil enables the introducer to kink resistant during normal use. The proximal end may also contain a 45° to 180° radius bend. This pre-bend allows the product to be directed by transferring distal torque to steer the proximal tip. The reinforced tube, through which a vessel dilator or catheter is introduced is available in 5,6,7 and 8 French sizes and in lengths ranging from 5 cm to 110 cm. {1}------------------------------------------------ ## Summary of Technical Characteristics as compared to the predicate device: The proposed change modifies the exterior dimensions of hemostasis valve body by and adds additional size offerings to include lengths up to 110cm. The device may also contain an additional Polyethylene dilator for user convenience. The proposed modification to the hemostasis valve improves the manufacturability of the device. This modification makes the device easier to handle by the user because of the reduction in bulk resulting by a change in length. The slight increase in overall length to 110cm allows the device to be used with devices that require longer length sheaths. The additional Polyethylene dilator improves the convenient use of the product when the user prefers a shorter or longer length dilator to accommodate patient anatomy. Test results indicate that the modified product is in conformance with Bipore specifications and is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR - 1 2008 Bipore, Inc. c/o Mr. Keith Paluch Director of Technical and Regulatory Affairs 31 Industrial Parkway Northvale, NJ 07647 Re: K073434 Bipore Accuflex™ Percutaneous Sheath Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II (two) Product Code: DYB Dated: March 17, 2008 Received: March 17, 2008 Dear Mr. Paluch: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 – Mr. Keith Paluch Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincercly vours. er R. Kirchner Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATION FOR USE 510(k) Number (if known): K073434 Device Name: Bipore Accuflex™ Percutaneous Sheath Introducer Intended Use / Indication For Use: The Bipore Accuflex™ Percutaneous Sheath Introducer is recommended for percutaneous introduction of various devices into arteries and/or veins for diagnostic and therapeutic procedures. Over-The-Counter Use Prescription Use ਮ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sunya R. Ladner (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number k073434 Confidential .
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...