K072698 · Confidant, Inc. · DRG · Nov 16, 2007 · Cardiovascular
Device Facts
Record ID
K072698
Device Name
CONFIDANT 2.5
Applicant
Confidant, Inc.
Product Code
DRG · Cardiovascular
Decision Date
Nov 16, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Indications for Use
Confidant 2.5 is intended for personal use by out-of-hospital patients as a means to retrospectively collect and record physiologic measurements from home monitoring devices (including blood glucose meters, blood pressure cuffs and weight scales). The data is transmitted to a database server where customized messages are generated by the system and returned to the patient. The returned messages contain objective observations and motivational information intended to help the patient better understand and manage their health. Confidant 2.5 is an accessory device that collects data from a range of supported home-monitoring devices. The data is collected from the supported devices and sent to a central database server, using standard wireless technologies. Upon receipt of newly submitted patient data, the Confidant Server software will generate and send one or more feedback messages directly to the patient's cell-phone. The feedback messages are selected by the system based on the patient's currently submitted and recent historic data. Confidant 2.5 does not provide diagnosis of any disease or medical condition. Confidant 2.5 is not intended to provide automated treatment decisions, or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional. Confidant 2.5 is not intended for emergency calls or for transmission or indication of any real-time alarms or time-critical data. This device is not intended as a substitute for direct medical supervision or emergency intervention. Confidant 2.5 is intended for over-the-counter use.
Device Story
Accessory system collecting physiologic data (glucose, blood pressure, weight) from home-monitoring devices; transmits data via wireless technology to central database server. Server software processes current and historic patient data to select and send automated feedback messages to patient's cell phone. Used by patients in home setting; OTC. Provides objective observations and motivational information to assist patient health management. Does not provide diagnosis, real-time alarms, or treatment decisions; requires professional healthcare oversight.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including electrical safety, EMC testing of the Confidant Connector, and software testing of the Confidant Collector and Server components, alongside device compatibility testing with supported home-monitoring hardware.
Technological Characteristics
System comprises Confidant Connector (hardware), Confidant Collector, and Confidant Server software. Utilizes standard wireless technologies for data transmission. Operates as an accessory to home-monitoring devices (glucose meters, BP cuffs, scales).
Indications for Use
Indicated for out-of-hospital patients for retrospective collection and recording of physiologic measurements (blood glucose, blood pressure, weight) from home monitoring devices to receive objective observations and motivational feedback via cell phone. Not for emergency use, real-time alarms, diagnosis, or automated treatment decisions.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
Honeywell HomMed Genesis OTC Monitor System (K061087)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for DC Confidant. The logo consists of two curved shapes resembling the letters 'D' and 'C'. To the right of the curved shapes, the word "CONFIDANT" is written in capital letters. The font appears to be a serif font.
# 510(K) SUMMARY
K072698
In accordance with 21 CFR 807.92, the following information constitutes Confidant's summary for the Confidant 2.5 System.
SUBMITTER'S NAME: ADDRESS: CONTACT PERSON: CONTACT PERSON TITLE: TELEPHONE NUMBER: FAX NUMBER: DATE OF SUBMISSION:
Confidant International, LLC 2530 Meridian Parkway, Suite 300 Daniel R. Plonski Director of Product Management (919) 806-4323 (919) 806-4802 September 20, 2007
## Identification of device 1
Proprietary Name: Confidant 2.5 Common Name: Physiological Transmitter and Receiver Classification Status: Class II per regulations 870.2910 Product Codes: DRG
## 2 Equivalent devices
Confidant Inc. believes that Confidant 2.5 is substantially equivalent to the following legally marketed devices:
Confidant 2.0 Honeywell HomMed Genesis K062215 OTC Monitor System K061087 Confidant Inc. Honeywell HomMed, LLC
## Description of the device 3
Confidant 2.5 is an accessory device that collects data from a range of supported home-monitoring devices. The data is collected from the supported devices and sent to a central database server, using standard wireless technologies. Upon receipt of newly submitted patient data, the Confidant Server software will generate and send one or more feedback messages directly to the patient's cell-phone. The feedback messages are selected by the system based on the patient's currently submitted and recent historic data.
