← Product Code [CJY](/productcode/CJY) · K072440

# VITROS CHEMISTRY PRODUCTS: CA SLIDES (MODEL: 145 0261), CALIBRATORS KIT 1 (MODEL: 188 2208) (K072440)

_Ortho-Clinical Diagnostics, Inc. · CJY · Sep 20, 2007 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072440

## Device Facts

- **Applicant:** Ortho-Clinical Diagnostics, Inc.
- **Product Code:** [CJY](/productcode/CJY.md)
- **Decision Date:** Sep 20, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 862.1145
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

VITROS Chemistry Products Ca Slides For in vitro diagnostic use only. VITROS Ca Slides quantitatively measure calcium (Ca) concentration in serum, plasma, and urine. VITROS Chemistry Products Calibrator Kit 1 For in vitro diagnostic use only. VITROS Calibrator Kit 1 is used to calibrate the VITROS Chemistry Systems for the quantitative measurement of BUN/UREA, Ca, CREA, GLU, LAC, Li, Mg, PHOS, SALI, THEO, and URIC.

## Device Story

VITROS Chemistry Products Ca slide; in vitro diagnostic reagent for quantitative measurement of calcium in serum, plasma, or urine. Modification involves change in manufacturing process and formulation of spreading layer component. Device used in clinical laboratory settings by trained personnel. Output provides quantitative calcium concentration to assist in clinical diagnosis and monitoring of calcium metabolism disorders.

## Clinical Evidence

Bench testing only. Verification and validation activities performed per design control procedures to assess impact of manufacturing and formulation changes. Results demonstrated adherence to predetermined acceptance criteria.

## Technological Characteristics

In vitro diagnostic reagent slide for calcium measurement. Modification to spreading layer formulation. Fundamental scientific technology unchanged from predicate. Conformance with 21 CFR 820.30 design controls.

## Regulatory Identification

A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

## Predicate Devices

- VITROS Chemistry Products Ca slide ([K961226](/device/K961226.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SPECIAL 510(k): Device Modification

ODE Review Memorandum (Decision Making Document is Attached)

To: THE FILE

RE: DOCUMENT NUMBER k072440

This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):

1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k961226
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.

This change was for a modification in the manufacturing process. The VITROS Chemistry Products Ca slide is being modified due to a change in the formulation of a component of the slide, the spreading layer.

4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and performance characteristics.

5. A Design Control Activities Summary which includes:

a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis FMECA
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:

i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.

6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.

---

**Source:** [https://fda-staging.innolitics.com/device/K072440](https://fda-staging.innolitics.com/device/K072440)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
