← Product Code [QSX](/productcode/QSX) · K073264

# D-STAT DRY CLEAR HEMOSTATIC BANDAGE, MODEL 3005 (K073264)

_Vascular Solutions, Inc. · QSX · Dec 7, 2007 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K073264

## Device Facts

- **Applicant:** Vascular Solutions, Inc.
- **Product Code:** [QSX](/productcode/QSX.md)
- **Decision Date:** Dec 7, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

The D-Stat Dry Clear is applied topically as an adjunct to manual compression and is indicated for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes and reducing the time-to-hemostasis in patients undergoing diagnostic endovascular procedures utilizing a 4-6 Fr. introducer sheath.

## Device Story

Topical hemostatic bandage; consists of lyophilized pad containing bovine-derived thrombin, sodium carboxymethylcellulose (CMC), and calcium chloride; includes transparent sterile bandage. Applied topically over bleeding source; used as adjunct to manual compression. Thrombin facilitates hemostasis via enzymatic conversion of fibrinogen to fibrin; CMC and calcium chloride support clot formation. Used in clinical settings by healthcare providers for vascular access sites. Provides physical barrier to blood flow; reduces time-to-hemostasis for patients undergoing diagnostic endovascular procedures.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Lyophilized pad containing bovine-derived thrombin, sodium carboxymethylcellulose (CMC), and calcium chloride. Includes transparent sterile bandage. Operates via enzymatic clotting cascade activation. Sterilized using same processes as predicate.

## Regulatory Identification

To temporarily control bleeding and cover external wounds.

## Predicate Devices

- D-Stat® Dry Topical Hemostat ([K030836](/device/K030836.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 10, 2023

Vascular Solutions, Inc. c/o James Chapman Regulatory Affairs Associate 907 South Lakewood Avenue Baltimore, Maryland 21224

Re: K073264

Trade/Device Name: D-Stat® Dry Clear, Hemostatic Bandage Regulatory Class: Unclassified Product Code: QSX

Dear James Chapman:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 7, 2007. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSX.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov.

Sincerely,

Julie A. Morabito -S

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Vascular Solutions, Inc. % Mr. James Chapman Regulatory Affairs Associate 907 South Lakewood Avenue Baltimore, Maryland 21224

DEC 7 2007

Re: K073264

Trade/Device Name: D-State® Dry Clear Hemostatic Bandage Regulatory Class: Unclassified Product Code: FRO Dated: November 19, 2007 Received: November 20, 2007

Dear Mr. Chapman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

#### Page 2 - Mr. James Chapman

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark M. Millerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use Statement

510(k) Number: K073264

Device Name: D-Stat

®

Dry Clear Hemostatic Bandage

Indications for Use:

The D-Stat Dry Clear is applied topically as an adjunct to manual compression and is indicated for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes and reducing the time-to-hemostasis in patients undergoing diagnostic endovascular procedures utilizing a 4-6 Fr. introducer sheath.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) (Division bigand Neurological Devices

103264

**510(k) Number:**

{4}------------------------------------------------

# Special 510(k) Summary (As required by 21 CFR 807.92(c))

K073264 Special 510(k) Number: Page 1 of 2

Date Prepared: November 19, 2007

#### Submitter Information

| Submitter's Name: | Vascular Solutions, Inc. |
|-------------------|--------------------------|
| Address:          | 6464 Sycamore Court      |
|                   | Minneapolis, MN 55369    |

Establishment Registration: 2134812

| Contact Person: | James Chapman                |
|-----------------|------------------------------|
|                 | Regulatory Affairs Associate |
| Phone:          | (763) 656-4300 ext. 380      |
| Fax:            | (763) 656-4253               |

#### Device Information

| Trade Name:          | D-Stat® Dry Clear, Hemostatic Bandage |
|----------------------|---------------------------------------|
| Common Name:         | Topical hemostat                      |
| Classification Name: | Unclassified                          |
| Product Code:        | FRO                                   |
| Regulation:          | Not Applicable                        |

### Predicate Device

The predicate device is the currently marketed D-Stat® Dry Topical Hemostat (K030836).

#### Device Description

The D-Stat® Dry Clear Hemostatic Bandage consists of a lyophilized pad containing bovinederived thrombin as an aid to hemostasis (King Pharmaceutical license number 0977), sodium carboxymethylcellulose (CMC), and calcium chloride. Included is a transparent sterile bandage attached to the primary packaging to apply over the hemostatic pad. The only difference between the D-Stat® Dry Clear Hemostatic Bandage and the predicate is the replacement of the opaque sterile bandage included with the device with a transparent sterile bandage.

The D-Stat® Dry Clear is applied directly over the source of bleeding, creating a physical barrier to blood flow through the application of adjunctive manual compression. The lyophilized components (thrombin, CMC, and calcium chloride) establish an environment in which a natural blood clot can build and form a physical barrier to bleeding.

OEC 7 2007

{5}------------------------------------------------

K073264

Vascular Solutions, Inc. D-Stat® Dry Clear Hemostatic Bandage Special 510(k) Pre-market Notification Page 2 of 2

The thrombin facilitates hemostasis by enhancing the surface-activated clotting cascade through enzymatic cleavage and conversion of fibrinogen to fibrin. This fundamental scientific technology is exactly the same as the predicate D-Stat Dry hemostatic bandage,

## Intended Use/Indications for Use

The D-Stat Dry Clear has the same indications for use and is intended to be used in the same manner as the predicate D-Stat Dry. The D-Stat Dry Clear is applied topically as an adjunct to manual compression and is indicated for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes and reducing the time-to-hemostasis in patients undergoing diagnostic endovascular procedures utilizing a 4-6 Fr. introducer sheath.

#### Summary of Design Control Activities

The risk analysis method used to assess the impact of the modification to the device was a Failure Modes and Effects Criticality Analysis. Based on the risk assessment, no design verification tests were required for inclusion of the sterile clear compression bandage to the D-Stat Dry Clear device.

A declaration of conformity with design controls is included in Section 7.

## Summary of Clinical Testing

No human clinical testing was required for this device.

## Statement of Equivalence

The D-Stat® Dry Clear Hemostatic Bandage and the currently marketed D-Stat® Dry Hemostatic Bandage have the following commonalities:

- · Both have the same indicated use,
- · Both use the same operating principle
- · Both incorporate the same basic device design.
- · Incorporate the same materials.
- · Have the same shelf life, and
- · Are packaged and sterilized using the same materials and processes

#### Conclusion

In summary, the D-Stat® Dry Clear Hemostatic Bandage is substantially equivalent to the currently marketed D-Stat® Dry Hemostatic Bandage based on a comparison of the indications for use and the technological characteristics of the device.

---

**Source:** [https://fda-staging.innolitics.com/device/K073264](https://fda-staging.innolitics.com/device/K073264)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).