← Product Code [IYE](/productcode/IYE) · K073273

# EQUAL DOSE 1.0 (K073273)

_Nucletron Corporation · IYE · Dec 11, 2007 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K073273

## Device Facts

- **Applicant:** Nucletron Corporation
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** Dec 11, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

EQUAL Dose 1.0 is a quality assurance (QA) tool intended for independent verification of the dose calculations performed during treatment planning for external beam therapy, prior to the clinical treatment start. The intended user is a medical physicist or someone well familiar with the dosimetric concepts that are of importance in external beam therapy.

## Device Story

EQUAL Dose 1.0 is a quality assurance (QA) software tool for independent verification of external beam radiation therapy dose calculations. Input data consists of treatment plans imported via DICOM RT Plan format. The software performs dose calculations for megavoltage photon beams (4-30 MV) delivered by standard medical linear accelerators. Users, typically medical physicists, can edit calculation parameters such as coordinates and depths of calculation points, though treatment delivery parameters are locked. The output provides a verification of the dose calculation, allowing the clinician to confirm the accuracy of the treatment plan before the clinical start. This process helps ensure the safety and precision of radiation delivery to the patient.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Software-based QA tool for radiation therapy. Operates on Windows XP platform. Inputs: DICOM RT Plan format. Supports megavoltage photon beams (4-30 MV).

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- DCM1.0 ([K011246](/device/K011246.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Traditional 510(k) Nucletron EQUAL Dose 1.0

K073273

Page 31
March 2007

Image /page/0/Picture/2 description: The image contains the word "Nucletron" in bold, sans-serif font, next to a square logo. The logo contains a black circle with two white circles on opposite sides of the black circle. The logo and the word "Nucletron" are aligned horizontally.

NUCLETRON B.V.

Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557133 Fax +31 318 550485

Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Traditional 510(k) section

DEC 1 1 2007

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION as required by section 807.92(c)

### Submitter of 510(k):

| Company name:        | Nucletron Corporation                                   |
|----------------------|---------------------------------------------------------|
| Registration number: | 1121753                                                 |
| Address:             | 8671 Robert Fulton Drive<br>Columbia, MD 21046          |
| Phone:               | 410-312-4100                                            |
| Fax:                 | 410-312-4197                                            |
| Correspondent:       | Lisa Dimmick<br>Director Assurance & Regulatory Affairs |

#### New Device Name:

| Trade/Proprietary Name: | EQUAL Dose 1.0                                |
|-------------------------|-----------------------------------------------|
| Common/Usual Name:      | Radiation Therapy Verification Tool           |
| Classification Name:    | System, Planning, Radiation Therapy Treatment |
| Classification:         | 21Cfr892.5050 Class II                        |

## Legally Marketed Device(s)

:

Our device is substantially equivalent to the legally marketed predicate device cited in the table below:

| Manufacturer | Device | 510(k) # |
|--------------|--------|----------|
| MDS Nordion  | DCM1.0 | K011246  |

#### Description:

MUV 1.0 is a quality assurance (QA) tool intended for independent verification of the dose calculations performed during treatment planning for external beam therapy, prior to the clinical treatment start. The intended user is a medical physicist or someone well familiar with the dosimetric concepts that are of importance in external beam therapy.

{1}------------------------------------------------

## Received Event (Event Succeeded)

| Date:       | 2007-08-30   | Time:   | 04:35 |
|-------------|--------------|---------|-------|
| Duration:   | 2 min 30 sec | Sender: | ---   |
| Fax Number: |              |         |       |

#### Traditional 510(k) Nucletron EQUAL Dose 1.0

Page 32 March 2007

The software requires beam data for individual treatment plans to be imported through the DICOM RT Plan format. The parameters associated with the treatment delivery on the accelerator can not be edited within the software, although parameters reliated to the calculation inside the patient/phantom, such as the coordinates and depths of the calculation points, are open for editing.

The software is validated for broad megavoltage photon beams in the range from 4 up to 30 MV, delivered by standard medical linear accelerators (linacs),

The software runs on a Windows XP platform.

Intended use:

The legally marked predicate device has a wider intended use as the new device cited:

DCM1.0 (510K # K011246)→ DCM is a three-dimensional radiotherapy dose engine for radiation dose planning of patients undergoing external beam treatment in the oncology clinic. Based on quality assured radiation therapy input data Dose Calculation Module (DCM) is used to plan radiation treatment with:

- Linear accelerators with X-ray energies from 4 to 50MV .
- . Cobalt-60 units

DCM will calculate dose for 3D radiotherapy treatment approaches of combined modality plans, asymmetric and non-copianar fields; total body irradiation; modality plans;
dynamic wodanar fields; total body irradiation; multi-leaf collimators; motorized and dynamic wedges; customized blocking.

EQUAL Dose 1.0 → MUV 1.0 is a quality assurance (QA) tool intended for independent thereove aries to the stime to a quiations performed during for external beam therapy, prior to the clinical treatment start.

The intended user is a medical physicist or someone well familiar with the dosimetric concepts that are of importance in external beam therapy.

## Summary of technological considerations:

The cleared predicate device, DCM1.0, 510(K)#: K011246, has a wide intended use than the new device EQUAL Dose 1.0. EQUAL Dose is merely a verification tool.

e

2007-08-30

Date

Name: Paul van den Biggelaar Title: Director Oncentra Nucletron B.V. Veenendaal, The Netherlands

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, representing the department's mission to protect the health of all Americans. The eagle is depicted in a simple, abstract style with bold lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# DEC 1 1 2007

Nucletron Corporation % Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Service 2307 E. Aurora Rd., Unit B7 TWINSBURG OH 44087

Re: K073273

Trade/Device Name: EQUAL Dose 1.0 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: November 20, 2007 Received: November 21, 2007

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.

Sincerely yours,

Nancy C Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Nucletron" in bold, black font next to a square logo. The logo contains a black circle with two white circles on either side. The logo is enclosed in a black square.

Nucletron Corporation 7080 Columbia Gateway Drive Columbia, MD 21046-2133 U.S.A. 410-312-4100 Phone 410-312-4197 Fax

# Indications for Use

510(k) Number

EQUAL Dose 1.0 Device Name

Indications for Use

EQUAL Dose 1.0 is a quality assurance (QA) tool intended for independent verification of the dose calculations performed during treatment planning for external beam therapy, prior to the clinical treatment start. The intended user is a medical physicist or someone well familiar with the dosimetric concepts that are of importance in external beam therapy.

Prescription Use _X___ (Part 21 CFR 801 subpart D) AND/OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Aori M. Whay
(Division Sign-Off)

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

---

**Source:** [https://fda-staging.innolitics.com/device/K073273](https://fda-staging.innolitics.com/device/K073273)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
