K073169 · Scimage, Inc. · LLZ · Jan 9, 2008 · Radiology
Device Facts
Record ID
K073169
Device Name
SCIMAGE PICOMENTERPRISE
Applicant
Scimage, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 9, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The PicomEnterprise software is intended for acceptance, transfer, display, storage and digital processing of medical images. Its hardware components may include digitizers, workstations, communications devices, computers, video monitors, magnetic, optical disk, or other digital data storage devices and hardcopy devices. The software components provide functions for performing operations related to image manipulation, enhancement, compression or quantification. To support the diagnostic interpretation of Mammography studies, PicomEnterprise will display the full fidelity DICOM image in a non-compressed format. Images will be rendered with patient and clinical information clearly displayed as part of the DICOM Overlay as required by MQSA, on monitors cleared by FDA for use in Digital Mammography. Lossy compressed mammography images and digitized film screen images should not be used for the purpose of primary diagnosis. Mammographic images may only be interpreted using an FDA approved monitor that offers at least five megapixel resolution and meets other technical specifications reviewed and accepted by FDA.
Device Story
PicomEnterprise is a multi-modality PACS software system for 2D, 3D, and 4D medical image management. It accepts medical images from various modalities, facilitating transfer, storage, and display within a Microsoft Windows environment. The system performs image manipulation, enhancement, compression, and quantification. It is used by clinicians in healthcare settings to support diagnostic interpretation. For mammography, it ensures display of full-fidelity, non-compressed DICOM images on FDA-cleared 5MP monitors, adhering to MQSA requirements for patient/clinical data overlays. The device benefits patients by providing a standardized, compliant platform for accurate diagnostic review of medical imaging studies.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and compliance with voluntary standards (DICOM, JPEG).
Technological Characteristics
Software-based PACS system operating in a Microsoft Windows environment. Supports DICOM 3.0 and JPEG standards. Features include multi-modality image reconstruction, display, and processing algorithms. Designed for integration with hardware including digitizers, workstations, and storage devices. Requires FDA-cleared 5MP monitors for primary mammography diagnosis.
Indications for Use
Indicated for the acceptance, transfer, display, storage, and digital processing of medical images, including mammography studies, for diagnostic interpretation by clinicians.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
NetraTM Workstation and NetraMDTM Software (K960911)
NetraTM Workstation and NetraMDTM Software (K003484)
{0}------------------------------------------------
K073169
pg. 1 of 2
Scimage
Sclmage, Inc. PicomEnterprise Software - 510(k) Premarket Notification
## 510(k) SUMMARY of Safety and Effectiveness
This following summary is provided as part of this Premarket Notification in compliance with and based on the format set forth in the Final Rule as published in the Federal Register, December 14, 1994. (See 21 CFR § 807.92)
(1) Submitters Name / Contact Information:
| Sclmage, Inc.<br>4916 El Camino Real<br>Suite 200<br>Los Altos, California 94022 | JAN - 9 20+ |
|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Sai P. Raya, Founder & CEO<br>Tel.: (650) 694-4858<br>Fax: (650) 694-4861<br>E-mail: sai_raya@scimage.com |
| Official Correspondent: | Regulatory Management Services<br>Gary J. Allsebrook - Consultant<br>16303 Panoramic Way<br>San Leandro, CA 94578<br>Tel/Fax: (510) 276-2648<br>E-Mail: regman10@comcast.net |
| Date prepared: | October 26, 2007 |
| (2) Name of device: | |
| Trade Name: | PicomEnterprise |
|----------------------|-------------------------------------------------------|
| Common Name: | Medical image workstation system, PACS. |
| Classification Name: | §892.2050 Picture archiving and communications system |
# (3) Identification of predicate device:
| Manufacturer | Device | 510(k) Number |
|---------------|--------------------------------------------|---------------|
| Scimage, Inc. | NetraTM Workstation and NetraMDTM Software | K960911 |
| Scimage, Inc. | NetraTM Workstation and NetraMDTM Software | K003484 |
| AMICAS | AMICAS Vision Series PACS 4.3 | K062477 |
#### (4) Description of the device:
The PicomEnterprise software is a multi-modality comprehensive two-, three- and four-dimensional image presentation software system intended for acceptance, transfer, display, storage and digital processing of medical images. The PicomEnterprise software combines reconstruction and display algorithms for medical image analysis in the familiar Microsoft Windows environment. PicomEnterprise offers full compliance with DICOM 3.0 standard that permit transfer of data from medical devices to storage server and then to other DICOM compliant devices.
{1}------------------------------------------------
K073169
pg. 2 of 2
# Scimage
## Sclmage, Inc. PicomEnterprise Software - 510(k) Premarket Notification
#### (5) A statement of the intended use of the device:
The PicomEnterprise software is intended for acceptance, transfer, display, storage and digital processing of medical images.
Its hardware components may include digitizers, workstations, communications devices, computers, video monitors, magnetic, optical disk, or other digital data storage devices and hardcopy devices.
The software components provide functions for performing operations related to image manipulation, enhancement, compression or quantification.
To support the diagnostic interpretation of Mammography studies, PicomEnterprise will display the full fidelity DICOM image in a non-compressed format. Images will be rendered with patient and clinical information clearly displayed as part of the DICOM Overlay as required by MQSA, on monitors cleared by FDA for use in Digital Mammography. Lossy compressed mammography images and digitized film screen images should not be used for the purpose of primary diagnosis. Mammographic images may only be interpreted using an FDA approved monitor that offers at least five megapixel resolution and meets other technical specifications reviewed and accepted by FDA
#### (6) Predicate Device Comparison:
The PicomEnterprise software is substantially equivalent to similar features in the predicate devices and has the same intended uses and technological characteristics. The new features included in the modified software do not affect the safety or effectiveness of the device.
The modified device complies with the following voluntary standards as "Special Controls" to ensure safe and effective use:
- . ACR/NEMA Digital Imaging and Communications in Medicine (DICOM) standard
- Joint Photographic Experts Group (JPEG) standard .
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design featuring three abstract human profiles facing to the right, with flowing lines representing movement or connection.
JAN - 9 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Scimage, Inc. % Mr. Gary J. Allsebrook Consultant Regulatory Management Services 16303 Panoramic Way SAN LEANDRO CA 94578-1116
Re: K073169
Trade/Device Name: PicomEnterprise Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 23, 2007 Received: November 9, 2007
#### Dear Mr. Allsebrook:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1796, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
ScImage Sclmage, Inc. PicomEnterprise - 510(k) Premarket Notification
# 2.0 FDA Indication for use form.
510(k) Number (if Known):
Device Name: PicomEnterprise
# Indications For Use:
The PicomEnterprise software is intended for acceptance, transfer, display, storage and digital processing of medical images.
Its hardware components may include digitizers, workstations, communications devices, computers, video monitors, magnetic, optical disk, or other digital data storage devices and hardcopy devices.
The software components provide functions for performing operations related to image manipulation, enhancement, compression or quantification.
To support the diagnostic interpretation of Mammography studies, PicomEnterprise will display the full fidelity DICOM image in a non-compressed format. Images will be rendered with patient and clinical information clearly displayed as part of the DICOM Overlay as required by MQSA, on monitors cleared by FDA for use in Digital Mammography. Lossy compressed mammography images and digitized film screen images should not be used for the purpose of primary diagnosis. Mammographic images may only be interpreted using an FDA approved monitor that offers at least five megapixel resolution and meets other technical specifications reviewed and accepted by FDA.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Hormal M. Whan
Division of Reproductive, Abdomin Radiological Devices 510(k) Number
Section 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.