K073017 · Stryker Instruments, Instruments Div. · FYA · Nov 30, 2007 · General, Plastic Surgery
Device Facts
Record ID
K073017
Device Name
STRYKER STRI-SHIELD FLYTE TOGAS
Applicant
Stryker Instruments, Instruments Div.
Product Code
FYA · General, Plastic Surgery
Decision Date
Nov 30, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Stryker Steri-Shield Flyte Toga, Pullover; Stryker Steri-Shield Flyte Toga, Zippered; Stryker Steri-Shield Flyte Toga with Peel-away Lens, Zippered; and Stryker Steri-Shield Flyte Togas with UV Lens, Zippered are components of a personal protection system and are intended to protect the patient, health care personnel and operating room personnel against contamination, exposure of infectious bodily fluids, the transfer of microorganisms and particulate material.
Device Story
Steri-Shield Flyte Togas are surgical gowns worn over Stryker Flyte Helmets; provide barrier protection in operating room environments. Togas feature integrated hoods with attached lenses (including peel-away or UV options) and gown sections covering front, back, and arms. Available in pullover or zippered configurations. Barrier performance meets AAMI PB70:2003 standards; Level 4 critical zones resist liquid and viral penetration (tested per ASTM F1671:2003); Level 1 zones resist liquid penetration (tested per AATCC 42:2000). Protects patients and staff from fluid/microorganism transfer. Used by surgical teams during procedures.
Clinical Evidence
Bench testing only. Barrier performance validated against AAMI PB70:2003 guidelines. Viral penetration resistance tested per ASTM F1671:2003 (bacteriophage Phi-X174) with 4% AQL. Liquid penetration resistance tested per AATCC 42:2000 with 4% AQL.
Technological Characteristics
Materials: Polypropylene or polyester non-woven fabric, copolymer polyester film, 35g SMMS polypropylene. Barrier performance: AAMI PB70:2003 Level 1 and Level 4. Testing standards: ASTM F1671:2003, AATCC 42:2000. Form factor: Pullover or zippered gown with integrated hood and lens. Sizes: S/M to 3XL.
Indications for Use
Indicated for use by patients, health care personnel, and operating room personnel as a personal protection system component to prevent contamination, exposure to infectious bodily fluids, and transfer of microorganisms and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Stryker T6 Hoods and Togas (K0063005)
Submission Summary (Full Text)
{0}------------------------------------------------
K07 3017
4100 E. Milham Avenue
Kalamazoo, MI 49001
t: 269 323 7700 ff: 800 965 6505
www.stryker.com
NOV 3 0 2007
Image /page/0/Picture/3 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and there is a registered trademark symbol to the upper right of the letter "r". The word is in black and the background is white.
Instruments
## 510(k) Summary
| Device Sponsor: | Stryker Instruments<br>4100 East Milham Avenue<br>Kalamazoo, Michigan 49001<br>(p) 269-323-7700<br>(f) 269-324-5412 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration No.: | 1811755 |
| Trade Name: | Stryker Steri-Shield Flyte Togas |
| Classification: | Sterile Surgical Gowns |
| Equivalent to: | K0063005-Stryker T6 Hoods and Togas |
| Device Description: | Device History |
| | Summary of Stryker Steri-Shield Flyte Togas |
| | The scope of this Special 510k material modification is limited<br>to material changes to the gown. The toga has a hood and gown<br>section. The hood covers the user's head. A lens is attached to the<br>front of the hood. The gown of the toga covers the user's front,<br>back and arms. The Stryker Steri-Shield Flyte Togas are tested to<br>meet applicable AAMI PB70:2003 standards. The AAMI standard<br>does not cover apparel for the head, face, and eyes. Therefore, the<br>hoods and lens (components of the togas) are exempt from<br>classification under the AAMI PB70:2003 standard. |
| | Stryker Steri-Shield Flyte Toga, Pullover |
| | The Stryker Steri-Shield Flyte Toga, Pullover is intended to be<br>worn over any Stryker Flyte Helmet. The Stryker Steri-Shield<br>Flyte Pullover togas provide Level 4 and Level 1 barrier protection<br>for the critical zones as set froth in the accompanying drafted IFUs<br>(Attachment 8 and as classified under the AAMI (Association for<br>the Advancement of Medical Instrumentation) guidelines for<br>barrier performance. This garment was tested for resistance to<br>bacteriophage Phi-X174 in accordance with ASTM F1671:2003 |
{1}------------------------------------------------
KITT
per the AAMI Barrier Classification System and demonstrate a passing result with an AQL of 4% under Procedure A. Level 1 zones resist liquid penetration per AATCC (American Association of Textile Chemist and Colorist) 42:2000 with an AQL (acceptable quality level) of 4%. Level 4 critical zones resist liquid and viral penetration. The Stryker Steri-Shield Flyte Toga Zippered garment was tested for resistance to bacteriophage Phi-X174 in accordance with ASTM (American Society for Testing and Materials) F1671:2003 with an AQL of 4% under Procedure A. The barrier material for the Stryker Steri-Shield Flyte Toga Pullover is constructed of layers of polypropylene or polyester non-woven fabric and copolymer polyester film. The Stryker Steri-Shield Flyte Toga, Pullover will be available in a variety of sizes ranging from Small/Medium thru 3X-Large sizes.
