← Product Code [PHX](/productcode/PHX) · K072804

# COMPREHENSIVE RS SHOULDER SYSTEM (K072804)

_Biomet Manufacturing Corp · PHX · Dec 12, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072804

## Device Facts

- **Applicant:** Biomet Manufacturing Corp
- **Product Code:** [PHX](/productcode/PHX.md)
- **Decision Date:** Dec 12, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3660
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Comprehensive® RS Shoulder System is indicated for use in patients whose shoulder ioint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. A cemented humeral stem must be used. The MacroBond®/HA RS Cleats are indicated only for uncemented biological fixation applications. The GT Baseplate components are intended for cementless application with the addition of screw fixation.

## Device Story

Comprehensive® RS Shoulder System is a reverse shoulder prosthesis for total shoulder replacement. Device replaces shoulder joint in patients with severe arthropathy and deficient rotator cuff. System components include humeral stem (cemented), MacroBond®/HA RS Cleats (uncemented), and GT Baseplate (cementless with screw fixation). Implanted by orthopedic surgeons in clinical/surgical settings. Device restores joint function by utilizing deltoid muscle to compensate for deficient rotator cuff. Benefits include pain relief and improved shoulder mobility for patients with otherwise limited surgical options.

## Clinical Evidence

No clinical data provided as a basis for substantial equivalence. Bench testing only.

## Technological Characteristics

Reverse shoulder prosthesis consisting of metal/polymer components. Features include cemented humeral stem, uncemented MacroBond®/HA RS Cleats, and cementless GT Baseplate with screw fixation. Semi-constrained design.

## Regulatory Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Predicate Devices

- Delta Shoulder and CTA™ Humeral Cups ([K021478](/device/K021478.md), [K050315](/device/K050315.md))
- Aequalis® Reversed Shoulder Prosthesis ([K030941](/device/K030941.md), [K041873](/device/K041873.md))
- Encore reverse Shoulder Prosthesis ([K041066](/device/K041066.md), [K051075](/device/K051075.md))
- Zimmer Trabecular Metal™ Reverse Shoulder System ([K052906](/device/K052906.md))
- Bio-Modular® Shoulder System ([K992119](/device/K992119.md), [K030710](/device/K030710.md), [K043100](/device/K043100.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized design of three human profiles facing to the right, with flowing lines suggesting movement or connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the design, emphasizing the organization's name and national affiliation.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 2, 2016

Biomet Manufacturing Corporation Ms. Patricia S. Beres Senior Regulatory Specialist 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581

Re: K072804

Trade/Device Name: Comprehensive® RS Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS Dated: September 28, 2007 Received: October 1, 2007

Dear Ms. Beres:

This letter corrects our substantially equivalent letter of December 12, 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

Page 2 - Ms. Patricia S. Beres

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.goy/Medica]Devices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Lori A. Wiggins -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510{k} Number (if known):___|イク72804

Device Name: Comprehensive® RS Shoulder System

Indications For Use:

The Comprehensive® RS Shoulder System is indicated for use in patients whose shoulder ioint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.

A cemented humeral stem must be used.

The MacroBond®/HA RS Cleats are indicated only for uncemented biological fixation applications. The GT Baseplate components are intended for cementless application with the addition of screw fixation.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Jubara Mreems
(Division Sign. Off)

Division of General, Restorative, and Neurological Devices

510(k) Number K022804 Page 1 of 1

4 - 1

{3}------------------------------------------------

K072804 (pg 1 of 2)

DEC 1 2 2007

Image /page/3/Picture/1 description: The image shows the logo for Biomet Orthopedics, Inc. The logo features the word "BIOMET" in a stylized, blocky font, with the letters connected to each other. Below the word "BIOMET" is the phrase "ORTHOPEDICS, INC." in a smaller, sans-serif font. The logo is black and white.

## 510(k) Summary

| Preparation Date:  | December 3, 2007                                        |
|--------------------|---------------------------------------------------------|
| Applicant/Sponsor: | Biomet Manufacturing Corp.                              |
| Contact Person:    | Patricia Sandborn Beres<br>Senior Regulatory Specialist |
| Proprietary Name:  | Comprehensive® RS Shoulder System                       |
| Common Name:       | Shoulder replacement components                         |

### Classification Name:

- Shoulder joint, metal/polymer, semi-constrained, cemented prosthesis ● (21 C.F.R. 888.3660) KWS
### Legally Marketed Devices To Which Substantial Equivalence Is Claimed:

- Delta Shoulder and CTA™ Humeral Cups (DePuy) K021478, K050315 ●
- K030941, K041873 Aequalis® Reversed Shoulder Prosthesis (Tomier) ●
- Encore reverse Shoulder Prosthesis (Encore) K041066. K051075
- Zimmer Trabecular Metal™ Reverse Shoulder System K052906 .
- Bio-Modular® Shoulder System (Biomet) K992119, K030710, ● K043100

Device Description: The Comprehensive® RS Shoulder System is intended for total shoulder replacement in a reverse shoulder.

#### Intended Use:

The Comprehensive® RS Shoulder System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.

A cemented humeral stem must be used.

Sailing Address: 658-18597 M 46581-0587

hipping Address:
5 East Bell Drive
arsaw, IN 46582

5 - 1

{4}------------------------------------------------

K072804 (pg 2 of 2)

The MacroBond®/HA RS Cleats are indicated only for uncemented blological fixation applications. The GT Baseplate components are intended for cementless application with the addition of screw fixation.

Summary of Technologies: The Comprehensive® RS Shoulder System have similar technologies as the predicate devices.

Non-Clinical Testing: Non-clinical laboratory testing was performed to determine substantial equivalence. The results indicated that the device was functional within its intended use.

Clinical Testing: None provided as a basis for substantial equivalence.

All trademarks of property of Blomet, Inc. except for the following: Delta CTA Is a trademark of Depuy Aequalis Is a trademark of Tomler S.A. Corporation Trabecular Metal are trademarks of Zimmer Trabecular Metal Technology, Inc

---

**Source:** [https://fda-staging.innolitics.com/device/K072804](https://fda-staging.innolitics.com/device/K072804)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
