K072693 · Toshiba America Medical Systems, In.C · JAK · Nov 1, 2007 · Radiology
Device Facts
Record ID
K072693
Device Name
CSNP-001A NEURO PACKAGE
Applicant
Toshiba America Medical Systems, In.C
Product Code
JAK · Radiology
Decision Date
Nov 1, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
This device is application software that permits cerebral perfusion imaging based upon dynamic volume sets. The software allows for visualization of apparent blood flow in the brain tissue and pictorially illustrates perfusion parameters (CBP - cerebral blood perfusion, CBV - cerebral blood volume, MTT - mean time to transit, TTP - time to peak of the time density curve, and DLY - delay of each voxel curve from the arterial curve). This device, when used by a trained physician, will assist in the assessment of the type and extent of cerebral perfusion disturbances. Additionally, the software allows for measurement of various values such as mean value, standard deviation, area and distance.
Device Story
Post-processing software package; processes dynamic volume sets acquired via CT; utilizes Time Density Curve (TDC) analysis; generates image maps and numeric data; visualizes cerebral blood perfusion (CBP), cerebral blood volume (CBV), mean transit time (MTT), time to peak (TTP), and delay (DLY); used by trained physicians in clinical settings; assists in assessing brain trauma or stroke; provides measurements including mean value, standard deviation, area, and distance; output aids clinical decision-making regarding perfusion disturbances.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Post-processing software package; operates on dynamic volume sets; provides visual and numeric perfusion parameters; conforms to IEC 60601-1 safety standards; developed under 21 CFR 820 quality system regulations.
Indications for Use
Indicated for use by trained physicians to perform brain perfusion studies, including whole brain studies, using time-sequential data to assess brain trauma or stroke.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
KC72693
Page 1 of 2
## Toshiba America Medical Systems, Inc. 510(k) Pre-market Notification; CSNP-001A Neuro Package
# 510(k) Summary
| Date: | September 20, 2007 | NOV - 1 2007 |
|---------------------------------------|-----------------------------------------------------------------------------------|--------------|
| Submitter's Name: | Toshiba America Medical Systems, Inc. | |
| Submitter's Address: | P.O. Box 2068, 2441 Michelle Drive,<br>Tustin, CA 92781-2068 | |
| Submitter's Contact: | Paul Biggins, Director Regulatory Affairs<br>(714)730-5000 | |
| Establishment Registration<br>Number: | 2020563 | |
| Device Proprietary Name: | CSNP-001A Neuro Package | |
| Common Name: | Scanner, Computed Tomography, X-Ray<br>[Fed. Reg. No. 892.1750, Pro. Code: 90JAK] | |
| Regulatory Class: | II (per 21 CFR 892.1750) | |
| Performance Standard: | 21 CFR Subchapter J,<br>Federal Diagnostic X-ray Equipment Standard | |
| Predicate Device(s): | CSCP-001A Cerebral Bloop Perfusion Analysis Package | |
| Reason For Submission | New device | |
### Description of this Device:
This device is a post processing package that allows the user to process dynamic volume sets using a Time Density Curve. The package allows visualization of the data in image map format. Additionally, the software provides numeric data via image analysis software.
### Summary of Intended Uses:
This device is application software that permits cerebral perfusion imaging based upon dynamic volume sets. The software allows for visualization of apparent blood flow in the brain tissue and pictorially illustrates perfusion parameters (CBP - cerebral blood perfusion, CBV - cerebral blood volume, MTT - mean time to transit, TTP - time to peak of the time density curve, and DLY - delay of each voxel curve from the arterial curve). This device, when used by a trained physician, will assist in the assessment of the type and extent of cerebral perfusion disturbances. Additionally, the software allows for measurement of various values such as mean value, standard deviation, area and distance.
{1}------------------------------------------------
## Toshiba America Medical Systems, Inc. 510(k) Pre-market Notification; CSNP-001A Neuro Package
### Technological Characteristics:
This device is similar to the predicate device in uses and applications. The primary difference is in the method used to obtain final results. Both devices are post processing software packages.
#### Safety and Effectiveness Concerns:
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820. Additionally this device is in conformance with the applicable parts of the IEC 60601-1 international safety standard.
#### Substantial Equivalence:
Based upon the above considerations TAMS believes that this device, CSNP-001A Neuro Package, is substantially equivalent to the predicate device and other devices already marketed in the US. This device and other similar marketed devices are post processing software that provide visual and numeric data. Additionally, this device has the same indications of previously marketed devices.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling a human figure with outstretched arms, accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The symbol is composed of three curved lines that converge to form the shape of a person.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Toshiba America Medical Systems. Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 24th Street NW BUFFALO MN 55313
NOV - 1 2007
Re: K072693
Trade/Device Name: CSNP-001A, NeuroPackage Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK and LLZ Dated: October 22, 2007 Received: October 23, 2007
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) for CSNP-001A Neuro Package
## Indications for Use
510(k) Number (if known):
072693
Device Name: CSNP-001A, NeuroPackage
Indications for Use:
The CSNP-001A is a post processing package that has the capability to perform brain perfusion studies to include whole brain studies, when used by a trained physician, by using data that is acquired on a time sequential basis. This package provides the physician with the following information that can then be measured and analyzed in the assessment of brain trauma or stroke:
TDC (time density curve)
CBF: Blood flow in the capillary vessels of the cerebral tissues.
CBV: Blood volume in the capillary vessels of the cerebral tissues.
MTT: Mean transit time of blood in the capillary vessels f the cerebral vessels.
TTP: Time to peak of the TDCs.
DLY: Delay of each voxel curve from the arterial curves.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of *_*_
ductive, Abdominal and Radiologic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.