← Product Code [LLZ](/productcode/LLZ) · K073306

# XORANCONNECT (K073306)

_Xoran Technologies, Inc. · LLZ · Jan 7, 2008 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K073306

## Device Facts

- **Applicant:** Xoran Technologies, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jan 7, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

XoranConnect™ is a web-based service designed to complement point-of-care medical imaging systems by providing point-of-expertise tools for viewing, interpretation, clinical review, analysis, distribution, processing and printing. Primary components include online multiplanar image viewing, offsite data storage, administrative reporting tools, and teleradiology workflow. It is intended as an extension to (but not limited to) the MiniCAT™ and xCAT™ CT imaging devices. XoranConnect is a medical imaging software system intended to display, edit, review, store, print and distribute images acquired from imaging devices such as Computed Tomography (CT), Magnetic Resonance (MR), Computed Radiography (CR), Ultrasound (US), Nuclear Medicine (NM), and other devices.

## Device Story

Web-based PACS service; complements point-of-care imaging systems (e.g., MiniCAT™, xCAT™). Inputs: medical images from CT, MR, CR, US, NM modalities. Architecture: XoranVault™ (storage servers, RAID5) and XoranConnect™ (study server, web server, web client). Operation: imaging modality transfers study to storage server; user accesses via web browser. Features: off-site data storage; 3D/multiplanar visualization; teleradiology workflow; administrative reporting. Used in clinical settings; operated by healthcare professionals. Output: visualized/manipulated images for clinical review, interpretation, and distribution. Benefits: enables remote point-of-expertise viewing and interpretation of imaging studies.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Web-based software system; runs on standard off-the-shelf PC and networking hardware. Architecture includes storage servers (RAID5) and web-based viewing client. Connectivity requires active Internet connection. Software-based image processing for 3D and multiplanar visualization.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Visage PACS/CS (Mercury Computer Systems, Inc.) ([K072205](/device/K072205.md))
- NovaPACS (NovaRad Corporation) ([K061920](/device/K061920.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Xoran Technologies. The logo features a stylized "X" in a box-like shape, followed by the word "ORAN" in bold, sans-serif font. Below "ORAN" is the word "TECHNOLOGIES" in a smaller, lighter font.

K073306
pg 1 of 2

XoranConnect™ 510(k) Premarket Notification

# 9 510(k) Summary

#### 9.1 Name and Address of Manufacturer

Xoran Technologies, Inc.

309 North First Street Ann Arbor, MI 48103

Phone: (734) 663 7194 (734) 663 8500 Fax:

### 9.2 Contact Information

Amy Kim Regulatory Affairs Manager

(734) 663 7194

Date of Submission: November 15, 2007

## 9.3 Product Identification

| Trade Name:        | XoranConnect™                                                                                                  |
|--------------------|----------------------------------------------------------------------------------------------------------------|
| Common Name:       | PACS                                                                                                           |
| Classification:    | Class II, Product code LLZ (Image Processing System - PACS, Sec.<br>892.2050).                                 |
| Guidance Document: | Guidance for the Submission Of Premarket Notifications for Medical Image<br>Management Devices (July 27, 2000) |

## 9.4 Substantially Equivalent (Predicate) Devices

XoranConnect™ software is substantially equivalent to the following predicate devices:

Visage PACS/CS (Mercury Computer Systems, Inc.), K072205

NovaPACS (NovaRad Corporation), K061920

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Xoran Technologies. The logo features a stylized "X" with the word "XORAN" in bold, sans-serif font to the right of the "X". Below "XORAN" is the word "TECHNOLOGIES" in a smaller, sans-serif font. The logo has a simple, modern design.

XoranConnect™ 510(k) Premarket Notification

# 9.5 Device Description

XoranConnect™ is a web-based service designed to complement point-of-care medical imaging systems by providing point-of-expertise viewing and interpretation tools. XoranConnect™ runs on standard offthe-shelf PC and networking hardware, all of which has been manufactured by third party reputable manufacturers.

The system performs these major functions.

- Off-site data storage allows user studies to be stored off-site. .
- Study viewer allows studies to be visualized and manipulated in 3D, multiplanar and other forms. .

To accommodate the major functions, the overall system architecture includes the following distinct components:

- XoranVault™ storage system (storage servers, dispatcher) .
- XoranConnect™ viewing system (study server, web server, web client) .

The storage (backup) system consists of one or more servers designed to store and serve stored imaging content on a RAID5 storage device. Imaging modality negotiates the connection and transfers study to the storage server.

Once stored, imaging data is ready to be served to the requesting clients. When a user logs into the system by visiting a XoranConnect web page, study server begins to serve the data, which in turn is displayed in the user's web browser window. Any storage server may simultaneously serve multiple requests for images.

It is the user's responsibility to ensure that the appropriate PC hardware and display device are used to run the XoranConnect™ software application. XoranConnect™ requires at all times an active Internet connection to operate. It is the user's responsibility to ensure that the connection that is being used provides the speed, stability and reliability required to utilize XoranConnect™ for any given purpose or application.

#### 9.6 Intended Use

XoranConnect™ is a web-based service designed to complement point-of-care medical imaging systems by providing point-of-expertise tools for viewing, interpretation, clinical review, analysis, distribution, processing and printing. Primary components include online multiplanar image viewing, offsite data storage, administrative reporting tools, and teleradiology workflow. It is intended as an extension to (but not limited to) the MiniCAT™ and xCAT™ CT imaging devices.

#### 9.7 Comparison With the Predicate Devices

XoranConnect™ has the same indication for use as the predicate devices and they perform very similar functions. Apart from differences in menus, usability, specific tool implementations, appearance and price, all of XoranConnect™ functions can be found in at least one of the predicate devices (most likely both). All three devices perform similar functions, while the predicate devices have many additional features.

#### 9.8 Conclusion

Functionality that XoranConnect™ provides is found on both predicate devices, and is therefore substantially similar to the predicate devices Visage PACS/CS (Mercury Computer Systems, Inc.), K072205, and NovaPACS (NovaRad Corporation), K061920. This design complies with 21 CFR 892.2050 and does not pose new safety risks.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing-like shapes above a body with two curved lines representing legs or a base.

JAN - 7 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Amy S. Kim, DDS Regulatory Affairs Manager Xoran Technologies, Inc. 309 N First Street ANN ARBOR MI 48103

Re: K073306

Trade/Device Name: XoranConnect™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 21, 2007 Received: November 30, 2007

Dear Dr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

K07-3306

XoranConnect™ Device Name:

Indications for Use:

XoranConnect is a medical imaging software system intended to display, edit, review, store, print and distribute images acquired from imaging devices such as Computed Tomography (CT), Magnetic Resonance (MR), Computed Radiography (CR), Ultrasound (US), Nuclear Medicine (NM), and other devices.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Vorquimthn

(División Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 073306 510(k) Number

Page 1 of 1

Division Sign-Off

510(k)

---

**Source:** [https://fda-staging.innolitics.com/device/K073306](https://fda-staging.innolitics.com/device/K073306)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
