TRUEPIX LCD MEDICAL DISPLAY MONITOR, MODELS TX-5M, TX-3M, TX-2M
K072532 · Dooyang Systems, Inc. · LLZ · Nov 2, 2007 · Radiology
Device Facts
Record ID
K072532
Device Name
TRUEPIX LCD MEDICAL DISPLAY MONITOR, MODELS TX-5M, TX-3M, TX-2M
Applicant
Dooyang Systems, Inc.
Product Code
LLZ · Radiology
Decision Date
Nov 2, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
Monochrome LCD Monitors (TX-5M, TX-3M, TX-2M) intended to be used in various kinds of medical image application for which the device complies with the performance specified by the manufacturer of the system. Not for use in mammography.
Device Story
Truepix LCD Medical Display Monitors (TX-5M, TX-3M, TX-2M) are monochrome TFT LCD displays used to visualize medical images like X-rays and MRIs. Monitors connect to PCs via graphics cards (Real Vision & Matrox Series) and feature USB connectivity. Optional photo-sensor (GFU12-SEQ01) and software enable luminance calibration. Devices are deployed in portrait or landscape orientation. Healthcare providers use these monitors to view diagnostic images to support clinical decision-making. Safety certified to EN60601-1.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Monochrome TFT LCD panels; resolutions 1200x1600 to 2560x2048; brightness 800-1000 cd/m2; contrast 600:1-700:1; USB interface; universal power supply (AC 100-240V); EN60601-1 safety certified; optional photo-sensor for luminance calibration.
Indications for Use
Indicated for display of medical images (e.g., X-ray, MRI) in clinical settings. Not for use in mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Page 1 of 2
### Exhibit 2 NOV - 2 2007 510K Summary: K072532 DOOYANG SYSTEMS INC. SK Techno Bldg. 7th Floor 16-4, Sungsoo 1-Ga, 2-Dong, Sungdong-Ku, Seoul, 133-710 Korea TEL : 82-2-489-2612 FAX : 82-2-467-4915 Contact: Yun-Hwoi, Ku /Managing Director Date : October 30, 2007
### 1. Identification of Device :
Proprietary-Trade Name: "Truepix LCD Medical Display Monitor" (Models : TX-5M, TX-3M, TX-2M)
Classification Name: System, Image Processing, Radiological, Product Code: LLZ Common/Usual Name: Radiological Image Processing System, LCD Monitor
#### 2. Equivalent Legally Marketed Devices:
Lumimed MM20, MM30 and MM50 Monitors, K052120
#### 3. Indication for Use (Intended Use) :
Monochrome LCD Monitors (TX-5M, TX-3M, TX-2M) intended to be used in various kinds of medical image application for which the device complies with the performance specified by the manufacturer of the system. Not for use in mammography.
### 4. Description of the Device
Monochrome LCD Monitors(TX-5M, TX-3M, TX-2M) are used to display images such as Xray, MRI images. These models have resolutions of: 2560x2048, 2048x1536, 1600x1200. These models have USB functions and an optional photo sensor GFU12-SEQ01 made by SEQUEL IMAGING INC. These models are cortified to the EN60601-1 medical safety standard. The monitors use universal Power Supply compatible with AC100V~240V, 50/60Hz. The graphics card needed for the personal computer is the Real Vision & Matrox Scries. The monitors may be deployed in the portrait or the landscape position.
{1}------------------------------------------------
KC72532
Page 2 of 2
:
| Items | Lumimed MM20, MM30<br>and MM50 Monitors, | New | | |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|-----------------------------------------------|--------------------------------------------------|
| | | Truepix TX-2M | Truepix TX-3M | Truepix IX-5M |
| 510(k) Numbers | K052120 | New | | |
| Manufactures | Heeyoung Cl. | Dooynag Systems Inc. | | |
| Panel Size and Type | 21.3" /20.1" TFT<br>monochrome LCD display | 21.3" TFT<br>monochrome CD<br>display | 21.3" TFT<br>monochrome LCD<br>display | 20.1" IFT<br>monochrome LCD<br>display |
| Pixel Pitch | 0.270mm x 0.270mm<br>0.2115mm x 0.2115mm<br>0.156mm x 0.156mm | 0.270mm x 0.270mm | 0.2115mm x<br>0.2115mm | 0.156mm x 0.156mm |
| Available Cabinet<br>Colors | Black | Black | Black | Black |
| Native Resolutions | 1200x1600(portrait)<br>1600x1200(landscape)<br>1536x2048 (portrait)<br>2048x1536(landscape) or<br>2048x2560 (portrait)<br>2560x2048(landscape) | 1200x1600 (portrait)<br>1600x1200 (landscape) | 1536x2048 (portrait)<br>2048x1536 (landscape) | 2048x2560 (portrait)<br>2560x2048<br>(landscape) |
| Brightness | 1000cd/m2 900cd/m2<br>850cd/m2 | 1000cd/m2 | 800cd/m2 | 850cd/m2 |
| Contrast Ratio | 700:1 700:1 600:1 | 700:1 | 700:1 | 600:1 |
| Dot Clock | 125MHz 130MHz 162MHz | 129MHz | 130MHz | 166.52MHz |
| Network Interface | USB<br>(1 Up, 2 Downstream) | USB<br>(1 Up, 4 Downstream) | USB<br>(1 Up, 4 Downstream) | USB<br>(1 Up, 4 Downstream) |
| Active Display Size<br>(HxV) | 432(H) x 324(V)<br>432(H) x 324(V)<br>399(H) x 319(V) | 433.152x324.864mm | 433.152x324.864mm | 399.36x319.49mm |
| Dimension (WxHxD) | 472 x 495 x92<br>472 x 495 x92<br>438 x 458 x 98 | 505 w x 505 h x 81 d | 505 w x 505 h x 81 d | 445 w x 376 h x 90 d |
| Luminance Calibration | Software (Optional)<br>Photo-sensor(Optional) | Software (Optional)<br>Photo-sensor(Optional) | Software (Optional)<br>Photo-sensor(Optional) | Software (Optional)<br>Photo-sensor(Optional) |
| Power | AC 100~240Volts/60Hz<br>DC 12V,6,67A | AC 100~240Volts/60Hz<br>DC 12V,6,67A | AC 100~240Volts/60Hz<br>DC 12V,6,67A | AC<br>100~240Volts/60Hz<br>DC 12V,6,67A |
## 5. Safety and Effectiveness, comparison to predicate device
### 6. Testing Information and Conclusion
In all matcrial respects, the Truepix LCD Medical Display Monitors are substantially equivalent to the predicate device. Testing was performed according to internal company procedures and the monitors were safety certificd to International Standard.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES'
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 2 2007
Doovang Systems, Inc. % Mr. Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates PO Box 7007 DEERFIELD IL 60015
Re: K072532
Trade/Device Name: Truepix LCD Medical Display Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 4, 2007 Received: September 17, 2007
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Page 2
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): _ K072532___________________________________________________________________________________________________________________________________________
Device Name: "Truepix LCD Medical Display Monitor"
Indications For Use: Monochrome LCD Monitors (TX-5M, TX-3M, TX-2M) intended to be used in various kinds of medical image application for which the device complies with the performance specified by the manufacturer of the system. These monitors are not intended for use in mammography.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
tomu m why
(Division Sign-Off)
Page 1 of 1
(Division Sign-Οπ)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K072532
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.