K072812 · Philips Medical Systems · MKJ · Feb 26, 2008 · Cardiovascular
Device Facts
Record ID
K072812
Device Name
PHILIPS HEARTSTART SMARTPADS III
Applicant
Philips Medical Systems
Product Code
MKJ · Cardiovascular
Decision Date
Feb 26, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The Philips HeartStart SmartPads III is indicated for external defibrillation, pacing, monitoring and cardioversion. The SmartPads III is intended for adults and children over 8 years of age or greater than 55 pounds (25kg) with all compatible defibrillators. The SmartPads III is also intended for children of all ages when used with a Philips Medical System model FRx equipped with an Infant/Child Key or with other compatible defibrillators which incorporate adjustable defibrillation energy functionality. The device is non-sterile and single-use only.
Device Story
Disposable, self-adhesive, single-use multifunction defibrillation pads; used for external defibrillation, pacing, cardioversion, and monitoring. Inputs: electrical signals from patient body; outputs: electrical therapy delivery and ECG monitoring signals to compatible defibrillators. Components: foam backing, laminated metallic substrate, conductive adhesive gel, cabling, molded connector. Used in clinical or emergency settings by healthcare providers or trained responders. Pads interface with defibrillator units to deliver energy or monitor cardiac rhythm; facilitates life-saving intervention for cardiac arrest or arrhythmia management.
Clinical Evidence
Bench testing only; no clinical data presented. Performance demonstrated to meet predefined criteria and equivalence to predicate.
Technological Characteristics
Disposable, self-adhesive, single-use pads. Construction: foam backing, laminated metallic substrate, conductive adhesive gel, cabling, molded connector. Non-sterile. Compatible with Philips defibrillators and adapters for other systems.
Indications for Use
Indicated for external defibrillation, pacing, monitoring, and cardioversion. Intended for adults and children >8 years or >55 lbs (25kg) with compatible defibrillators; indicated for children of all ages when used with Philips FRx with Infant/Child Key or other compatible defibrillators with adjustable energy functionality.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Predicate Devices
Philips Medical Systems HeartStart SmartPads II Multifunction Defibrillation Pads (K050004)
Submission Summary (Full Text)
{0}------------------------------------------------
#### FES 26
# 510(k) Summary Philips HeartStart SmartPads III Multifunction Defibrillation Pads
### 1. 510(k) Submission Number: K072812
- 2. Date Summary Prepared: February 22, 2008
### 3. Submitter's Name and Address:
Philips Medical Systems 2301 Fifth Avenue, Suite 200 Seattle, WA 98121
## 4. Contact Person:
Larry Walker Philips Medical Systems 2301 Fifth Avenue, Suite 200 Seattle WA 98121 Telephone: (206) 664-5000 Fax: (206) 664-5001
### 5. Device Name:
| Proprietary Name: | Philips HeartStart SmartPads III Multifunction Defibrillation Pads |
|----------------------|----------------------------------------------------------------------------------------------------------------|
| Common Name: | Multifunction Defibrillation Pad |
| Classification Name: | (MKJ) Automated External Defibrillator (Class III) (MLN) Multifunction Electrocardiograph Electrode (Class II) |
### 6. Predicate Device
The legally marketed device, to which Philips Medical Systems claims equivalence, is the Philips Medical Systems HeartStart SmartPads II Multifunction Defibrillation Pads, cleared as a part of the Philips Medical System HeartStart model FRx Automated External Defibrillator (K050004).
The design and the intended use of the HeartStart SmartPad III is substantially equivalent in safety and performance to the device named above.
CONFIDENTIAL
{1}------------------------------------------------
# 7. Device Description Summary
The Philips Medical Systems HeartStart SmartPads III is a disposable, selfadhesive, single use disposable defibrillator pad set, designed for external defibrillation, external pacing, cardioversion and monitoring with current and future Philips Medical Systems defibrillators. The SmartPads III are also compatible with other defibrillators through currently marketed Philips adapters.
The HeartStart SmartPad III consists of foam backing, laminated metallic substrate, conductive adhesive gel, cabling and molded connector. The SmartPad III are packaged in pairs within a water-vapor proof, heat sealed, non-transparent aluminized pouches.
## 8. Intended Use
The Philips HeartStart SmartPads III is indicated for external defibrillation, pacing, monitoring and cardioversion.
The SmartPads III is intended for adults and children over 8 years of age or greater than 55 pounds (25kg) with all compatible defibrillators. The SmartPads III is also indicated for children of all ages when used with a Philips Medical System model FRx equipped with an Infant/Child Key or with other compatible defibrillators which incorporate adjustable defibrillation energy functionality
## 9. Indications for use
The Philips HeartStart SmartPads III is indicated for external defibrillation, pacing, monitoring and cardioversion.
The SmartPads III is intended for adults and children over 8 years of age or greater than 55 pounds (25kg) with all compatible defibrillators. The SmartPads III is also intended for children of all ages when used with a Philips Medical System model FRx equipped with an Infant/Child Key or with other compatible defibrillators which incorporate adjustable defibrillation energy functionality.
The device is non-sterile and single-use only.
# 10. Comparison of Technology Characteristics
The Phillips HeartStart SmartPads III employs the same fundamental scientific technology as the currently available Philips HeartStart SmartPad II.
{2}------------------------------------------------
# 11. Data Used in Determination of Substantial Equivalence
Testing demonstrated the Philips HeartStart SmartPads III perform in a manner substantially equivalent to the predicate SmartPads II and is acceptable for its indicated use and meets the predefined criteria.
Testing demonstrated the performance of the SmartPads III is acceptable for its indicated use and the meets predefined criteria.
# 12. Conclusion
The proposed modifications do not present new issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of the HHS symbol, which is a stylized human figure embracing the world, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion around the symbol. The symbol is black, and the text is also black.
FEB 2 6 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Philips Medical Systems c/o Mr. Larry Walker Regulatory Affairs Specialist 2301 Fifth Avenue, Suite 200 Seattle, WA 98121
Re: K072812
Trade/Device Name: Philips Heartstart Smartpads III Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillators Regulatory Class: III (three) Product Code: MKJ, MLN Dated: January 18, 2008 Received: January 22, 2008
Dear Mr. Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Larry Walker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bfimmimor for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 11. Indications for Use
510(k) Number (if known): K072812
Device Name: Philips HeartStart SmartPads III Multifunction Defibrillator Pads
Indications For Use:
The Philips HeartStart SmartPads III is indicated for external defibrillation, pacing, monitoring and cardioversion.
The SmartPads III is intended for adults and children over 8 years of age or greater than 55 pounds (25kg) with all compatible defibrillators. The SmartPads III is also intended for children of all ages when used with a Philips Medical System model FRx equipped with an Infant/Child Key or with other compatible defibrillators which incorporate adjustable defibrillation energy functionality
The device is non-sterile and single-use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
t D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Or
Concurrence of CDRH, Office of Device Evaluation (ODE)
B. Zimmerman
ovascur Devices
PREMARKET NOTIFICATION PHILIPS MEDICAL SYSTEMS, HEARTSTREAM SMARTPAD III 510(K) PAGE 22 OF 27 CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.