← Product Code [LPH](/productcode/LPH) · K072417

# SMITH & NEPHEW MIS HIP STEM (K072417)

_Smith & Nephew, Inc. · LPH · Jan 10, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072417

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Jan 10, 2008
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as theumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitits with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew MIS Hip Stem components are intended for single use only and are to be implanted without bone cement.

## Device Story

MIS Hip Stem; straight, tapered, proximally loading femoral stem; designed for uncemented implantation. Used in primary and revision total hip arthroplasty. Implanted by orthopedic surgeons in clinical/OR settings. Provides structural replacement for damaged femoral bone; restores hip joint function. Modular neck options allow anatomical customization. Single-use device.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design, material, and technological characteristics compared to predicate devices.

## Technological Characteristics

Straight, tapered, proximally loading femoral stem. Materials: Titanium alloy (Ti-6Al-4V) stem, Cobalt Chrome neck. Surface: Roughcoat porous coating. Sizes: 1-9. Modular neck design. Uncemented fixation. No software or electronic components.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- Anthology Hip Stem ([K052792](/device/K052792.md))
- Profemur TL Hip Stem ([K060358](/device/K060358.md))
- Mayo Conservative Hip ([K030733](/device/K030733.md))
- Lateralized Taperloc Microplasty ([K062994](/device/K062994.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K072417

# Smith & Nephew, Inc. Summary of Safety and Effectiveness MIS Hip Stem

### Contact Person and Address

Rishi Sinha Regulatory Affairs Specialist Smith & Nephew, Inc. Orthopaedic Reconstruction 1450 Brooks Road Memphis, TN 38116 (901)399-6054

Date of Summary: 08/17/2007

JAN 10 2008

Name of Device: Smith & Nephew MIS Hip Stem Common Name: MIS Hip Stem

### Device Description

The Smith & Nephew MIS Stem is a straight, tapered, proximally loading stem designed to match the geometry of the femur. The stems are proportionally sized and shaped in sizes 1 through 9 and have modular neck options to address patient anatomy. The stems are manufactured from titanium alloy (Ti-6Al-4V) and the necks are manufactured from Cobalt Chrome. Each femoral stem is proximally coated with Smith & Nephew's Roughcoat porous coating.

## Device Classification

21 CFR 888.3358 Hip joint metal semi-constrained porous-coated Uncemented prosthesis -Class II

# Indications for Use

Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as theumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitits with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew MIS Hip Stem components are intended for single use only and are to be implanted without bone cement.

# Substantial Equivalence Information

The overall design of the Smith & Nephew MIS Hip Stem is substantially equivalent to previously cleared devices listed below.

| MANUFACTURER                    | DESCRIPTION                      | 510(K)  | CLEARANCE DATE |
|---------------------------------|----------------------------------|---------|----------------|
| Smith & Nephew, Inc.            | Anthology Hip Stem               | K052792 | 10/07/2005     |
| Wright Medical Technology, Inc. | Profemur TL Hip Stem             | K060358 | 5/10/2006      |
| Zimmer Inc.                     | Mayo Conservative Hip            | K030733 | 5/1/2003       |
| Biomet Manufacturing Corp.      | Lateralized Taperloc Microplasty | K062994 | 5/9/2007       |

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle symbol on the right and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the left side of the eagle. The eagle symbol is composed of three curved lines that resemble the shape of an eagle's head and neck. The text is written in a simple, sans-serif font.

JAN 10 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Smith & Nephew, Inc. % Mr. Rishi Sinha Regulatory Affairs Specialist 1450 Brooks Road Memphis, TN 38116

Re: K072417 Trade/Device Name: MIS Hip Stem Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH Dated: January 4, 2008 Received: January 7, 2008

Dear Mr. Sinha:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. Rishi Sinha

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Mark M. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: MIS Hip Stem

Indications for Use:

Total hip components are indicated for Uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitits with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew MIS Hip Stem components are intended for single use only and are to be implanted without bone cement.

Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Valerie Owen

Page 1 of and Ne

**510(k) Number** K072417

---

**Source:** [https://fda-staging.innolitics.com/device/K072417](https://fda-staging.innolitics.com/device/K072417)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
