← Product Code [GEX](/productcode/GEX) · K073332

# V-RASER DIODE LASER SYSTEM (K073332)

_Hoya Conbio, Inc. · GEX · Feb 14, 2008 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K073332

## Device Facts

- **Applicant:** Hoya Conbio, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Feb 14, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Incision, excision, ablation, vaporization and coagulation of soft tissue. Indicated for the treatment of vascular lesions.

## Device Story

V-RASER Diode Laser System is a modified Diodent Micro 980; utilizes solid-state Gallium Aluminum Arsenide (GaAlAs) semiconductor diode to produce 980nm invisible laser energy. System includes fiber optic beam delivery, removable flexible HCS (hard clad silica) optical treatment fiber, and lightweight handpiece with interchangeable tips for spot size adjustment. Operator controls laser via key switch, Ready Mode, and footswitch; color LCD touch-screen allows adjustment of output levels. Software calculates treatment fluence based on selected spot size. Device operates in pulsed mode. Used by clinicians for soft tissue procedures; provides precise tissue interaction for incision, excision, ablation, vaporization, and coagulation. Benefits include targeted treatment of vascular lesions.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Solid-state GaAlAs semiconductor diode laser; 980nm wavelength; pulsed mode operation. Delivery via HCS optical fiber and handpiece with interchangeable tips. Control via LCD touch-screen interface. Electrical requirements and physical specifications similar to predicate Diodent Micro 980.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Diodent Micro 980 ([K063384](/device/K063384.md))
- Iridex Varilite ([K041930](/device/K041930.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K073332

pg 1 of 2

## 510(k) Summary or 510(k) Statement 5.

V-RASER Diode Laser System 510k Summary

Image /page/0/Picture/4 description: The image shows the logo for HOYA ConBio. The text "HOYA" is in large, bold, black letters, followed by "ConBio" in a smaller, thinner font. There is a horizontal line underneath the text. The logo appears to be for a company or organization named HOYA ConBio.

47733 Fremont Boulevard, Fremont, California 945 510-445-4500 www.conbio.com 510-445-4550 Fax

| Submitter:             | HOYA ConBio, Inc.<br>47733 Fremont Blvd.<br>Fremont, California 94538<br>Phone: 510-445-4500<br>Fax: 510-445-4550                                                                                                                                                                                                                                                                                                                                                                                                                            |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:               | Mr. Jim Green<br>Vice President of Engineering                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Date Summary Prepared: | November 21, 2007                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Trade Name:     | V-RASER Diode Laser System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common Name:           | Dermatology Laser System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Classification Name:   | Instrument, surgical, powered, laser<br>79-GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification Code:   | 878.4810 Laser surgical instrument for use in general and<br>plastic surgery and in dermatology (1) A carbon dioxide<br>laser for use in general surgery and in dermatology is a<br>laser device intended to cut, destroy, or remove tissue by<br>light energy emitted by carbon dioxide.<br>(2) An argon laser for use in dermatology is a laser device<br>intended to destroy or coagulate tissue by light energy<br>emitted by argon.                                                                                                     |
| Equivalent Device:     | Diodent Micro 980 (K063384), Iridex Varilite (K041930),                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Description:    | The V-RASER Diode System is a modified Diodent Micro<br>980. It will use similar specifications (laser medium,<br>excitation method, fiber optic beam delivery, electrical<br>requirements, physical specifications) as the Diodent Micro<br>980. The laser source of this device is a solid-state Gallium<br>Aluminum Arsenide (GaAlAs) semiconductor diode. It<br>produces invisible laser energy at the 980-nanometer<br>wavelength. The delivery system consists of removable<br>flexible HCS (hard clad silica) optical treatment fiber |

1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1

{1}------------------------------------------------

K0733332

|                                  | assembly connected to a lightweight, hand piece that has<br>two different size end pieces (tips) to adjust the laser beam<br>to specific spot sizes. Activation occurs when the operator<br>enables the laser by a key switch, enters the Ready Mode<br>and presses the footswitch. Releasing the footswitch<br>suspends laser treatment.  A color LCD and touch-screen<br>display panel allows the operator to adjust or set the laser<br>output level. The laser  operates in a pulsed mode and will<br>have a removable stainless steel end piece for establishing<br>the proper focal length for the treatment. The hand piece<br>will be manually attached (i.e., screwed on) to a connector<br>to the fiber assembly. The spot size will be selected on the<br>control panel and the software will calculate the treatment<br>fluence accordingly. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use:                    | Incision, excision, ablation, vaporization and coagulation of<br>soft tissue.  Indicated for the treatment of vascular lesions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Comparison:                      | The V-RASER Diode Laser System is comparable to its<br>predicate and parent device in terms of its indications for<br>use, technical specifications, operating performance<br>features, and general design features.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Nonclinical Performance<br>Data: | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Clinical Performance<br>Data:    | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Additional Information:          | None requested at this time.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or other bird-like figure, composed of three curved lines.

FEB 1 4 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

HOYA ConBio, Inc. % Liza Burns and Associates Ms. Liza Burns Regulatory/Clinical Consultant 19722 Westview Drive Twain Harte, California 95383

Re: K073332

Trade/Device Name: V-RASER Diode Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: November 21, 2007 Received: November 21, 2007

Dear Ms. Burns:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Liza Burns

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark N. Mulkerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K073332 pg 1 of 1

## Indications for Use

510(k) Number (if known):

V-RASER Diode Laser System Device Name:

Intended Use:The V-Raser is intended for Incision, excision, ablation, vaporization and coagulation of soft tissue.

Specific Indications: Treatment of vascular lesions.

Prescription Use Over-The-Counter Use × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Philip Van Off

Page __ of ___________________________________________________________________________________________________________________________________________________________________

(Division Sign-Off Division of General, Restorative, and Neurological Devices

510(k) Number L013352

12

---

**Source:** [https://fda-staging.innolitics.com/device/K073332](https://fda-staging.innolitics.com/device/K073332)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).