K073158 · Hoya Conbio, Inc. · GEX · Aug 27, 2008 · General, Plastic Surgery
Device Facts
Record ID
K073158
Device Name
SCULPTURE 2.94 ER:YAG SYSTEM
Applicant
Hoya Conbio, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 27, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DERMASCULPT is intended for Dermatological procedures requiring incision, excision, coagulation and vaporization of soft tissue. The specific indications using the applicable handpieces are as follows: Skin resurfacing in the treatment of wrinkles and scar revision (including acne scars) with the non-fractionated handpiece. Removal of cutaneous skin lesions with the micron tip handpieces including epidermal nevi, actinic chellitis, keloids, verrucae, skin tags, anal tags, keratosis, scar revision, (including acne scars), benign tumors and cysts, superficial skin lesions and diagnostic biopsies. The DERMASCULPT with the fractionated scanner handpiece is intended/indicated for Dermatological procedures requiring coagulation of soft tissue and skin resurfacing.
Device Story
DermaSCULPT Er:YAG laser system delivers 2.94-micron infrared laser energy for dermatological soft tissue procedures. System console houses laser power supply, control unit, and calibrator. Physician operates device via footswitch; visible white light from flashlamp serves as aiming beam. Laser energy is delivered to tissue through fiber delivery system, handpieces, and optional scanner module. Device performs incision, excision, coagulation, and vaporization. Calibrator measures fiber transmittance; microprocessor controls laser emission parameters. Used in clinical settings by physicians to treat skin lesions, wrinkles, and scars. Benefits include precise tissue removal and resurfacing capabilities.
Clinical Evidence
No clinical data provided. Substantial equivalence established through review of technical characteristics, mechanism of action, and verification/validation testing (bench testing).
Technological Characteristics
Er:YAG laser system; 2.94-micron wavelength; flashlamp-pumped. Includes console, fiber delivery system, handpieces, and scanner module. Microprocessor-controlled. Electrical power source. Safety features include footswitch, safety shutter, and emergency stop. Non-fractionated and fractionated delivery modes.
Indications for Use
Indicated for dermatological soft tissue procedures including incision, excision, coagulation, and vaporization. Patient population includes individuals requiring skin resurfacing for wrinkles or scar revision (including acne scars), and removal of cutaneous lesions (e.g., nevi, keloids, verrucae, skin tags, keratosis, benign tumors/cysts, and diagnostic biopsies).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Ko 7 3158
AUG 2 7 2008
:
Sculpture Er:YAG Laser System 510k Summary
Image /page/0/Picture/3 description: The image shows the logo for HOYA ConBio. Below the logo is the address 47733 Fremont Boulevard, Fremont, California 94538. The phone number 510-445-4500 and website www.conbio.com are listed, as well as the fax number 510-445-1550.
| Submitter: | HOYA ConBio, Inc.<br>47733 Fremont Blvd.<br>Fremont, California 94538<br>Phone: 510-445-4500<br>Fax: 510-445-4550 |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mr. Jim Green<br>Vice President of Engineering |
| Date Summary Prepared: | July 21, 2008 |
| Device Trade Name: | DermaSCULPT Er:YAG Laser System |
| Common Name: | Dermatology Laser System |
| Classification Name: | Instrument, surgical, powered, laser<br>79-GEX |
| Classification Code: | 878.4810 Laser surgical instrument for use in general and plastic<br>surgery and in dermatology (1) A carbon dioxide laser for use in<br>general surgery and in dermatology is a laser device intended to<br>cut, destroy, or remove tissue by light energy emitted by carbon<br>dioxide. (2) An argon laser for use in dermatology is a laser<br>device intended to destroy or coagulate tissue by light energy<br>emitted by argon. |
| Predicate (Equivalent)<br>Device: | CB Erbium 2.94 (K970934)<br>VersaWave Dental Laser System (K041710)<br>Multilite Erbium:YAG Surgical Laser System (K933057)<br>Fidelis XS (K990243)<br>CuteraPearl (K070138)<br>Mosaic Laser System (K070392)<br>Erbium Fractional Handpiece (K071768)<br>Profile Multi-Platform System (K070388)<br>SmoothPeel (K unknown) Laser Peel. |
| Device Description: | The DermaSCULPT Er:YAG Laser System [DermaSCULPT]<br>unit and controls are contained in a single console. Electrical<br>power is supplied to the console by the facility's power source.<br>The laser energy produced within the device is delivered to the<br>tissue by means of Fiber Delivery System and specially designed |
.