## Stryker Steri-Shield Flyte Toga, Zippered: Stryker Steri-Shield Flyte Toga with Peel-away Lens, Zippered; Stryker Steri-Shield Flyte Toga with UV Lens, Zippered.
The Stryker Steri-Shield Flyte Toga, Zippered; Stryker Steri-Shield Flyte Toga with Peel-away Lens, Zippered; and Stryker Steri-Shield Flyte Togas with UV Lens. Zippered are intended to be worn over any Stryker Flyte Helmet. The Stryker Steri-Shield Flyte Togas provide protection as specified by the AAMI (Association for the Advancement of Medical Instrumentation) Barrier Classification System. Level 1 zones resist liquid penetration per AATCC (American Association of Textile Chemist and Colorist) 42:2000 with an AQL (acceptable quality level) of 4%. Level 4 critical zones resist liquid and viral penetration. The Stryker Steri-Shield Flyte Toga Zippered garment was tested for resistance to bacteriophage Phi-X174 in accordance with ASTM (American Society for Testing and Materials) F1671:2003 with an AQL of 4% under Procedure A. The Level 4 barrier material is constructed of layers of polypropylene or polyester non-woven fabric and copolymer polyester film. The Level 1 material on the back of the zippered togas is constructed of 35 gram SMMS polypropylene. The Stryker Steri-Shield Flyte Toga. Zippered will be available in a variety of sizes ranging from Small/Medium thru 3X-Large sizes.
Indications for Use: The Stryker Steri-Shield Flyte Togas are components of a personal protection system and are intended to protect the patient, health care personnel and operating room personnel against
{2}------------------------------------------------
ー…( 다.
각자
| | | contamination, exposure of infectious bodily fluids, the transfer of<br>microorganisms and particulate material. |
|--|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Substantial Equivalence<br>(SE) Rational: | The Stryker Steri-Shield Flyte Togas are substantially equivalent to<br>devices in commercial distribution. |
| | | Stryker Steri-Shield Flyte Togas have an equivalent intended use,<br>patient contact materials, operating principles and physical<br>specifications as compared to predicate devices. |
| | Safety and Effectiveness: | Based upon the comparison to the predicate devices, the Stryke<br>Steri-Shield Flyte Togas are substantially equivalent to legally<br>marketed devices. The Stryker Steri-Shield Flyte Togas do not<br>raise any new safety or efficacy concerns. |
| | Submitted by: | Paulette D. Johnson<br>Regulatory Analyst<br>Stryker Instruments<br>4100 East Milham Avenue<br>Kalamazoo, MI 49009<br>p: 269-323-7700, ext 3084<br>f: 269-324-5412 |
tourities Signature
Date Submitted:
10-18-07
---
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 3 0 2007
Ms. Paulette D. Johnson Regulatory Analyst Stryker Instruments 4100 East Milham Avenue Kalamazoo, Michigan 49001
Re: K073017
Trade/Device Name: Stryker Steri-Shield Flyte Togas Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FYA Dated: November 19, 2007 Received: November 20, 2007
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Johnson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clue
Chia-Lin Pi, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K073017
Device Name: Stryker Steri-Shield Flyte Togas (Stryker Steri-Shield Flyte Toga, Pullover; Stryker Steri-Shield Flyte Toga, Zippered; Stryker Steri-Shield Flyte Toga with Peel-away Lens, Zippered; and Stryker Steri-Shield Flyte Togas with UV Lens, Zippered)
Indications for Use:
The Stryker Steri-Shield Flyte Toga, Pullover; Stryker Steri-Shield Flyte Toga, Zippered; Stryker Steri-Shield Flyte Toga with Peel-away Lens, Zippered; and Stryker Steri-Shield Flyte Togas with UV Lens, Zippered are components of a personal protection system and are intended to protect the patient, health care personnel and operating room personnel against contamination, exposure of infectious bodily fluids, the transfer of microorganisms and particulate material.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
S. H. Murphy, Jr.
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.