← Product Code [LFL](/productcode/LFL) · K072904

# SONOCA 180/195 WOUND CARE SYSTEM (K072904)

_Soering GmbH · LFL · May 21, 2008 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072904

## Device Facts

- **Applicant:** Soering GmbH
- **Product Code:** [LFL](/productcode/LFL.md)
- **Decision Date:** May 21, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

The SONOCA 180/185 Wound Care System is an instrument intended for selected ultrasound dissection and fragmentation of tissue, wound debridement (acute and chronic wounds, burns, diseased or necrotic tissue) and cleansing irrigation of the site for removal of debris, exudates, fragments, and other matter.

## Device Story

Sonoca 180/185 is an ultrasonic wound care system; uses low-frequency power ultrasound (25 kHz) to treat wound tissue. System consists of an ultrasonic generator and a handpiece with a piezoelectric crystal (PZT transducer) that converts electrical energy into mechanical vibrations. Device operates by applying ultrasound via an irrigation solution (0.9% NaCl or Ringer solution) to the wound surface. Mechanism of action is cavitation; formation and implosion of bubbles in the irrigation fluid create turbulence and currents that loosen necrotic tissue and fibrin layers while flushing away debris. Used in clinical settings for wound debridement and tissue fragmentation. Healthcare providers move the receptacle over the wound to perform cleansing. Benefits include gentle removal of necrotic tissue and debris from acute and chronic wounds or burns.

## Clinical Evidence

Bench testing only. The device complies with electrical safety and electromagnetic compatibility standards including IEC 950, CISPR 22, IEC-801-2/3, IEEE 1003.1, IEC 601-1, IEC 601-2-2, and DIN EN 61847.

## Technological Characteristics

Piezoelectric (PZT) transducer; 25 kHz ultrasonic frequency; irrigation-based delivery (0.9% NaCl or Ringer solution); electrical safety compliance per IEC 601-1, IEC 601-2-2, and IEC 950; electromagnetic compatibility per CISPR 22 and IEC-801-2/3.

## Predicate Devices

- SORING GMBH SONOCA 180/190 ([K050776](/device/K050776.md))
- AR1000 ULTRASONIC WOUND THERAPY SYSTEM ([K062544](/device/K062544.md))
- AUSS-6 ULTRASONIC SURGICAL ASPIRATOR SYSTEM ([K012753](/device/K012753.md))
- SORING GMBH SONOCA 300 ([K992026](/device/K992026.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K072904

## 510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

Date Prepared:

Many 8, 2008

Submitter's Information:

Predicate Device:

Mr. Jan Schueller-lwersen, Manager, Regulatory & Quality Soring GmbH Medizintechnik Justus-v.Liebig 2 25451 Quickborn Germany

Telephone: 49 4106-6100-0 Fax: 49 4106-6100-0 Email: info@soering.com

Trade Name, Common Name, Classification:

Trade Name: Common name: Classification name:

195 Sonoca™ 180/195 Wound Care System Ultrasound wound cleaner Low energy ultrasound wound cleaner, Instrument, ultrasonic Surgical

instrument, ultrasonic

surgical

K050776

Device instrument, ultrasonic wound cleaner, Classification surgical ultrasound Name 510(k) Number K062544 K012753

| Device Name                             | SORING GMBH<br>SONOCA 180/190 | AR1000 ULTRASONIC<br>WOUND THERAPY<br>SYSTEM | AUSS-6 ULTRASONIC<br>SURGICAL<br>ASPIRATOR SYSTEM | SORING GMBH<br>SONOCA 300     |
|-----------------------------------------|-------------------------------|----------------------------------------------|---------------------------------------------------|-------------------------------|
| Applicant                               | SORING GMBH<br>MEDIZINTECHNIK | AROBELLA MEDICAL,<br>LLC                     | MISONIX, INC.                                     | SORING GMBH<br>MEDIZINTECHNIK |
| Classification<br>Product Code          | LFL                           | NRB, FQH, LFL                                | LFL                                               | LFL<br>BTA                    |
| Decision Date                           | 11/13/2001                    | 01/03/2007                                   | 06/06/2005                                        | 09/23/1999                    |
| Classification<br>Advisory<br>Committee | General & Plastic<br>Surgery  | General & Plastic<br>Surgery                 | General & Plastic<br>Surgery                      | General & Plastic<br>Surgery  |

Device Description:

Ultrasonic-Assisted Wound Treatment using the Sonoca 180 / 185 applies low-frequency power ultrasound in conjunction with an irrigation solution via a moving receptacle directly to the wound tissue. An ultrasonic generator transfers electric energy to a receptacle where a high-precision piezoelectric crystal system (PZT transducer) transforms this energy into mechanic vibrations: 25,000 vibrations / second (= 25 kHz).

MAY 2 1 2008

pump, portable,

aspiration (manual or

powered)

K992026

{1}------------------------------------------------

K072904 Page 2 of 2

## 510(k) Summary of Safety and Effectiveness

The receptacle is continuously moved over the wound surface. This allows gentle loosening of necrotic tissue and fibrin layers and flushes them away with the irrigation solution (0.9 % NaCl or Ringer solution). The liquid is used for transmitting the ultrasound as well as for wound irrigation. Cavitation is the underlying mode of action of ultrasound: the formation and disintegration of cavities (bubbles) in liquids due to pressure fluctuations, i.e. formation and subsequent implosion of these bubbles causes turbulences and currents on the wound surface, which help to loosen necrotic tissue and fibrin layers.

## Indications for Use:

The SONOCA 180/185 Wound Care System is an instrument intended for selected ultrasound dissection and fragmentation of tissue, wound debridement (acute and chronic wounds, burns, diseased or necrotic tissue) and cleansing irrigation of the site for removal of debris, exudates, fragments, and other matter.

#### Performance Data:

The subject device complies with IEC 950 - Safety of Information Technology Equipment, CISPR 22, class A - Electromagnetic Compatibility, IEC-801-2, IEC-801-3 - Electromagnetic Compatibility, IEEE 1003.1 - General Electrical Safety for medical devices, IEC 601-1 -Electrical Safety for medical devices using RF-power, IEC 601-2-2 - Ultrasonic surgical devices, DIN EN 61847

#### Conclusion:

Similar to the predicate devices, the Sonoca 180/185 does not control any life sustaining functions or services.

The new device and the predicate devices are substantially equivalent in the areas of general function, and intended use. Any differences between the devices will not affect safety or efficacy.

Based on the information supplied in this 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

2 2009 JUL

Soring GmbH % Delta Quality Consulting Mr. Carl Thomas 1600 Manchester Way Corinth, Texas 76210

Re: K072904 Trade/Device Name: Sonoca 180/185 Wound Care System Regulatory Class: Unclassified Product Code: LFL Dated: May 12, 2008 Received: May 19, 2008

Dear Mr. Thomas:

This letter corrects our substantially equivalent letter of May 21, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page - Mr. Carl Thomas

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

for D.B. D

D.e

Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

# Indications for Use

K072904 510(k) Number (if known):

Device Name:

Soring GmbH, SONOCA 180/185 Wound Care System

Indications for Use:

The SONOCA 180/185 is an instrument intended for selected ultrasound dissection and fragmentation of tissue, wound debridement (acute and chronic wounds, burns, diseased or necrotic tissue) and cleansing irrigation of the site for removal of debris, exudates, fragments, and other matter.

Prescription Use _ × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(Ni

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark H. Milkman

Sign-( Division of General, Restorative, and Neurological Devices

**510(k) Number**

---

**Source:** [https://fda-staging.innolitics.com/device/K072904](https://fda-staging.innolitics.com/device/K072904)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
