VECTOR VISION HIP

K072716 · Brainlab AG · HAW · Oct 12, 2007 · Neurology

Device Facts

Record IDK072716
Device NameVECTOR VISION HIP
ApplicantBrainlab AG
Product CodeHAW · Neurology
Decision DateOct 12, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

BrainLAB VectorVision hip is intended to be an intraoperative image guided localization system. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space either on a patient's preoperative image data being processed by a VectorVision or a Ci™ workstation or on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, X-ray, MR based model of the anatomy. The system aids the surgeon to accurately navigate a hip endoprosthesis to the preoperatively or intraoperatively planned position. Example orthopedic surgical procedures include but are not limited to: Total Joint Replacement (TJR) Revision surgery of TJR Tumor resection and bone/joint reconstruction

Device Story

Intraoperative image-guided localization system for orthopedic surgery. Inputs: preoperative CT/X-ray/MR image data or intraoperatively acquired bone surface landmarks via tracked freehand probe. Operation: system registers patient anatomy to virtual 3D models; tracks surgical instruments (reamers, inserters, rasps, saws) using passive marker sensor system; reconstructs 3D mechanical axes/planes of femur and pelvis. Output: real-time navigation visualization on workstation display. Used in OR by surgeons to guide hip endoprosthesis placement. Benefits: improved accuracy in implant alignment relative to planned position.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Stereotaxic instrument; passive marker tracking system; workstation-based software; 3D anatomical modeling; CT/X-ray/MR data integration; intraoperative landmark acquisition; supports integration of implant/tool data.

Indications for Use

Indicated for patients undergoing stereotactic orthopedic surgery, including Total Joint Replacement (TJR), revision TJR, tumor resection, and bone/joint reconstruction, where rigid anatomical structures (skull, long bone, vertebra) can be referenced to CT, X-ray, or MR-based models.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## K072716 510 (k) Summary of Safety and Effectiveness for VectorVision® hip #### Manufacturer: | Manufacturer: | BrainLAB AG | |-----------------------------|--------------------------------------------------------------------------------------------------------| | Address: | Kapellenstr. 12<br>85622 Feldkirchen<br>Germany<br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 33 | | Contact Person: | Mr. Per Persson | | Summary Date: | August 30, 2007 | | Device Name: | | | Trade name: | VectorVision hip | | Common/Classification Name: | VectorVision, BrainLAB Image Guided Surgery System / Instrument<br>Stereotactic | Predicate Devices: Vector Vision® Hip 3.0 (K 040368) Navigation SW Hip 3.1 on CI (K-052213) Device Classification Name: Instrument, Stereotaxic Regulatory Class: Class II #### Intended Use: BrainLAB VectorVision hip is intended to be an intraoperative image guided localization system. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space either on a patient's preoperative image data being processed by a VectorVision or a Ci™ workstation or on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, X-ray, MR based model of the anatomy. The system aids the surgeon to accurately navigate a hip endoprosthesis to the preoperatively planned position. Example orthopedic surgical procedures include but are not limited to: Total Joint Replacement (TJR) Revision surgery of TJR Tumor resection and bone/joint reconstruction {1}------------------------------------------------ #### Device Description: BrainLAB VectorVision® Hip is intended to enable operational planning and navigation in orthopedic surgery. It links a surgical instrument, tracked by flexible passive markers to virtual computer image space on an individual 3D-model of the patient's bone, which is either based on a patients preoperative image data or generated through acquiring multiple landmarks on the bone surface. VectorVision® Hip uses the registered landmarks to navigate the needed surgical tools like cup reamer, cup inserter, stem rasp, bone saw and the implant to the planned position. If no implant data is available it is possible to provide information in order to achieve a generally targeted alignment relative to the bone orientation as defined by the operating surgeon. VectorVision® Hip allows 3-dimensional reconstruction of the relevant mechanical axes and planes of femur and pelvis and alignment of the implants. The VectorVision® Hip software has been designed to read in data of implants and tools if provided by the implant manufacturer and offers to individually choose the prosthesis during each surgery. The VectorVision® Hip software registers the patient data needed for planning and navigating the surgery intra-operatively within the CT free module. The System can be used to generally alian tool orientations according to the anatomy described and defined by the landmarks acquired by the surgeon. Using the CT based module the patient data can be used additionally for surgery, the patient data is then is provided by the CT data stored on a standard data storage media. #### Substantial equivalence: VectorVision® hip has been verified and validated according to the BrainLAB procedures for product design and development. The validation proves the safety and effectiveness of the information provided by BrainLAB in this 510 (k) application was found to be substantially equivalent with the predicate devices Vector Vision® Hip 3.0 (K 040368) and Navigation SW Hip 3.1 on CI (K-052213). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three wavy lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 BrainLAB AG % Mr. Benjamin Wagner Mechanical Design Engineer Kapellenstraße 12 85622 Feldkirchen Germany OCT 1 2 2007 Re: K072716 Trade/Device Name: VectorVision hip Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: September 10, 2007 Received: September 12, 2007 Dear Mr. Wagner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Benjamin Wagner This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Mark N. Melkerso Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use # 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name: VectorVision hip #### Indications For Use: BrainLAB VectorVision hip is intended to be an intraoperative image quided localization system. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space either on a patient's preoperative image data being processed by a VectorVision or a Ci™ workstation or on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, X-ray, MR based model of the anatomy. The system aids the surgeon to accurately navigate a hip endoprosthesis to the preoperatively or intraoperatively planned position. Example orthopedic surgical procedures include but are not limited to: - . Total Joint Replacement (TJR) - . Revision surgery of TJR - . Tumor resection and bone/joint reconstruction | Prescription Use<br>(Per 21 CFR 801 Subpart D) | X | |------------------------------------------------|---| |------------------------------------------------|---| AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) | | |------------------------------------------------|--| |------------------------------------------------|--| (Division Sign-Off) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Division of General, Restorative, Concurrence of CDRH, Office and Neurological Devices STORY Num
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...