← Product Code [QSZ](/productcode/QSZ) · K073251

# HYALOMATRIX PA (K073251)

_Fidia Advanced Biopolymers Srl · QSZ · Dec 14, 2007 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K073251

## Device Facts

- **Applicant:** Fidia Advanced Biopolymers Srl
- **Product Code:** [QSZ](/productcode/QSZ.md)
- **Decision Date:** Dec 14, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

HYALOMATRIX® is indicated for the management of wounds including: partial and full-thickness wounds; second-degree burns; pressure ulcers; venous ulcers; diabetic ulcers; chronic vascular ulcers; tunneled/undetermined wounds; surgical wounds (donor sites/grafts, post-Moh's surgery , post-laser surgery, podiatric, wound dehiscence); trauma wounds (abrasions, lacerations, skin tears); and draining wounds. The device is intended for one-time use.

## Device Story

HYALOMATRIX is a sterile, flexible, bilayered wound dressing. Input: wound surface. Transformation: nonwoven HYAFF 11 (benzyl ester of hyaluronic acid) pad acts as a biodegradable scaffold for cellular invasion and capillary growth; semipermeable silicone membrane controls water vapor loss and provides tear strength. Output: physical wound covering. Used in clinical settings by healthcare providers. Benefits: promotes healing via scaffold structure and moisture management. One-time use.

## Clinical Evidence

Bench testing only. Biocompatibility assessed via in vitro and in vivo studies: cytotoxicity, hemolysis, acute oral/dermal toxicity, irritation, sensitization, and genotoxicity assays. Implantation studies confirmed biodegradability of HYAFF 11 within 4 months. Pyrogenicity tested via LAL. Silastic Medical Adhesive and Loctite cured film components also underwent comprehensive biocompatibility testing (cytotoxicity, sensitization, implantation, etc.).

## Technological Characteristics

Bilayered dressing: nonwoven HYAFF 11 (benzyl ester of hyaluronic acid) pad and semipermeable silicone membrane. Sterile, flexible, conformable. Biodegradable contact layer. No electronic components or software.

## Regulatory Identification

Intended as a physical barrier to cover the wound and provide a moist wound environment. Intended to be left on wounds after topical application.

## Predicate Devices

- Bilayer Matrix Wound Dressing ([K001508](/device/K001508.md))
- Oasis Wound Matrix ([K021792](/device/K021792.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Anika Therapeutics, S.r.l.

K073251
510(k) K073251
HYALOMATRIX®

## 510(k) SUMMARY

DEC 1 4 2007

Anika Therapeutics, S.r.l. (former Fidia Advanced Biopolymers, S.r.l) 510(k) Owner: Via Ponte dell Fabbrica 3/B 35031 Abano Terme PADOVA, ITALY Contact: Dr. Adriano Zanotti Regulatory Affairs Phone: +39-049-82 32 650

+39-049 86 11 580

HYALOMATRIX™ KC (Laserskin) Wound Dressing

HYALOMATRIX®

Wound Dressing Product Code FRO

Unclassified

Date Summary Prepared:

May 19, 2011 Revisel

Trade Name: Common/Classification Name:

Classification:

Fax:

Predicate Devices:

Device:

K001508 Bilayer Matrix Wound Dressing Integra LifeSciences Corporation

Fidia Advanced Biopolymers S.r.l.

K021792 Oasis Wound Matrix

Cook Biotech, Inc. K061711

Device

Description:

Lateral View

matrix acts as a scaffold for cellular invasion and capillary growth. HY ALOMATRIX is indicated for the management of wounds including: Intended Use: partial and full-thickness wounds; second-degree burns; pressure ulcers; venous ulcers; diabetic ulcers; chronic vascular ulcers; tunneled/undetermined wounds; surgical wounds (donor sites/grafts, post-Moh's surgery , post-laser surgery, podiatric, wound dehiscence); trauma wounds (abrasions, lacerations, skin tears); and draining wounds. The device is intended for one-time use.

HY ALOMATRIX is a bilayered, sterile, flexible, and conformable wound

dressing that acts as an advanced wound care device. It is comprised of a nonwoven pad entirely composed of HYAFF 11, a benzyl ester of hyaluronic acid, and a semipermeable silicone membrane, which controls water vapor loss, provides a flexible covering for the wound surface, and adds increased tear strength to the device. The HYAFF 11 wound contact layer biodegradable

Technological Similar to the predicate devices, HYALOMATRIX is a biologically-derived Characteristics: wound dressing that provides a scaffold for in-growing cells, and like the Integra Bilayer Matrix Wound Dressing, it contains a semipermeable silicone

May 19, 2011

CONFIDENTIAL

{1}------------------------------------------------

layer to prevent moisture loss.

