K072817 · Smith & Nephew, Inc. · LPH · May 15, 2008 · Orthopedic
Device Facts
Record ID
K072817
Device Name
ECHELON TITANIUM HIP SYSTEM
Applicant
Smith & Nephew, Inc.
Product Code
LPH · Orthopedic
Decision Date
May 15, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Echelon Titanium Hip System is indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips clamaged as a result of Irauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; old, remote osteomyelitis with an extended drainage-free period, in which case, the patient should be warned of an above normal danger of infection postoperatively, treatments of nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteolorny, or Girdlestone resection; fracture-dislocation of the hip; and correction of deformity. The Echelon Titanium Hip System is intended for single use only and implanted without bone cement.
Device Story
Echelon Titanium Hip System; uncemented prosthetic femoral stem for total hip replacement. Device implanted by orthopedic surgeons in clinical/hospital settings. System features primary and revision stem options with 10/12 neck taper for compatibility with 10/12 Taper Femoral Heads. Design and material composition derived from Anthology Hip Stem (K052792). Intended to replace damaged hip joint components; restores joint function; alleviates pain; improves patient mobility. Single-use device.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by technological comparison and design similarities to legally marketed predicate devices.
Technological Characteristics
Uncemented femoral stem prosthesis. Materials: Titanium. Features porous coating and surface finish identical to Anthology Hip Stem (K052792). Includes 10/12 neck taper. Designed for primary and revision total hip arthroplasty. Single-use.
Indications for Use
Indicated for patients requiring primary or revision hip surgery due to trauma, noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant), inflammatory degenerative joint disease (rheumatoid arthritis, secondary arthritis, congenital dysplasia), remote osteomyelitis, nonunion, femoral neck/trochanteric fractures, endoprosthesis, femoral osteotomy, Girdlestone resection, fracture-dislocation, or deformity correction. Contraindicated in patients with active infection.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Hip Joint Prosthesis (Richards Modular Hip System) (K924100)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY SMITH & NEPHEW ECHELON TITANIUM HIP SYSTEM
MAY 1 5 2008
| SUBMITTER'S NAME: | Smith & Nephew, Inc., Orthopaedic Division |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER'S ADDRESS: | 1450 Brooks Road, Memphis, TN 38116 |
| SUBMITTER'S TELEPHONE NUMBER: | 901-399-6055 |
| CONTACT PERSON: | Marlon D. Ridley |
| DATE SUMMARY PREPARED: | January 16, 2008 |
| TRADE OR PROPRIETARY DEVICE NAME: | Echelon Titanium Hip System |
| COMMON OR USUAL NAME: | Prosthetic Hip Joint System-Porous Femoral Stem |
| CLASSIFICATION NAME AND REFERENCE: | 21 CFR 888.3358 Hip Joint Metal/Polymer/Metal Semi-<br>Constrained Porous-Coated Uncemented Prosthesis - Class II<br>21 CFR 888.3350 Hip Joint Metal/Polymer Semi-Constrained<br>Cemented Prosthesis - Class II<br>21 CFR 888.3353 Hip Joint Metal/Ceramic/Polymer Semi-<br>Constrained - Class II<br>21 CFR 878.3300 Surgical Mesh - Class II |
| PRODUCT CODE AND PANEL CODE: | Orthopaedics 87 / LPH / MBL / MEH / JDI / JDJ |
## PRC
### DEVICE INFORMATION:
### प्र INTENDED USE:
The Echelon Titanium Hip System is indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips clamaged as a result of Irauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; old, remote osteomyelitis with an extended drainage-free period, in which case, the patient should be warned of an above normal danger of infection postoperatively, treatments of nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteolorny, or Girdlestone resection; fracture-dislocation of the hip; and correction of deformity.
The Echelon Titanium Hip System is intended for single use only and implanted without bone cement.
### ന് DEVICE DESCRIPTION:
The body of the Echelon Titanium Hip System femoral stems is similar in shape to the Anthology Hip Stem as previously cleared under premarket notification K052792. The material, porous coaling and surface finish of the Echelon Titanium Hip System fernoral stems are identical to the Anthology Hip Stern (K052792). The Echelon Titanium Hip System femoral stems are offered in primary and revision options featuring a 10/12 neck taper for use with the 10/12 Taper Femoral Heads.
### SUBSTANTIAL EQUIVALENCE INFORMATION: ﻝ
The substantial equivalence of the Echelon Thanium Hip System is supported by II's Similarities in design features, overall indications, and material composition to table of legally marketed devices for total hip replacement.
{1}------------------------------------------------
| COMPANY | PREDICATE DEVICE NAME | 510(κ) | CLEARANCE<br>DATE |
|------------------------------------|--------------------------------------------------------------|---------|-------------------|
| Smith & Nephew, Inc. | ANTHOLOGY® Hip Stems | K052792 | 10/07/2005 |
| Wright Medical<br>Technology, Inc. | Profemur Renaissance Total Hip System | K051995 | 08/22/2005 |
| Ortho Development<br>Corporation | Encompass Cemented Hip Stem and 10/12 Cobalt<br>Chrome Heads | K050637 | 09/23/2005 |
| Smith & Nephew, Inc. | Global Taper Tapered Hip System | K963509 | 01/27/1997 |
| Smith & Nephew, Inc. | Hip Joint Prosthesis (Richards Modular Hip System) | K924100 | 04/02/1993 |
### SUMMARY OF TECHNOLOGICAL COMPARISON: D.
The intended use, design, and materials of the Echelon Titanium Hip System are similar to the legally marketed devices listed above. The Echelon Titanium Hip System components have the same technological characteristics as each of the legally marketed devices listed above as indicated for total hip replacement.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and tail. The eagle's head is facing left. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Smith & Nephew, Inc. % Mr. Jason Sells Project Manager, Regulatory Affairs 1450 Brooks Road Memphis, TN 38116
K072817 Re: Trade/Device Name: Echelon Titanium Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, MBL, JDI, MEH, JDJ Dated: May 12, 2008 Received: May 13, 2008
Dear Mr. Sells:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
MAY 15 2008
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Jason Sells
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Mullison
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
ිංගිකරන පිළිබඳ විය Forsity & Presidents IFK 1 3 5 3 6 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 Marshap Res. Tra Raker 80 St A 1000 2019 12 22 23 23 www.analib naposuwoom
- Smith&nephew
# Indications for Use
一次不可是
# 510(k) Number (if known):
Device Name: Echelon Titanium Hip System
# Indications for Use:
The Echelon Titanium Hip System is indicated for individuals undergoing primary and revision surgery where other treatments or devices have falled in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NID)D) or any of its composite diagnoses of osteoarthrilis, avascular necrosis, traumatic anhrifis slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
Hip components are also indicated for inflammatory degenerative joint disease including theumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; old, remote osteomyelitis with an extended drainage-free period, which case, the patient should be warned of an above normal danger of infection postoperatively; treatments of nonunion, femoral neck fracture and trochariteric fractures of the proximal fernur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osfeotomy, or Girdlestone resection; fracture distocation of the hip; and correction of deformity.
The Echelon Titanium Hip System is intended for single use only and implanted vittho bone cement
Prescription Use IPart 21 CFR 801 Subpart Di
Over-The-Counter Use AND/OR 21 CFR 807 Subpan CI
IPLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
DRH, Office of Device Evaluation (ODE) Concurrance of
Division Sign-Off Division of General, Restorative and Neurological Devices
510(k) Number K072817
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.