K073306 · Xoran Technologies, Inc. · LLZ · Jan 7, 2008 · Radiology
Device Facts
Record ID
K073306
Device Name
XORANCONNECT
Applicant
Xoran Technologies, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 7, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
XoranConnect™ is a web-based service designed to complement point-of-care medical imaging systems by providing point-of-expertise tools for viewing, interpretation, clinical review, analysis, distribution, processing and printing. Primary components include online multiplanar image viewing, offsite data storage, administrative reporting tools, and teleradiology workflow. It is intended as an extension to (but not limited to) the MiniCAT™ and xCAT™ CT imaging devices. XoranConnect is a medical imaging software system intended to display, edit, review, store, print and distribute images acquired from imaging devices such as Computed Tomography (CT), Magnetic Resonance (MR), Computed Radiography (CR), Ultrasound (US), Nuclear Medicine (NM), and other devices.
Device Story
Web-based PACS service; complements point-of-care imaging systems (e.g., MiniCAT™, xCAT™). Inputs: medical images from CT, MR, CR, US, NM modalities. Architecture: XoranVault™ (storage servers, RAID5) and XoranConnect™ (study server, web server, web client). Operation: imaging modality transfers study to storage server; user accesses via web browser. Features: off-site data storage; 3D/multiplanar visualization; teleradiology workflow; administrative reporting. Used in clinical settings; operated by healthcare professionals. Output: visualized/manipulated images for clinical review, interpretation, and distribution. Benefits: enables remote point-of-expertise viewing and interpretation of imaging studies.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Web-based software system; runs on standard off-the-shelf PC and networking hardware. Architecture includes storage servers (RAID5) and web-based viewing client. Connectivity requires active Internet connection. Software-based image processing for 3D and multiplanar visualization.
Indications for Use
Indicated for use by clinicians to display, edit, review, store, print, and distribute medical images acquired from CT, MR, CR, US, NM, and other imaging modalities. Prescription use only.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Xoran Technologies. The logo features a stylized "X" in a box-like shape, followed by the word "ORAN" in bold, sans-serif font. Below "ORAN" is the word "TECHNOLOGIES" in a smaller, lighter font.
K073306
pg 1 of 2
XoranConnect™ 510(k) Premarket Notification
# 9 510(k) Summary
#### 9.1 Name and Address of Manufacturer
Xoran Technologies, Inc.
309 North First Street Ann Arbor, MI 48103
Phone: (734) 663 7194 (734) 663 8500 Fax:
### 9.2 Contact Information
Amy Kim Regulatory Affairs Manager
(734) 663 7194
Date of Submission: November 15, 2007
## 9.3 Product Identification
| Trade Name: | XoranConnect™ |
|--------------------|----------------------------------------------------------------------------------------------------------------|
| Common Name: | PACS |
| Classification: | Class II, Product code LLZ (Image Processing System - PACS, Sec.<br>892.2050). |
| Guidance Document: | Guidance for the Submission Of Premarket Notifications for Medical Image<br>Management Devices (July 27, 2000) |
## 9.4 Substantially Equivalent (Predicate) Devices
XoranConnect™ software is substantially equivalent to the following predicate devices:
Visage PACS/CS (Mercury Computer Systems, Inc.), K072205
NovaPACS (NovaRad Corporation), K061920
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Xoran Technologies. The logo features a stylized "X" with the word "XORAN" in bold, sans-serif font to the right of the "X". Below "XORAN" is the word "TECHNOLOGIES" in a smaller, sans-serif font. The logo has a simple, modern design.
XoranConnect™ 510(k) Premarket Notification
# 9.5 Device Description
XoranConnect™ is a web-based service designed to complement point-of-care medical imaging systems by providing point-of-expertise viewing and interpretation tools. XoranConnect™ runs on standard offthe-shelf PC and networking hardware, all of which has been manufactured by third party reputable manufacturers.
The system performs these major functions.
- Off-site data storage allows user studies to be stored off-site. .
- Study viewer allows studies to be visualized and manipulated in 3D, multiplanar and other forms. .
To accommodate the major functions, the overall system architecture includes the following distinct components:
- XoranVault™ storage system (storage servers, dispatcher) .
- XoranConnect™ viewing system (study server, web server, web client) .
The storage (backup) system consists of one or more servers designed to store and serve stored imaging content on a RAID5 storage device. Imaging modality negotiates the connection and transfers study to the storage server.
Once stored, imaging data is ready to be served to the requesting clients. When a user logs into the system by visiting a XoranConnect web page, study server begins to serve the data, which in turn is displayed in the user's web browser window. Any storage server may simultaneously serve multiple requests for images.
It is the user's responsibility to ensure that the appropriate PC hardware and display device are used to run the XoranConnect™ software application. XoranConnect™ requires at all times an active Internet connection to operate. It is the user's responsibility to ensure that the connection that is being used provides the speed, stability and reliability required to utilize XoranConnect™ for any given purpose or application.
#### 9.6 Intended Use
XoranConnect™ is a web-based service designed to complement point-of-care medical imaging systems by providing point-of-expertise tools for viewing, interpretation, clinical review, analysis, distribution, processing and printing. Primary components include online multiplanar image viewing, offsite data storage, administrative reporting tools, and teleradiology workflow. It is intended as an extension to (but not limited to) the MiniCAT™ and xCAT™ CT imaging devices.
#### 9.7 Comparison With the Predicate Devices
XoranConnect™ has the same indication for use as the predicate devices and they perform very similar functions. Apart from differences in menus, usability, specific tool implementations, appearance and price, all of XoranConnect™ functions can be found in at least one of the predicate devices (most likely both). All three devices perform similar functions, while the predicate devices have many additional features.
#### 9.8 Conclusion
Functionality that XoranConnect™ provides is found on both predicate devices, and is therefore substantially similar to the predicate devices Visage PACS/CS (Mercury Computer Systems, Inc.), K072205, and NovaPACS (NovaRad Corporation), K061920. This design complies with 21 CFR 892.2050 and does not pose new safety risks.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing-like shapes above a body with two curved lines representing legs or a base.
JAN - 7 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Amy S. Kim, DDS Regulatory Affairs Manager Xoran Technologies, Inc. 309 N First Street ANN ARBOR MI 48103
Re: K073306
Trade/Device Name: XoranConnect™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 21, 2007 Received: November 30, 2007
Dear Dr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K07-3306
XoranConnect™ Device Name:
Indications for Use:
XoranConnect is a medical imaging software system intended to display, edit, review, store, print and distribute images acquired from imaging devices such as Computed Tomography (CT), Magnetic Resonance (MR), Computed Radiography (CR), Ultrasound (US), Nuclear Medicine (NM), and other devices.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Vorquimthn
(División Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 073306 510(k) Number
Page 1 of 1
Division Sign-Off
510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.