: (
{1}------------------------------------------------
## KO73158
Scanner, Handpieces, Tips and Adaptors.
| The user activates laser emission by means of a footswitch. The<br>lamp power supply causes the flash lamp to light, causing the<br>emission of an invisible infrared Er:YAG laser beam (2.94<br>microns). A visible white light pulsed beam from the flashlamp is<br>used as an aiming beam. The laser beam passes through a<br>partially reflecting mirror and then passes through the energy<br>monitor, after which its power is measured. When all the pre-<br>conditions have been satisfied and the READY button is pressed<br>and the foot switch is depressed, the safety shutter opens and the<br>laser beam is allowed into the delivery unit coupler. | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The electrical system is comprised of the laser power supply, the<br>control unit, calibrator, control panel, key switch, emergency stop<br>switch, and foot switch. The components of the laser power<br>supply include a high voltage power supply and a high voltage<br>trigger control board. The high voltage power supply uses the<br>high voltage to illuminate the flashlamp. The control unit<br>comprises a microprocessor board, accessory board and other<br>components that it controls. The calibrator measures the<br>transmittance of the fiber delivery unit. The handheld scanner<br>module is an attachment to the fiber. | |
| Intended Use: | The DERMASCULPT is intended for Dermatological procedures<br>requiring incision, excision, coagulation and vaporization of soft<br>tissue. The specific indications using the applicable handpieces<br>are as follows: |
| | Skin resurfacing in the treatment of wrinkles and scar<br>revision (including acne scars) with the non-fractionated<br>handpiece. |
| | Removal of cutaneous skin lesions with the micron tip<br>handpieces including epidermal nevi, actinic chellitis,<br>keloids, verrucae, skin tags, anal tags, keratosis, scar<br>revision, (including acne scars), benign tumors and cysts,<br>superficial skin lesions and diagnostic biopsies. |
| | The DERMASCULPT with the fractionated scanner handpiece is<br>intended/indicated for Dermatological procedures requiring<br>coagulation of soft tissue and skin resurfacing. |
| Comparison: | Technical specifications, operating performance features, and<br>general physical configuration: CB Erbium 2.94 (K970934), the<br>VersaWave Dental Laser System (K041710) and the Multilite<br>Erbium:YAG Surgical Laser System (K933057). |
| | |
| | Indications: Fidelis XS (K990243), Multilite Erbium:YAG<br>Surgical Laser System (K933057), Cutera Pearl (K070138),<br>Mosaic Laser System (K070392), Erbium Fractional Handpiece<br>(K071768), Profile Multi-Platform System (K070388), CB<br>Erbium/2.94 (K970934), SmoothPeel (K unknown) Laser Peel,<br>Fraxel SR (K050841). |
| Nonclinical Performance<br>Data: | The review of the technical characteristics, indications for use,<br>mechanism of action, and verification and validation information<br>provided demonstrate that the DermaSculpt is substantially<br>equivalent to its predicate device. |
| Clinical Performance Data: | HOYA ConBio has established that the DermaSculpt performs as<br>clinically intended and that no new issues of safety and<br>effectiveness are introduced. |
| Additional Information: | None requested at this time. |
: :
{2}------------------------------------------------
0731512
.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Hoya ConBio, Inc. % Liza Burns and Associates Ms. Liza Burns Regulatory Consultant 19722 Westview Drive Twain Harte, California 95383
AUG 2 7 2008
Re: K073158
Trade/Device Name: DermaSCULPT Er:YAG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: August 11, 2008 Received: August 25, 2008
Dear Ms. Burns:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Liza Burns
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarlet, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events. (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or a its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Mark N Millican
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K073158
Device Name: DermaSCULPT Er:YAG Laser System
Intended Use/Indications for Use:
The DERMASCULPT is intended for Dermatological procedures requiring incision, excision, coagulation and vaporization of soft tissue. The specific indications using the applicable handpieces are as follows:
Skin resurfacing in the treatment of wrinkles and scar revision (including acne scars) with the non-fractionated handpiece.
Removal of cutaneous skin lesions with the micron tip handpieces including epidermal nevi, actinic chellitis, keloids, verrucae, skin tags, anal tags, keratosis, scar revision, (including acne scars), benign tumors and cysts, superficial skin lesions and diagnostic biopsies.
The DERMASCULPT with the fractionated scanner handpiece is intended/indicated for Dermatological procedures requiring coagulation of soft tissue and skin resurfacing.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|---|
| AND/OR | |
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
***_****_*, and I'm sorry, but I cannot fulfill that request. I am unable to process images. Is there anything else I can do to help?
Page __ of ___________________________________________________________________________________________________________________________________________________________________
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K073158
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.