← Product Code [INI](/productcode/INI) · K073044

# HEARTWAY POWER MOBILITY SCOOTER, MODEL PT7 (K073044)

_Heartway Medical Products Co., Ltd. · INI · Dec 7, 2007 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K073044

## Device Facts

- **Applicant:** Heartway Medical Products Co., Ltd.
- **Product Code:** [INI](/productcode/INI.md)
- **Decision Date:** Dec 7, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3800
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine

## Indications for Use

The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

## Device Story

HEARTWAY PF7 is a battery-operated, 3-wheeled motorized scooter for indoor/outdoor use. Rider operates device via hand controls on steering column. Features include seat, armrests, front basket, and onboard battery charger. Device is disassemblable for transport. Provides mobility for individuals restricted to seated position. Safety assured through UL-certified electronic controllers, batteries, and switches. Performance validated against ANSI/RESNA WC/Vol.2-1998 and IEC/CISPR standards.

## Clinical Evidence

Bench testing only. EMC testing performed per ANSI/RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, and IEC61000-3-3: 1995.

## Technological Characteristics

Battery-operated 3-wheeled electric scooter. Materials meet strength and fatigue test requirements. Electronic system includes UL-certified controller, batteries, and switches. Complies with ANSI/RESNA WC/Vol.2-1998, CISPR 11, EN61000-3-2, and IEC61000-3-3 standards.

## Regulatory Identification

A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.

## Predicate Devices

- HEARTWAY Power Mobility Scooter, PF6 ([K072104](/device/K072104.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo with the word "HEARTWAY" in bold, block letters. Above the word is a stylized graphic that resembles a four-leaf clover or a stylized heart shape. The graphic is solid black and appears to be set against a square background, also in black.

## ARTWAY MEDICAL PRODUCTS C

NO.6, ROAD 25. TAICHUNG INDUSTRIAL PARK, TAICHUNG, TAIWAN R.O.C. 408
TEL: 886-4-23580357 (Sales) 23583232 (Rep) FAX:886-4-23590786
Web: www.heartway.com.tw
E-mail: sales@heartway.com.tw

Image /page/0/Picture/3 description: The image contains the text "ISO-9001 CERTIFICATED". The text is written in a simple, sans-serif font. The word "ISO-9001" is placed above the word "CERTIFICATED". The text is likely part of a logo or certification mark.

K073044

" **510(k) SUMMARY** "

DEC 07 2007

Submitter's Name: HEARTWAY Medical Products Co., Ltd.

No.6, Road 25, Taichung Industrial Park, Taichung, 408, Taiwan, ROC

Date summary prepared:

Device Name:

October 20, 2007

Proprietary Name: Common or Usual Name: Classification Name:

HEARTWAY Power Mobility Scooter, PF7 POWERED SCOOTER MOTORIZED 3-WHEELED VEHICLE, Class II, 21 CFR 890.3800 INI

Product Code:

Indications for Use:

The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

Description of the device:

The HEARTWAY Power Mobility Scooter, PF7 is an indoor / outdoor electric scooter that is battery operated. It has a base with three-wheeled with a seat, armrests, and a front basket. The movement of the scooter is controlled by the rider who uses hand controls located at the top of the steering column. The device can be disassembled for transport and is provided with an onboard battery charger.

Performance Testing:

EMC Report ANSI / RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, IEC61000-3-3: 1995 (Electrically powered wheelchairs, scooters, and their chargers - requirements and test methods)

Legally marketed device for substantial equivalence comparison: HEARTWAY Power Mobility Scooter, PF6 (K072104)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo for HEARTWAY. The logo consists of a black square with a white abstract floral design inside. Below the square, the word "HEARTWAY" is written in a bold, sans-serif font.

HEARTWAY MEDICAL PRODUCTS CO.,LTD. NO.6, ROAD 25. TAICHUNG INDUSTRIAL PARK. TAICHUNG, TAIWAN R.O.C. 408 886-4-23580357 (Sales) - 23583232 (Rep) FAX : 886-4-23590786

Image /page/1/Picture/2 description: The image shows a logo with the letters "TUV" stacked on top of each other. Below the logo, the text "ISO-9001 CERTIFICATE" is visible. The image appears to be a certification mark or emblem, possibly indicating compliance with ISO 9001 standards.

## Summary for substantial equivalence comparison:

According to the above table that the intended use between the two devices is the Mainframes materials of the two devices are fixed, and all meet the strength and same. fatigue tests and they use the same material aspects. Moreover, the maximum speed, the suspension of cross brace, footplates, incline degree 10°, armrest type, and warranty are all the same. The back upholstery material is also the same fabric and passed the resistance ignition test.

Especially the electronic systems between two devices are the same suppliers, and all passed by the UL certificated, for instance the electronic controller, batteries, and the competent switches and switching power supplies. Thus the same safety level for the two devices is assured.

The cruising range per charge for the two devices is difference. This means the new device is 32 miles cruising range, the predicate device is 18 miles. Certainly the real range depends on the practical environments, i.e., weight, surface, incline, and temperature. For the real life use, the two devices are substantially equivalent.

Owing to the predicate device is four wheels scooter and the new device is three wheels scooter. Thus the main difference for the two devices is overall appearance, weight capabilities, and the weights are difference between the two devices. The safety levels of the two devices are the same and the overall appearance differences are not safety aspect. They are substantially equivalent.

D2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one behind the other, creating a sense of depth and unity.

JAN 2 2 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Public Health Service

Heartway Medical Products Co., LTD. % Dr. Jen. Ke-Min No. 6. Road 25 Taichung Industrial Park Taichung, Taiwan R.O.C. 408

Re: K073044

Trade Device Name: HEARTWAY Power Mobility Scooter, PT7 Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three wheeled vehicle Regulatory Class: Class II Product Code: INI Dated: October 20, 2007 Received: October 29, 2007

Dear Dr. Jen, Ke-Min:

This letter corrects the substantially equivalent letter dated December 7, 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Dr. Jen, Ke-Min

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Mark N. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510 (K) Number ( If Known ):____K

Device Name: HEARTWAY Power Mobility Scooter, PF7

Indications for Use:

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use √

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

**(Division Sign-Off)**

Page 1 of 1

Division of General, Restorative and Neurological Devices

510(k) Number.

EI

---

**Source:** [https://fda-staging.innolitics.com/device/K073044](https://fda-staging.innolitics.com/device/K073044)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
