← Product Code [IYE](/productcode/IYE) · K072454

# INTEGRA RADIONICS INTERFIX PATIENT ADAPTERS (K072454)

_Integra Radionics, Inc. · IYE · Oct 16, 2007 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072454

## Device Facts

- **Applicant:** Integra Radionics, Inc.
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** Oct 16, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

For cranial fixation during CT scanning and treatment with the TomoTherapy HiArt System.

## Device Story

Interfix Patient Adapters consist of two mechanical components: a CT adapter for scanner tabletops and a tabletop adapter for the TomoTherapy HiArt System. Device functions as a physical interface to secure Integra Radionics headrings (e.g., Gill-Thomas-Cosman, UCHR, HRA-IM, TLC Pediatric Frame) to imaging and treatment platforms. Used in clinical settings by healthcare providers to maintain consistent patient positioning between diagnostic CT scanning and radiation therapy. Adapters aid in laser alignment; however, prescan procedures remain required for final positioning verification. Benefits include improved patient stability and reproducibility of position during the radiation treatment planning and delivery workflow.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Mechanical cranial fixation adapters. Components: CT adapter and TomoTherapy HiArt tabletop adapter. Designed to interface with existing Integra Radionics headring systems (Gill-Thomas-Cosman, UCHR, HRA-IM, TLC Pediatric Frame). Non-powered, passive mechanical hardware.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- LCM-2 Linac Couch-Mount System ([K924632](/device/K924632.md))
- Knife Radiosurgery System ([K912630](/device/K912630.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Integra Radionics Interfix™ Patient Adapters 510(k) Summary

### Submitter's Name and Address:

Integra Radionics 22 Terry Avenue Burlington, MA 01803 781-565-1227 (Telephone) 781-238-0645 (Fax)

### Contact Person and Telephone Number:

Kevin J. O'Connell Regulatory Affairs Manager Integra Radionics, Inc. Tel.: (781) 565-1227

### Date Summary was Prepared: August 30, 2007.

#### Name of the Device:

| Trade Name:          | Integra Radionics Interfix™ Patient Adapters.                     |
|----------------------|-------------------------------------------------------------------|
| Common Name:         | Stereotactic Radiation Treatment Planning System and Accessories. |
| Classification Name: | Accelerator, Linear, Medical<br>21 CFR 892.5050, Product Code IYE |

## Classification Panel: Radiology

#### Substantial Equivalance:

The modified device is intended for cranial fixation during CT scanning and treatment with the Tomo Therapy HiArt system. Cranial fixation is within the indications for use of the predicate devices. The technological characteristics are similar to those found with the following predicate devices: LCM-2 Linac Couch-Mount System cleared via 510(k) K924632 on April 7, 1993. Knife Radiosurgery System cleared via 510(k) K912630 on November 4, 1991.

The Interfix Patient Adapters consists of two adapters, the InterFix™ CT Adapter which attaches to the CT scanner tabletop and the InterFix™ Tabletop Adapter for attachment to the TomoTherapy HiArt System tabletop. The adapters are designed to accept the Integra Radionics headrings, such as the Gill-Thomas-Cosman relocatable headring. Universal Compact Head Ring 

{1}------------------------------------------------

(UCHR), Intubation Headring Assembly (HRA-IM), and Tarbell-Loeffler Cosman (TLC) Pediatric Frame.

The purpose of the Interfix system is to hold a patient securely and in the same position during the diagnostic CT scan and TomoTherapy HiArt Treatment. The adapters can be used to aid the user in aligning the patient to the lasers; however, the prescan is still required to ensure that the patient is in the correct position.

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the address for the Food and Drug Administration. The address is 9200 Corporate Blvd, Rockville MD 20850. The text is black and the background is white.

# OCT 16 2007

Mr. Kevin J. O'Connell Manager Regulatory Affairs Integra Radionics. Inc. 22 Terry Ave. BURLINGTON MA 01803

Re: K072454

Trade/Device Name: Integra Radionics Interfix™ Patient Adapters Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: TYE Dated: August 30, 2007 Received: August 31, 2007

#### Dear Mr. O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/9 description: The image shows a circular seal with the text "1906-2006" at the top. The letters "PA" are in the center of the seal, with the word "Occidental" underneath. There are four stars below the word "Occidental". The seal is surrounded by a dotted border.

moting Public Hoults lating and OP,

{3}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely vours,

Nancy Chogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

## Indications for Use Statement

510(k) Number (if known): K072454

Device Name: Integra Radionics Interfix™ Patient Adapters Indications For Use:

For cranial fixation during CT scanning and treatment with the TomoTherapy HiArt System.

PRESCRIPTION USE X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Aomi M. Whay

Division of Reproductive, Abdominal and Radiological Device 510(k) Number

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K072454](https://fda-staging.innolitics.com/device/K072454)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com