K073225 · Hansen Medical, Inc. · DXX · Jun 30, 2008 · Cardiovascular
Device Facts
Record ID
K073225
Device Name
SENSEI CATHETER CONTROL SYSTEM
Applicant
Hansen Medical, Inc.
Product Code
DXX · Cardiovascular
Decision Date
Jun 30, 2008
Decision
SESU
Submission Type
Traditional
Regulation
21 CFR 870.1290
Device Class
Class 2
Indications for Use
The Sensei Robotic Catheter System, Artisan Catheter, and accessories are intended to facilitate manipulation, positioning and control, for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems, using the following percutaneous mapping catheters: the Polaris-Dx™ Steerable Diagnostic catheters made by Boston Scientific Corporation and the Livewire™ Electrophysiology catheters made by St. Jude Medical.
Device Story
Sensei Catheter Control System is a master/slave robotic system for percutaneous catheter manipulation in heart atria. Physician operates system from a remote workstation, remaining seated away from x-ray radiation source. System enables precise positioning and visualization of steerable catheter tips. Modified version integrates 3D electro-anatomic maps from third-party mapping systems onto the central workstation screen. Device facilitates electrophysiological data collection; does not perform ablation. Benefits include reduced radiation exposure for physicians and improved catheter control. Used in clinical electrophysiology settings.
Clinical Evidence
No clinical data provided. Substantial equivalence based on design and engineering data.
Technological Characteristics
Master/slave robotic catheter control system. Includes Artisan steerable guide catheter, remote catheter manipulator, and workstation. Integrates third-party 3D electro-anatomic mapping data. Connectivity includes interface with external mapping/recording systems. Class II device.
Indications for Use
Indicated for facilitating manipulation, positioning, and control of percutaneous mapping catheters (Polaris-Dx or Livewire) to collect electrophysiological data within the heart atria. Contraindicated for use with cardiac ablation catheters for treating cardiac arrhythmias, including atrial fibrillation.
Regulatory Classification
Identification
A steerable catheter control system is a device that is connected to the proximal end of a steerable guide wire that controls the motion of the steerable catheter.
{0}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## 510(k) Number:
## Applicant Information:
| Owner Name: | Hansen Medical, Inc. |
|-------------------|--------------------------------------------------------------------------|
| Address: | 380 N. Bernardo Ave.<br>Mountain View, CA. 94043<br>Office: 650-404-5800 |
| Contact Person: | Doug Worth, RAC |
| Phone Number: | 650 404 5800 |
| Facsimile Number: | 650 404 5901 |
| Date Prepared: | 11/14/2007 (revised 1/7/08) |
## Device Information:
| Classification: | Class II |
|----------------------|------------------------------------------------------------|
| Trade Name: | Sensei Catheter Control System |
| Common name: | Steerable Catheter Control System |
| Classification name: | System, Catheter Control, Steerable, (21 CFR 870.1290/DXX) |
## Predicate Devices:
The modified Hansen Medical Sensei Catheter Control system is substantially equivalent in intended use and method of operation to the earlier Sensei system (K052480).
## Device Description:
The Hansen Medical Sensei Catheter Control, Artisan Catheter and accessories and are designed to facilitate manipulation, positioning and control of mapping percutaneous catheters within the atria of the heart. The fundamental concept of the system is based on a master/slave control system that enables and visualizes positioning of a steerable catheter tip at a desired point inside the heart, whilst enabling a physician to remain seated and away
K073225 Hansen Medical
Attachment 2, Page 2 of 3 Indication for Use and 510(k) Summary
{1}------------------------------------------------
from the x-ray radiation source The Catheter Control System has been modified to display 3D electroanotomic maps from a 30 party mapping system in the center screen of the Sensei system.
### Intended Use:
The Sensei Robotic Catheter System, Artisan Catheter, and accessories are intended to facilitate manipulation, positioning and control, for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems, using the following percutaneous mapping catheters: the Polaris-Dx™ Steerable Diagnostic catheters made by Boston Scientific Corporation and the Livewire™ Electrophysiology catheters made by St. Jude Medical.
### Comparison to Predicate Device(s):
The modified Hansen Medical Sensei Catheter Control System is substantially equivalent to the predicate device.
#### Substantial equivalence:
Based upon the indications for use and the design and engineering data provided in this pre-market notification, the Hansen Medical Sensei Robotic Catheter System, Artisan Catheter and have been shown to be substantially equivalent to a currently marketed predicate device.
K073225 Hansen Medical
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines above a wavy base.
Public Health Service
JUN 3 0 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Hansen Medical, Inc. c/o Doug Worth, RAC Manager, Regulatory Affairs 380 N. Bernardo Ave. Mountain View, CA 94043
Re: K073225
> Trade/Device Name: Sensei Catheter Control System, Model 02057 Regulation Number: 21 CFR 870.1290 Regulation Name: Steerable Catheter Control System Regulatory Class: Class II Product Code: DXX and DRA Dated: May 29, 2008 Received: May 30, 2008
Dear Mr. Worth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling, on the packaging for the Artisan Steerable Guide Catheter and Sheath, on the Remote Catheter Manipulator, and the Workstation:
The safety and effectiveness of this device for use with cardiac ablation catheters, in the treatment of cardiac arrhythmias including atrial fibrillation, have not been established.
Furthermore, this warning must be prominently displayed on the Remote Catheter Manipulator, Workstation, all labeling, including pouch box, and carton labels, instructions for use and other
{3}------------------------------------------------
Page 2 -- Mr. Doug Worth, RAC
promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
NOCETIN
Donna-Bea Tillman, Ph.D., M.P.A. Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): Ko13225
- Hansen Medical Sensei Catheter Control System Device Name: Hansen Medical Artisan Steerable guide Catheter
#### Indications for Use:
The Sensei Robotic Catheter System, Artisan Catheter, and accessories are intended to facilitate manipulation, positioning and control, for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems, using the following percutaneous mapping catheters: the Polaris-Dx™ Steerable Diagnostic catheters made by Boston Scientific Corporation and the Livewire™ Electrophysiology catheters made by St. Jude Medical.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Deeth
K073225 Hansen Medical
Attachment 2, Page 1 of 3 Indication for Use and 510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.