← Product Code [FFK](/productcode/FFK) · K072422

# CYBERWAND STERILE PADS, MODEL: CW-RBPBX (K072422)

_Cybersonics, Inc. · FFK · Oct 24, 2007 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072422

## Device Facts

- **Applicant:** Cybersonics, Inc.
- **Product Code:** [FFK](/productcode/FFK.md)
- **Decision Date:** Oct 24, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.4480
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The CyberWand™ Lithotripter with STERILE Probes is intended to be used for the intracorporeal fragmentation of urinary tract calculi in the kidney, ureter, bladder and biliary tracts.

## Device Story

CyberWand™ Lithotripter is an electrohydraulic lithotripter used for fragmentation of urinary tract calculi. Device utilizes sterile probes to deliver energy for stone disintegration within kidney, ureter, bladder, and biliary tracts. Operated by physicians in clinical settings. Device functions by converting electrical energy into mechanical energy to break down calculi. Output is physical fragmentation of stones, facilitating removal. Clinical benefit includes minimally invasive treatment of urinary tract stones.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use.

## Technological Characteristics

Electrohydraulic lithotripter; utilizes sterile probes for energy delivery; class II device (21 CFR 876.4480); product code FFK.

## Regulatory Identification

An electrohydraulic lithotriptor is an AC-powered device used to fragment urinary bladder stones. It consists of a high voltage source connected by a cable to a bipolar electrode that is introduced into the urinary bladder through a cystoscope. The electrode is held against the stone in a water-filled bladder and repeated electrical discharges between the two poles of the electrode cause electrohydraulic shock waves which disintegrate the stone.

## Special Controls

The special control for this device is FDA's “Guidance for the Content of Premarket Notifications for Intracorporeal Lithotripters.”

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle with three wavy lines extending from its head, representing the department's mission to protect the health of all Americans and provide essential human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## OCT 2 4 2007

Mr. William Stoll QA/RA Manager Cybersonics, Inc. 5368 Kuhl Road ERIE PA 16510

Re: K072422

Trade/Device Name: CyberWand™ Lithotripter with STERILE Probes Regulation Number: 21 CFR §876.4480 Regulation Name: Electrohydraulic lithotripter Regulatory Class: II Product Code: FFK Dated: September 13, 2007 Received: October 2, 2007

Dear Mr. Stoll:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.

Sincerely yours,

Nancy Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Attachment 2

510(k)
Number
(if known)

K072422

Device Name CyberWand™ Lithotripter with STERILE Probes

Indications
for Use
The CyberWand™ Lithotripter with STERILE Probes is intended to be
used for the intracorporeal fragmentation of urinary tract calculi in the
kidney, ureter, bladder and biliary tracts.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
NEEDED

Indications for Use Statement

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801. 109)
OR
Over-The-Counter Use

(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number- K072422

16/19

---

**Source:** [https://fda-staging.innolitics.com/device/K072422](https://fda-staging.innolitics.com/device/K072422)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
