K073000 · Clareblend, Inc. · GEX · Oct 1, 2008 · General, Plastic Surgery
Device Facts
Record ID
K073000
Device Name
LUMAPROBE
Applicant
Clareblend, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 1, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LumaProbe is intended to provide light to the body. LumaProbe (Red-631) is generally indicated for the treatment of superficial, benign vascular, and pigmented lesions. LumaProbe (Blue-415) is generally indicated for the treatment of dermatological conditions and specially indicated for the treatment of mild to moderate inflammatory acne vulgaris. LumaProbe (IR-830) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
Device Story
LumaProbe is a light therapy system utilizing high-spectral-purity LEDs to deliver specific wavelengths of light to the skin. The system consists of a control unit with power supplies and attachable hand probes (Red 631nm, Blue 415nm, IR 830nm). The device provides uniform, hot-spot-free illumination. Operators move the probes over the skin surface to deliver light. Probes feature an integrated STOP button for immediate power removal. The device is intended for clinical use to treat dermatological conditions, vascular/pigmented lesions, or provide temporary pain relief and increased local circulation. Healthcare providers use the device to deliver targeted light therapy, which may benefit patients by reducing acne inflammation, treating skin lesions, or alleviating musculoskeletal discomfort.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use comparisons to predicate devices.
Technological Characteristics
LED-based light therapy system. Wavelengths: Red (631+/-4nm), Blue (415+/-5nm), IR (830nm). System includes power supply, control unit, and attachable hand probes. Features manual STOP button on probes. Standalone device.
Indications for Use
Indicated for patients with superficial, benign vascular and pigmented lesions (Red-631); mild to moderate inflammatory acne vulgaris and dermatological conditions (Blue-415); and minor muscle/joint pain, arthritis, muscle spasm, stiffness, or need for increased local blood circulation (IR-830).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K073600
OCT 01 2008
## FDA 510K Summary of Safety and Effectiveness for LumaProbe
- 1. General Information
| Submitter: | Clareblend Inc.<br>3555 Airway Dr.<br>Suite 307<br>Reno, Nevada 89511<br>800-334-7126<br>775-332-3852 Fax |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | C/O Jill Creasy<br>Aesthetica-Tech<br>675 Pine Street<br>Elgin, IL 60123<br>847-429-9631<br>847-429-9633 Fax<br>contact@aestheticatech.com |
| Summary Preparation Date: | October 22, 2007 |
| Names | |
| Device Name: | LumaProbe Model # 7002-01 with hand probes (Models: 7201-415,<br>7204-631, and 7205-830 |
| Classification Name: | Laser instrument, surgical, powered device; GEX, ILY<br>FDA Class II category <p>Although this device is not a laser, the specifications developer feels this is the closest applicable classification name.</p> |
#### 3. Predicate Device
2.
LumaProbe is substantially equivalent to the Omnilux Blue (K030883), Omnilux Revive (K030426), Omnilux Plus (K043317)
#### 4. Device Description
The LumaProbe is a device that utilizes Light Emitting Diodes to provide LED light to the body. The LumaProbe Red, IR and Blue are a visible light source of high spectral purity. They provide uniform or "hot-spot" free illumination. The output is pre-tuned to one wavelength with a narrow spectral bandwidth. The output wavelength of Red is 631 +/- 4nm, IR is 830 +/- nm and Blue is 415 +/- 5nm. The LumaProbe unit contains the power supplies and the control unit; it has the capabilities to power and sequence up to 2 attachable probes depending on the product model, that deliver the light to the skin as they are moved over the skin surface. The STOP button directly on the probes allows the user to immediately remove all power to the probes.
#### 5. Indications for Use:
LumaProbe (Red-631) is generally indicated for the treatment of superficial, benign vascular and pigmented lesions
LumaProbe (Blue-415) is generally indicated for the treatment of dermatological conditions and specially indicated for the treatment of mild to moderate inflammatory acne vulgaris.
LumaProbe (IR-830) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
{1}------------------------------------------------
# Ko 73000
#### Comparison of Technological Differences: 6.
The intended use and technological characteristics of the LumaProbe system are virtually identical to the intended use and technological characteristics of the listed equivalent devices. Any differences between the LumaProbe and the equivalent devices have no significant influence on safety or effectiveness of the LumaProbe product.
#### 7. Conclusions
Based upon an analysis of the overall performance characteristics for the LumaProbe, Clareblend, Inc. believes that no significant differences exist between this system and the predicate systems quoted, therefore, the LumaProbe Light Therapy device does not impose any new safety or effectiveness concerns.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 0 1 2008
Clareblend, Inc. % Aesthetica-Tech Ms. Jill Creasy Medical Device Consultant 675 Pine Street Elgin, Illinois 60123
Re: K073000
Trade/Device Name: LumaProbe Model # 7002-01 with hand probes (Models: 7201-415, 7204-631, and 7205-830) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX, ILY Dated: September 15, 2008 Received: September 18, 2008
Dear Ms. Creasy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Jill Creasy
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
## 510(k) Number (if known): K_073000
Device Name: LumaProbe Model # 7002-01 with hand probes (Models: 7201-415, 7204-631, and 7205-830
## Indications for Use:
The LumaProbe is intended to provide light to the body.
- LumaProbe (Red-631) is generally indicated for the treatment of superficial, benign vascular, and pigmented A. lesions
- LumaProbe (Blue-415) is generally indicated for the treatment of dermatological conditions and specially в. indicated for the treatment of mild to moderate inflammatory acne vulgaris.
- LumaProbe (R-830) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis C. and muscle spasm, relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
Prescription Use X AND/OR (21 CFR Part 801 Subpart D)
Over-The-Counter Use (21 CFR Part 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*Milke*
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
K073000
510(k) Number
Page 1 of 1 -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.