K072629 · Integrated Surgical Systems, Inc. · OJP · Aug 6, 2008 · Neurology
Device Facts
Record ID
K072629
Device Name
DIGIMATCH ROBODOC SURGICAL SYSTEM
Applicant
Integrated Surgical Systems, Inc.
Product Code
OJP · Neurology
Decision Date
Aug 6, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DigiMatch™ ROBODOC® Surgical System is intended for use as a device, which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intraoperative procedures. The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan. The preoperative planning software and robotic surgical tool is used as an alternative to manual planning and broaching/reaming techniques for femoral canal preparation in primary total hip arthroplasty (THA). The DigiMatch ROBODOC Surgical System is indicated for Orthopedic procedures in which the broaching/reaming in primary total hip arthroplasty (THA) may be considered to be safe and effective and where references to rigid anatomical structures may be made.
Device Story
System comprises ORTHODOC preoperative planning workstation and ROBODOC robotic surgical tool. Inputs: patient CT scan data. ORTHODOC software allows surgeon to assess femoral anatomy, select implant size, and determine positioning. ROBODOC, an electromechanical arm, implements the plan under direct surgeon control. Used in clinical settings for primary total hip arthroplasty (THA) as an alternative to manual broaching/reaming. Surgeon monitors robotic implementation to ensure precise femoral canal preparation. Benefits include improved implant fit, alignment, and reduced gaps between bone and implant compared to manual techniques.
Clinical Evidence
Evidence includes bench, animal, cadaver, and clinical studies. Cadaver studies showed reduced gaps between implant and bone vs. manual. Animal study (20 greyhounds) showed better cortical bone continuity and implant apposition. Two U.S. clinical trials (G920035, G000071) demonstrated safety and efficacy; primary endpoint of Harris Hip Scores showed equivalence between ROBODOC and manual control groups.
Technological Characteristics
System includes ORTHODOC workstation and ROBODOC electromechanical arm with end effector. Uses CT scan data for preoperative planning. Registration via point-to-surface method. Software-driven control of robotic arm for precise implementation of surgical plan. Class II device.
Indications for Use
Indicated for patients requiring primary total hip arthroplasty (THA) where femoral canal preparation via broaching/reaming is appropriate and rigid anatomical structures can be referenced.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
### Summary of Safety and Effectiveness K072629
AUG - 6 2008
| Submitter: | Integrated Surgical Systems, Inc. |
|------------------------|----------------------------------------------------|
| Address: | 1433 N. Market Blvd., #1, Sacramento, CA 95834 USA |
| Phone Number: | 916-285-9943 |
| Fax Number: | 916-575-7918 |
| Contact Person: | Russ Hibbert |
| Date Submitted: | September 7, 2007 |
| Trade Name: | DigiMatch™ ROBODOC® Surgical System |
| Common Name: | Stereotaxic Instrument |
| Device Classification; | Class II |
| Regulation Number: | 21 CFR 882.4560 |
#### Substantial Equivalence Claim
The DigiMatch™ ROBODOC® Surgical System is substantially equivalent to the Voyager/Tactile Guidance System -- CT (K052851), the Frameless Neuromate (K991081) and the da Vinci Surgical System (K043153).
#### Device Description
The DigiMatch™ ROBODOC® Surgical System consists of the 510(k)-cleared (K960685) ORTHODOC® Preoperative Planning Workstation (ORTHODOC), and a Robot composed of an electromechanical arm, electronics control cabinet, display monitor and miscellaneous accessories (ROBODOC). The System (ORTHODOC and ROBODOC) is a threedimensional graphical preoperative planner and implementation tool and is indicated as an alternative to template planning and manual broaching for treatment of patients who require primary total hip arthroplasty (THA). The ORTHODOC component of the System allows a surgeon to preoperatively assess a patient's femoral anatomy and state of hip disease, select an optimally sized femoral stem implant from a library of prostheses, and determine where the implant should be positioned within the femur. Then ROBODOC component of the System, under direct control by the surgeon, will precisely implement the preoperative plan.
#### Indications for Use / Intended Use Statement
The DigiMatch™ ROBODOC® Surgical System is intended for use as a device, which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures, The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan.
The preoperative planning software and robotic surgical tool is used as an alternative to manual planning and broaching/reaming techniques for femoral canal preparation in primary total hip arthroplasty (THA).
The DigiMatch ROBODOC Surgical System is indicated for Orthopedic procedures in which the broaching/reaming in primary total hip arthroplasty (THA) may be considered to be safe and effective and where references to rigid anatomical structures may be made.
{1}------------------------------------------------
## Summary of Safety and Effectiveness K072629
# Comparison of Technological Characteristics and Principles of Operation
Table 1 provides a comparison of technological characteristics and principles of operation between the DigiMatch™ ROBODOC® Surgical System and its predicate devices, the Voyager/Tactile Guidance System - CT, the Frameless Neuromate and the da Vinci Surgical System.