Biocompatibility The biocompatibility of HYAFF 11 (powder or film), one component of Data: HY ALOMATRIX, was assessed in a series of in vitro and in vivo studies.

> Cytotoxicity and hemolysis studies demonstrate that HYAFF 11 films are neither cytogenic nor hemolytic. Acute oral and dermal toxicity studies on HYAFF 11 powder indicate that the lethal dose of HYAFF 11 is greater than 5,000 mg/kg and 2,000 mg/kg, respectively. Irritation studies demonstrate that HYAFF 11 powder is not an ocular or dermal irritant, and a sensitization study provides evidence for the lack of a sensitizing effect. No evidence of genotoxicity was observed in three in vitro assays or one in vivo study. An implantation study of up to 1 year found no evidence of treatment-related toxicity, and demonstrated that the HYAFF 11 film was degraded within 4 months. The Limulus Amebocyte Lysate ("LAL") test is used to assess the pyrogenicity of HYAFF 11 as part of the finished product specifications, and the endotoxin levels are within specification (≤0.2 EU/mg).

No toxicity was observed in USP XXII (Plastic Containers tests) testing, indicating that any potential leachable substances present in HYAFF 11 from the packaging used for HYALOMATRIX have no effect on the safety of this device.

Biocompatibility testing on the Silastic® Medical Adhesive Silicone, Type A included cytoxicity in multiple cell culture assays; pyrogenicity; U.S. P. Class V testing; hemolysis; 90-day implantation; and guinea pig skin sensitization. The material produced cytotoxic effects in multiple cell culture assays when placed in direct contact and when extracts of the material were tested. Pyrogenicity, U.S.P. Class V, and hemolysis assays were negative. Implantation studies with cured material resulted in no noteworthy histological findings, and saline extracts did not produce a skin sensitization response in guinea pigs.

Biocompatibility testing on the Loctite cured film included intracutaneous irritation; acute toxicity; cytotoxicity; hemocompatibility; hemolysis; implantation; and U.S.P. physicochemical tests. The material was nonirritating, not acutely toxic; non-cytotoxic; non-hemolytic; hemocompatible; produced no treatment-related histopathological findings upon implantation; and passed the U.S.P. physicochemical tests.

Performance Data:

Clinical experience HYALOMATRIX as a wound dressing supports the safety and effectiveness of the device.

Conclusions:

The performance data discussed above demonstrate that the device is as safe and effective, and performs as well as the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of a human figure in profile, with three overlapping figures suggesting community and support.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUL 19 2011

Fidia Advanced Biopolymers S.r.l. % Morgan, Lewis & Bockius, LLP Sharon A. Segal, Ph.D. 1111 Pennsylvania Avenue, NW Washington, District of Columbia 20004

Re: K073251

Trade/Device Name: HYALOMATRIX® PA Regulation Number: Unclassified Product Code: FRO Dated (Date on orig SE ltr): November 19, 2007 Received (Date on orig SE ltr): November 26, 2007

Dear Dr. Segal:

This letter corrects our substantially equivalent letter of December 14, 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Sharon A. Segal. Ph.D.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1118001/phm.for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFFR, Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollofree minber (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Mark A. Millman

Mark Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

## Indications for Use

510(k) Number K073251 (if known):

HYALOMATRIX ® Device Name:

HYALOMATRIX® is indicated for the management of wounds Indications for Use: including: partial and full-thickness wounds; second-degree burns; pressure ulcers; venous ulcers; diabetic ulcers; chronic vascular ulcers; tunneled/undetermined wounds; surgical wounds (donor sites/grafts, post-Moh's surgery , post-laser surgery, podiatric, wound dehiscence); trauma wounds (abrasions, lacerations, skin tears); and draining wounds. The device is intended for one-time use.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Kromeda MXM
(Division Sign Off)

Division Sign Division of Surgical. Orthopedic. and Restorative Devices

510(k) Number K073251

---

**Source:** [https://fda-staging.innolitics.com/device/K073251](https://fda-staging.innolitics.com/device/K073251)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
