← Product Code [DTR](/productcode/DTR) · K072583

# PLEGIOX CARDIOPLEGIA HEAT EXCHANGER, WITH SAFELINE COATING, MODEL(S) CHX30, BSQ-CHX 30 (K072583)

_Maquet Cardiopulmonary, AG · DTR · Apr 4, 2008 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072583

## Device Facts

- **Applicant:** Maquet Cardiopulmonary, AG
- **Product Code:** [DTR](/productcode/DTR.md)
- **Decision Date:** Apr 4, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4240
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Plegiox heat exchanger is used to set and maintain the temperature (for given flow rates and within the given temperature range) of blood cardioplegic and crystalloid cardioplegic solutions during extracorporeal circulation. The product is designed for single use only, for an application period of no longer than 6 hours.

## Device Story

Plegiox is a sterile, single-use cardioplegia heat exchanger with an integrated bubble trap; used during extracorporeal circulation to regulate temperature of blood or crystalloid cardioplegic solutions. Device may include Safeline coating to reduce surface tension on blood-contacting surfaces. Operated by perfusionists or clinical staff in surgical settings. Heat exchange occurs via fluid-to-fluid thermal transfer; bubble trap removes air from perfusion circuit. Output is temperature-controlled cardioplegic solution delivered to patient heart to induce/maintain cardioplegia. Benefits include precise thermal management and air removal during cardiopulmonary bypass.

## Clinical Evidence

Bench testing only. In-vitro testing performed to demonstrate safety and effectiveness compared to predicate devices.

## Technological Characteristics

Cardiopulmonary bypass heat exchanger; integrated bubble trap; optional Safeline coating for surface tension reduction. Single-use, sterile, pyrogen-free. Class II device (21 CFR 870.4240).

## Regulatory Identification

A cardiopulmonary bypass heat exchanger is a device, consisting of a heat exchange system used in extracorporeal circulation to warm or cool the blood or perfusion fluid flowing through the device.

## Predicate Devices

- CSC 14 Blood Cardioplegia System, Cobe Cardiovascular Inc. ([K012898](/device/K012898.md))
- Jostra RotaFlow Centrifugal Pump with Safeline Coating, Maquet Cardiopulmonary AG ([K061072](/device/K061072.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# MAQUET

# 510 (K) Summary [as required by 21 CFR 807.92(c) ]

Maquet Cardiopulmonary AG Submitter: Hechinger Strasse 38 72145 Hirrlingen APR - 4 2008 Germany Katrin Schwenkglenks Contact Person:

Phone: +49 7478 921-151 Fax: +49 7478 921-400 E-mail: katrin.schwenkglenks@maquet-cp.com

September 07, 2007

Plegiox Cardioplegia Heat Exchanger with and Device Trade Name: without Safeline Coating

Cardioplegia Heat Exchanger Common/Usual Name:

Classification Names:

Predicate Devices:

Date Prepared:

Cardiopulmonary Bypass Heat Exchanger (21 CFR 870.4240, product code: DTR)

CSC 14 Blood Cardioplegia System, Cobe Cardiovascular Inc. (K012898) and Jostra RotaFlow Centrifugal Pump with Safeline Coating (K061072), Maquet Cardiopulmonary AG.

030(v2)

### Device Description

The Plegiox is a product, which is delivered to the end user in a pyrogen-free and sterile way. The product is determined for single use only. Re-sterilization and re-use are forbidden. The Plegiox is a cardioplegia heat exchanger with integrated bubble trap.

### Statement of Indications for Use

The Plegiox heat exchanger is used to set and maintain the temperature (for given flow rates and within the given temperature range) of blood cardioplegic and crystalloid cardioplegic solutions during extracorporeal circulation.

The product is designed for single use only, for an application period of no longer than 6 hours.

## Statement of Technical Characteristics Comparison

Maquet Cardiopulmonary AG has compared indications for use, design, specifications, performance characteristics and safety of the Plegiox Heat Exchanger with and without Safeline Coating and of the predicate devices.

In-vitro testing on safety and effectiveness was executed to demonstrate that the Plegiox with and without Safeline Coating described in this submission is substantially equivalent to the named predicate devices.

{1}------------------------------------------------

Indications for use for the Plegiox Cardioplegia Heat Exchanger, Maquet Cardiopulmonary AG:

"The Plegiox heat exchanger is used to set and maintain the temperature (for given flow rates and within the given temperature range) of blood cardioplegic and crystalloid cardioplegic solutions during extracorporeal circulation.

The product is designed for single use only, for an application period of no longer than 6 hours.

CSC 14 Cardioplegia Heat Exchanger, Cobe Cardiovascular Inc.:

"CSC14 is recommended for use as a heating/cooling device and bubble trap for cardioplegia and clear fluid perfusion in extracorporeal circulation associated with cardiopulmonary bypass. It is suggested not to use CSC 14 for more than 6 hours.

CSC 14 can be used in combination with the devices listed in paragraph 1. Devices which can be used with CSC 14."

#### Safeline Coating:

The Safeline Coating applied to the Plegiox Cardioplegia Heat Exchanger has the same intended use as the predicate RotaFlow Centrifugal Pump with Safeline Coating from Maquet Cardiopulmonary. The indications for use of the Safeline Coating is "To reduce the surface tension on blood contact surfaces." for both devices.

#### Comparison of the performance

The substantial equivalence in performance of the Plegiox Cardioplegia Heat Exchanger with the predicate device has been shown by in-vitro testing. The Plegiox Cardioplegia Heat Exchanger has shown to be as effective as the predicate device.

Prepared by Maquet Cardiopulmonary AG, Hirrlingen, Germany

Confidential

040(v2)

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 4 2008

Maquet Cardiopulmonary AG c/o Ms. Katrin Schwenkglenks Regulatory Affairs Manager Hechinger Strasse 38 72145 Hirrlingen, Germany

Re: K072583

Plegiox Cardioplegia Heat exchanger with and without Safeline Regulation Number: 21 CFR 870.4240 Regulation Name: Cardiopulmonary bypass heat exchanger Regulatory Class: Class II (two) Product Code: DTR Dated: March 14, 2008 Received: March 19, 2008

Dear Ms. Schwenkglenks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Katrin Schwenkglenks

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrf/industry/support/index.html.

Sincerely yours,

Dmna R. bchmer

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): __ Ko72583 Device Name: Plegiox Cardiplegia Heat Exchanger with and without Safeline Coating

Indications for Use:

The Plegiox heat exchanger is used to set and maintain the temperature (for given flow rates and within the given temperature range) of blood cardioplegic and crystalloid cardioplegic solutions during extracorporeal circulation.

The product is designed for single use only, for an application period of no longer than 6 hours.

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Donna R. Seidel

ivision of Cardiovascular Devices

510(k) Number K072583

Page 1 of

029(v2

(Posted November 13, 2003)

---

**Source:** [https://fda-staging.innolitics.com/device/K072583](https://fda-staging.innolitics.com/device/K072583)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).