HYPODERMIC NEEDLE-PRO EDGE SAFETY DEVICE

K073188 · Smiths Medical Asd, Inc. · FMI · Feb 8, 2008 · General Hospital

Device Facts

Record IDK073188
Device NameHYPODERMIC NEEDLE-PRO EDGE SAFETY DEVICE
ApplicantSmiths Medical Asd, Inc.
Product CodeFMI · General Hospital
Decision DateFeb 8, 2008
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is intended for injection or aspiration of fluids using a Luer lock or Luer slip syringe. The needle protection device covers the needle after use to help prevent needle sticks. The EDGE™ safety device may be removed from the syringe when required by a specific medical procedure.

Device Story

Hypodermic needle with integrated safety protection mechanism; one-piece design featuring needle hub and protective sheath connected by living hinge. Used for fluid injection or aspiration. Operated by clinicians (e.g., phlebotomists) via one-handed technique; needle pressed into sheath post-procedure; lugs on hub snap into clips on sheath for containment; includes needle retaining hooks. Sheath features arrow indicator for bevel orientation. Device can be removed from syringe for specific procedures and discarded in sharps container. Benefits include reduction of accidental needle sticks.

Clinical Evidence

Simulated clinical use studies conducted with phlebotomists. Results confirmed device usage required no substantial change to existing clinical techniques and allowed for safe removal from the syringe after activation.

Technological Characteristics

Hinged-style protective sheath; one-piece hub/sheath design; Luer lock/slip compatibility. Conforms to ISO 594/1:1986(E) (general Luer requirements), ISO 594-2:1998(E) (Luer lock fittings), and ISO 7864:1993(E) (sterile hypodermic needles). Manual activation; mechanical snap-lock containment.

Indications for Use

Indicated for injection or aspiration of fluids in patients requiring hypodermic needle use. Compatible with Luer lock or Luer slip syringes. No specific age or gender contraindications noted.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Smiths Medical. The logo is in two parts, with "Smiths" in a bold, sans-serif font and "medical" in a smaller, lighter font. Above the logo is the text "K073/88" in a handwritten style. bringing technology to life # SECTION 5, 510(k) Summary ## Company Information: Smiths Medical ASD, Inc. 10 Bowman Drive Keene, NH 03431 (603) 352-3812, prompt 4, ext 2493 Contact: Brian D. Farias Regulatory Affairs Manager Smiths Medical ASD Inc. Critical Care Division 10 Bowman Drive Keene, NH 03431 0724, USA T: +1 603 352 3812 F: +1 603 355 8157 www.smiths-medical.com Summary Prepared: November 9, 2007 #### Product Name: Trade Name: Hypodermic Needle-Pro® EDGE™ Safety Device Common Name: Hypodermic Needle with attached needle protection. Accessory to Blood Sampling Syringes. Classification Name: Hypodermic Single Lumen Needles (21 CFR 880.5570, Product Code FMI). Accessory to Tubes, Vials, Systems, Serum Separators, Blood Collection (21 CFR 862.1675. Product Code JKA). # Predicate Device(s): K071785 (Smiths Medical ASD, Inc.) Hypodermic Needle-Pro® EDGE™ Safety Device and EDGE™ Safety Device with Syringe K911037 (Smiths Medical ASD, Inc.- formerly Concord/Portex) Safety Needle Sheath (modification) #### Device Description: This device is intended for injection or aspiration of fluids. The needle protection device covers the needle after use to help prevent needle sticks. The device features a "one-piece" design of needle hub and protective sheath with a living hinge. The needle cannula is permanently affixed into the hub. The sheath has an "arrow" indicating the bevel orientation, i.e. when the sheath is oriented to the right, the bevel is in the "up position". After the procedure is completed, the needle is pressed into the sheath using a one-handed technique. As the needle enters the protective sheath, the lugs on the needle hub snap into the clips of the protective sheath resulting in a "bottom snap" and the needle is contained within the sheath. In addition to the "bottom snap", the needle protection sheath for needles equal to or less than !" has one needle retaining hook (except 30g) and the sheath for needles longer than 1" has two hooks. The EDGE™ safety device may be removed from the syringe when required by a specific medical procedure. An example is removal of the EDGE™ safety device from a blood sampling syringe. The EDGE™ safety device is then discarded into a sharps container. {1}------------------------------------------------ K073188 ### Indications for Use: Hypodermic Needle-Pro® EDGE" Safety Device: This device is intended for injection or aspiration of fluids using a Luer lock or Luer slip syringe. The needle protection device covers the needle after use to help prevent needle sticks. The EDGE™ safety device may be removed from the syringe when required by a specific medical procedure. #### Technological Characteristics: The proposed and predicate devices all employ the same hinged style protective sheath that is manually activated after use. ### Non-Clinical Data: This abbreviated 510(k) submission declares conformance to the following standards: ISO 594/1:1986(E), International Standard, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment-Part 1: General requirements ISO 594-2:1998(E), International Standard, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment-Part 2: Lock Fittings ISO 7864:1993(E), International Standard, Sterile hypodermic needles for single use #### Clinical Data: Simulated clinical use studies were conducted which confirmed that the EDGE™ safety device could be used by phlebotomists with no substantial change to their technique and could be safely removed from the syringe after activation. #### Conclusion: The bench testing and simulated clinical use studies demonstrate that the proposed device is safe and effective and is substantially equivalent to the predicate devices. Very truly yours. SMITHS MEDICAL ASD, INC. Brian D. Farias Regulatory Affairs Manager {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an abstract symbol, resembling a person embracing another, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB -8 2008 Mr. Brian D. Farias Regulatory Affairs Manager Smiths Medical ASD, Incorporated 10 Bowman Drive Keene, New Hampshire 03431 Re: K073188 Trade/Device Name: Hypodermic Needle-Pro EDGE" Safety Device Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: November 9, 2007 Received: November 13, 2007 Dear Mr. Farias: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Farias Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050. 'This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Clues Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # SECTION 4, Indications for Use Statement # Indications for Use 510(k) Number (if known): _ K 0 73/88 Device Name: Hypodermic Needle-Pro® EDGE™ Safety Device Indications for Use: This device is intended for injection or aspiration of fluids using a Luer lock or Luer slip syringe. The needle protection device covers the needle after use to help prevent needle sticks. The EDGE™ safety device may be removed from the syringe when required by a specific medical procedure. Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Chr. v. Page 1 of 1 (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices K473188 510(k) Number:
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...