Confidant 2.5 currently supports several models of glucose meters, noninvasive blood pressure cuffs and weight scales.
Page 1 OF 3
{1}------------------------------------------------
# 4 Intended use
Confidant 2.5 is intended for personal use by out-of-hospital patients as a means to retrospectively collect and record physiologic measurements from home monitoring devices (including blood glucose meters, blood pressure cuffs and weight scales). The data is transmitted to a database server where customized messages are generated by the system and returned to the patient. The returned messages contain objective observations and motivational information intended to help the patient better understand and manage their health.
Confidant 2.5 is an accessory device that collects data from a range of supported home-monitoring devices. The data is collected from the supported devices and sent to a central database server, using standard wireless technologies. Upon receipt of newly submitted patient data, the Confidant Server software will generate and send one or more feedback messages directly to the patient's cell-phone. The feedback messages are selected by the system based on the patient's currently submitted and recent historic data.
Confidant 2.5 does not provide diagnosis of any disease or medical condition.
Confidant 2.5 is not intended to provide automated treatment decisions, or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional.
Confidant 2.5 is not intended for emergency calls or for transmission or indication of any real-time alarms or time-critical data. This device is not intended as a substitute for direct medical supervision or emergency intervention.
Confidant 2.5 is intended for over-the-counter use.
page 2 of 3
2530 Meridian Parkway, Ste 300 Durham, NC 27713 (W) 919.806.4620 (F) 919.806.4802
{2}------------------------------------------------
## Technological characteristics, comparison to predicate 5 device.
Confidant 2.5 utilizes the same technology as one or both of the predicate devices (Confidant 2.0, K062215 and/or Honeywell HomMed Genesis OTC Monitor System, K061087) including:
- The same supported monitoring devices �
- The same operating features .
- The same fundamental technology .
### Discussion of functional and safety testing. 6
Testing of Confidant 2.5 included electrical safety and EMC testing of the Confidant Connector component; software testing of the Confidant Collector and Confidant Server components; and low-level, device compatibility testing with each of the supported monitoring devices. The test results demonstrate that Confidant 2.5 is in compliance with the applied standards and that it performed within its specifications and functional requirements.
#### Conclusion 7
Based on the comparison of intended use, supported monitoring devices, operational features and technology and the results of electrical safety. EMC, device compatibility and performance testing, it is our conclusion that Confidant 2.5 is as safe, as effective and performs as well as the legally marketed predicate devices.
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of a bird-like figure, which is the symbol of the HHS.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 6 2007
Confidant International, LLC c/o Mr. Daniel R. Plonski Director of Product Management 2530 Meridian Parkway, Suite 300 Durham, NC 27713
Re: K072698
Device Name: Confidant 2.5 Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II (two) Product Code: DRG Dated: September 20, 2007 Received: September 24, 2007
Dear Mr. Plonski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Daniel R. Plonski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
B.Jemimson for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K072698
Device Name: Confidant 2.5
Indications For Use: Confidant 2.5 is intended for personal use by out-of-hospital patients as a means to retrospectively collect and record physiologic measurements from home monitoring devices (including blood glucose meters, blood pressure cuffs and weight scales). The data is transmitted to a database server where customized messages are generated by the system and returned to the patient. The returned messages contain objective observations and motivational information intended to help the patient better understand and manage their health.
Confidant 2.5 is an accessory device that collects data from a range of supported home-monitoring devices. The data is collected from the supported devices and sent to a central database server, using standard wireless technologies. Upon receipt of newly submitted patient data, the Confidant Server software will generate and send one or more feedback messages directly to the patient's cell-phone. The feedback messages are selected by the system based on the patient's currently submitted and recent historic data.
Confidant 2.5 does not provide diagnosis of any disease or medical condition.
Confidant 2.5 is not intended to provide automated treatment decisions, or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional.
Confidant 2.5 is not intended for emergency calls or for transmission of any real-time alarms or timecritical data. This device is not intended as a substitute for direct medical supervision or emergency intervention.
Confidant 2.5 is intended for over-the-counter use.
Prescription Use AND/OR Over-The-Counter Use X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhumima
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.