The ASC R-Port II Laparoscopic Access Device is intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.
Device Story
ASC R-Port II is a sterile, single-use laparoscopic access port; used during minimally invasive abdominal surgery. Device consists of a retracting portion to maintain abdominal incision and a valve portion to maintain pneumoperitoneum. Input: surgical site access; Output: port for multiple laparoscopic instruments/cameras. Operated by surgeons in an OR setting. Delivery via Verres needle insufflation or Hasson cut-down incision. Device retracts tissue to allow instrument passage while ensuring gas seal. Benefits include facilitating multi-instrument access through a single incision, maintaining abdominal pressure, and providing a standardized alternative to traditional trocars.
Clinical Evidence
Bench testing only. Biocompatibility and verification testing demonstrated the device functions as intended and is safe and effective for its intended use.
Technological Characteristics
Sterile, single-use laparoscopic access port. Comprised of a retracting portion for incision management and a valve portion for pneumoperitoneum maintenance. Mechanical design; no electronic or software components. Dimensions and materials are consistent with the predicate R-Port (K070158).
Indications for Use
Indicated for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery in patients undergoing such procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
K073170
# 510(k) Summary for the Advanced Surgical Concepts (ASC) R-Port II Laparoscopic Access Device
#### 1. SUBMITTER/510(K) HOLDER
| Advanced Surgical Concepts | | |
|------------------------------------|------------------|-------------|
| Unit 4 Sunnybank Centre | | |
| Upper Dargle Road | | |
| Bray, County Wicklow | | |
| Ireland | | |
| Establishment Registration Number: | 9616720 | DEC 03 2007 |
| Contact Person: | Tanya Kavanagh | |
| Telephone: | 353-(0)1-2864777 | |
| Date Prepared: | November 8, 2007 | |
#### 2. DEVICE NAME
| Proprietary Name: | ASC R-Port II Laparoscopic Access Device |
|----------------------|---------------------------------------------|
| Common/Usual Name: | Laparoscopic Accessory |
| Classification Name: | Endoscopic Accessory and Surgical Retractor |
#### 3. PREDICATE DEVICES
- . R-Port Laparoscopic Access Device subject of K070158
- . Taut Inc. ADAPt Laparoscopic Port and Accessory subject of K010007
- . Ethicon Endopath III Trocar System subject of K032676
- ASC Ecotract Device subject of K010711 .
### DEVICE DESCRIPTION 4.
The Advanced Surgical Concepts (ASC) R-Port II Laparoscopic Access Device is intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery. The proposed ASC R-Port II Laparoscopic Access Device is identical in design to the R-Port Laparoscopic Access Device, which has been cleared for marketing under K070158. Two modifications have been made to the use of the proposed ASC R-Port II Laparoscopic Access device. ASC has made very minor
{1}------------------------------------------------
changes to the indications for use and the recommended method for delivery of the device although the design of the device is identical to the one described in K070158. The proposed ASC R-Port II Laparoscopic Access Device and the original R-Port Laparoscopic Access Device are otherwise identical in design, mechanism of action and operating principles.
The proposed ASC R-Port II Laparoscopic Access Device is a laparoscopic instrumentation access port that is used to perform the same function as other port systems and standard trocars. The original R-Port was offered with four different Introducer components for deployment but the proposed ASC R-Port II is offered with only one Introducer and a new delivery method. The new recommended delivery method for the proposed ASC R-Port II consists of initially insufflating the abdomen using a standard Verres Needle before creating the port incision. This is common practice with standard trocar insertions. In addition, as described in K070158, the ASC R-Port II can also be deployed after the creation of a Hasson cut-down incision. This type of incision is also commonly used for the deployment of standard trocars. The Distal Ring of the ASC R-Port II is delivered through the Hasson incision using the Introducer component of the ASC R-Port II which is the same Injector Introducer cleared for marketing under K070158.
The ASC R-Port II Laparoscopic Access Port is a sterile, disposable laparoscopic instrument port which performs two functions, as follows:
- . It retracts a small abdominal incision to allow multiple laparoscopic instruments to pass through to the abdomen
- It ensures that pneumoperitoneum is maintained in the abdomen during the . surgical procedure whether or not a laparoscopic instrument is passing through the port
Like the predicate R-Port Laparoscopic Access Device, the ASC R-Port II Laparoscopic Access Device is comprised of the following:
- a retracting portion which retracts an abdominal incision to allow the passage of . laparoscopic instruments
- . a valve portion which maintains the pneumoperitoneum established for the surgical procedure
{2}------------------------------------------------
## 5. INTENDED USE
The ASC R-Port II Laparoscopic Access Device is intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.
### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The ASC R-Port II provides an access path for laparoscopic instruments through a small incision in the abdominal wall identical to the function of the predicate ASC R-Port Laparoscopic Access Device (K070158) and similar to the predicate trocars. A standard trocar consists of a bladeless dissecting-tipped component for creating a small incision in an abdominal wall, and a rigid cannula, which keeps the small incision open, thereby providing an access path for laparoscopic instruments through the small incision in the abdominal wall. To deploy a trocar, a skin incision is created at the desired location in the patient's abdomen. The trocar creates its own incision through to the abdomen using its bladeless dissecting-tipped member which is rotated back and forth as it is advanced through the tissue layers. Once it has penetrated through to the abdomen, the bladeless dissecting-tipped member is removed, leaving the rigid cannula through which the laparoscopic instruments can be introduced. There is a valve system on the trocar to maintain pneumoperitoneum, whether an instrument is present in the trocar or not.
The proposed ASC R-Port II and the predicate ASC R-Port Laparoscopic Access Device are laparoscopic instrument access ports that are used to perform the same function as a trocar. The ASC R-Port II functions to both retract a small abdominal incision to allow laparoscopic instruments to pass through to the abdomen, and to maintain the pneumoperitoneum in the abdomen during the surgical procedure whether or not a laparoscopic instrument is passing through the port. The only difference between the proposed ASC R-Port II device and the predicate ASC R-Port Laparoscopic Access Device (K070158) is that the ASC R-Port II allows for insertion of multiple instruments or cameras at one time, and provides an additional method of delivery. Other than those modifications, the proposed ASC R-Port II and predicate ASC R-Port Laparoscopic Access Device are identical.
Like the predicate devices, the ASC R-Port II is a sterile, single-use device. The use of the ASC R-Port II and the predicates are identical in that they allow the passage of laparoscopic instrumentation while maintaining pneumoperitoneum. Although insertion of the ASC R-Port II and the predicate trocars differs, the ASC R-Port II insertion does not affect safety and effectiveness of the device incisions are made consistently and safely during laparoscopic surgery.
{3}------------------------------------------------
### 7. PERFORMANCE TESTING
Biocompatibility and verification testing have been performed which demonstrated that the ASC R-Port II Laparoscopic Access Device functions as intended and is safe and effective for its intended use.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
DEC 0 3 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Advanced Surgical Concepts % Medical Device Consultants, Inc. Ms. Mary McNamara-Cullinane 49 Plain Street North Attleboro, Massachusetts 02760
Re: K073170
Trade/Device Name: ASC R-Port II Laparoscopic Access Device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: November 8, 2007 Received: November 13, 2007
Dear Ms. McNamara-Cullinane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
# Page 2 - Ms. Mary McNamara-Cullinane
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Muriel M. Millican
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: ASC R-Port II Laparoscopic Access Device
Indications for Use:
The ASC R-Port II Laparoscopic Access Device is intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Division Sign-Off)**
(Division of General, Restorative, and Neurological D
**510(k) Number**
November 8, 2007
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.