#### Comparison of Technological Characteristics and Principles of Operation Table 1
| | Patient | Pre- | Surgical Plan | Machine | Patient/Robot | Robot |
|---------------------------------------------------------|--------------------|---------------------|-------------------------------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| | Image | Surgical | Data | Instructions | Registration | Electromechanic |
| | Data | Plan | | | Requirement | al Arm(s) |
| DigiMatch™<br>ROBODOC®<br>Surgical System | Yes<br>CT Scan | Yes<br>pre-surgery | Yes<br>high level<br>operative plan | Yes<br>robotic arms<br>driven by<br>validated control<br>software and<br>hardware | Yes<br>point to surface<br>registration | Yes<br>Robot with single<br>electromechanical<br>arm and end<br>effector implement<br>control file<br>instructions |
| Voyager<br>Tactile Guidance<br>System - CT<br>(K052851) | Yes<br>CT Scan | Yes<br>pre-surgery | Yes<br>high level<br>operative plan | Yes<br>robotic arms<br>driven by<br>validated control<br>software and<br>hardware | Yes<br>surface to<br>surface<br>registration | Yes<br>Robot with single<br>electromechanical<br>arm and end<br>effector implement<br>control file<br>instructions |
| Frameless<br>Neuromate<br>(K991081) | Yes<br>CT Scan | Yes<br>pre-surgery | Yes<br>high level<br>operative plan | Yes<br>robotic arms<br>driven by<br>validated control<br>software and<br>hardware | Yes<br>fiducial marker<br>registration | Yes<br>Robot with single<br>electromechanical<br>arm and end<br>effector implement<br>control file<br>instructions |
| da Vinci®<br>Surgical System<br>(K043153) | No<br>visual image | No<br>intra-surgery | No<br>control file<br>developed in<br>real time during<br>surgery | Yes<br>robotic arms<br>driven by<br>validated control<br>software and<br>hardware | Yes<br>visual<br>registration | Yes<br>Robot with<br>multiple<br>electromechanical<br>arms and end<br>effector implement<br>control file<br>instructions |
{2}------------------------------------------------
### Summary of Safety and Effectiveness K072629
#### Summary of Performance Testing
ISS has completed bench, animal, cadaver and clinical testing that assess the safety and efficacy of the ROBODOC System.
#### Nonclinical Studies
Nonclinical cadaver and animal studies provide an accurate means of determining cavity accuracy and placement. The value of such studies is that they allow a measurement of the gap between the bone and implant. They also permit histological examinations of the cavity/implant interface, which has value in determining whether the implant apposition within the cavity is sufficiently narrow.
#### Cadaver Study
The cadaver study compared the accuracy of cutting the femur with the ROBODOC versus manual broaching. The results showed that the ROBODOC was more accurate than manual broaching and led to significantly reduced gaps between the implant and bone.
#### Animal Study
The animal study, involving 20 male greyhounds, also supports the conclusion that the ROBODOC is more accurate than manual broaching in cutting the femur. Histological examination revealed many instances of fracture or osteotomy in the control group, compared to none in the ROBODOC group. In addition, histological examination revealed that the ROBODOC group had more consistent cortical bone continuity, better apposition of implants with cortical and cancellous bone, and appeared to have less cancellous bone hypertrophy.
### Clinical Studies
The First U.S. Trial (G920035) was conducted using the "pin-based" technology, which required three locator (fiduciary) pins to register the robotic device to the patient. The primary endpoints of efficacy that demonstrated precise and reliable implantation of the preoperative plan were fit and alignment of the femoral implant. Primary clinical efficacy was demonstrated showing equivalence in Harris Hip Scores between the ROBODOC group and the manual control group.
The Second U.S. Trial (IDE # G000071) was undertaken with the "pinless" DigiMatch™ ROBODOC® Surgical System. The DigiMatch™ technology modified the ROBODOC System so as not to require pre-operative placement of fiduciary pins. The data from this trial also confirmed the safety and efficacy of the ROBODOC System. As with the First IDE Trial, primary clinical efficacy was demonstrated showing equivalence in Harris Hip Scores between the ROBODOC group and the manual control group.
### Conclusion
The DigiMatch™ ROBODOC® Surgical System, which is substantially equivalent to the Voyager/Tactile Guidance System - CT, the Frameless Neuromate and the da Vinci Surgical System, has been shown to be safe and effective for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 6 2008
Intergrated Surgical Systems, Inc. % Mr. Russ Hibbert Director, Clinical & Regulatory Affairs 1433 N. Market Boulevard #1 Sacramento, California 95834
Re: K072629
Trade/Device Name: DigiMatch™ ROBODOC® Surgical System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: OJP, HAW Dated: March 31, 2008 Received: April 3, 2008
Dear Mr. Hibbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Russ Hibbert
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 10gally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarlet, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events. (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Mullison
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): Unknown at this time K072629
Device Name: DigiMatch™ ROBODOC® Surgical System
Indications for Use:
The DigiMatch™ ROBODOC® Surgical System is intended for use as a device, which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intraoperative procedures. The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan.
The preoperative planning software and robotic surgical tool is used as an alternative to manual planning and broaching/reaming techniques for femoral canal preparation in primary total hip arthroplasty (THA).
The DigiMatch ROBODOC Surgical System is indicated for Orthopedic procedures in which the broaching/reaming in primary total hip arthroplasty (THA) may be considered to be safe and effective and where references to rigid anatomical structures may be made.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Wilkerson
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 1 of I
510(k) Number K0 to
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.