← Product Code [KRO](/productcode/KRO) · K072367

# REPIPHYSIS LIMB SALVAGE PROXIMAL FEMUR AND TOTAL FEMUR (K072367)

_Wrightmedicaltechnologyinc · KRO · Jan 9, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072367

## Device Facts

- **Applicant:** Wrightmedicaltechnologyinc
- **Product Code:** [KRO](/productcode/KRO.md)
- **Decision Date:** Jan 9, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3510
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The REPIPHYSIS Limb Salvage Proximal Femur and REPIPHYSIS Limb Salvage Total Femur are indicated for cemented procedures where radical resection and replacement of the proximal or total femur is required in skeletally immature patients (weighing up to 67 lbs) with osteosarcoma. REPIPHYSIS Limb Salvage Proximal and Total Femur devices are intended for cemented use only.

## Device Story

Limb salvage system for proximal or total femur replacement in skeletally immature patients; utilizes internal expansion mechanism to accommodate growth. Proximal femur configuration includes femoral stem, housing with expansion mechanism, and neck; total femur configuration adds tibial base, axial bushing, axial pin, and distal hinge femur housing. Custom assembled for each patient; implanted by orthopedic surgeons during surgical resection. Expansion mechanism allows for post-operative adjustment to match patient growth, potentially reducing need for multiple revision surgeries. Cemented fixation required.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design, material, and engineering analysis.

## Technological Characteristics

Modular femoral prosthesis; includes expansion mechanism for growth adjustment. Components: femoral stem, housing, neck, tibial base, axial bushing, axial pin, distal hinge. Materials: Metal/Polymer. Fixation: Cemented. Configuration: Custom assembled per patient. Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- REPIPHYSIS Limb Salvage System ([K072367](/device/K072367.md))
- GUARDIAN Limb Salvage System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(K) SUMMARY JAN - 9 2008 OF SAFETY AND EFFECTIVENESS

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the REPIPHYSIS® Limb Salvage Proximal Femur and Total Femur.

| Submitted By:                       | Wright Medical Technology, Inc.                                                                    |
|-------------------------------------|----------------------------------------------------------------------------------------------------|
| Date:                               | June 27, 2007                                                                                      |
| Contact Person:                     | Ehab Esmail                                                                                        |
|                                     | Director, Regulatory Affairs                                                                       |
| Proprietary Name:                   | REPIPHYSIS® Limb Salvage Proximal Femur and<br>Total Femur                                         |
| Common Name:                        | Limb Salvage System                                                                                |
| Classification Name and Reference:  | 21 CFR 888.3510 Prosthesis, Knee, Femorotibial,<br>Constrained, Cemented, Metal/Polymer - Class II |
|                                     | 21 CFR 888.3350 Prosthesis, Hip, Semi-<br>Constrained, Metal/Polymer Cemented- Class II            |
| Device Product Code and Panel Code: | Orthopedics/87/JDI                                                                                 |
|                                     | Orthopedics/87/KRO                                                                                 |

#### DEVICE INFORMATION

#### A. INTENDED USE

The REPIPHYSIS Limb Salvage Proximal Femur and REPIPHYSIS Limb Salvage Total Femur are indicated for cemented procedures where radical resection and replacement of the proximal or total femur is required in skeletally immature patients (weighing up to 67 lbs) with osteosarcoma.

REPIPHYSIS Limb Salvage Proximal and Total Femur devices are intended for cemented use only.

{1}------------------------------------------------

#### B. DEVICE DESCRIPTION

The REPIPHYSIS® Limb Salvage Proximal Femur is designed to replace the proximal femur with an expansion mechanism that has the identical internal components as the expansion mechanism in the previously cleared REPIPHYSIS® Limb Salvage System. The REPIPHYSIS® Limb Salvage Total Femur is designed to replace the entire femur with an expansion mechanism in the distal femoral portion that is identical to the expansion mechanism in the previously cleared REPIPHYSIS® Limb Salvage System.

### REPIPHYSIS® Limb Salvage Proximal Femur

The REPIPHYSIS® Limb Salvage Proximal consists of a femoral stem, a femoral housing with expansion mechanism, and a femoral neck designed to be used with all previously cleared WMT 12/14 taper femoral heads. The components of the REPIPHYSIS® Proximal Femur are substantially equivalent to the components of the GUARDIAN™ Limb Salvage System for proximal femur replacement. The REPIPHYSIS® Limb Salvage Proximal Femur is custom assembled for each patient.

## REPIPHYSIS® Limb Salvage Total Femur

The REPIPHYSIS® Limb Salvage Total Femur consists of a tibial base, axial bushing, axial pin, distal hinge femur housing with expansion mechanism, and a proximal femoral component. The components of the REPIPHYSIS® Total Femur are substantially equivalent to the components of the GUARDIAN™ Limb Salvage System for total femur replacement. The tibial base, axial bushing, axial pin, and distal hinge femur housing with expansion mechanism are identical to the components of the REPIPHYSIS® Limb Salvage System. The REPIPHY SIS® Limb Salvage Total Femur is custom assembled for each patient.

#### C. SUBSTANTIAL EQUIVALENCE INFORMATION

The material and design features of the proposed REPIPHYSIS® Proximal Femur and Total Femur are substantially equivalent to the REPIPHYSIS® Limb Salvage System. The surgical technique and intended use for the REPIPHYSIS® Proximal Femur and Total Femur are substantially equivalent to the GUARDIAN™ Limb Salvage System. The safety and effectiveness of the REPIPHYSIS® Limb Salvage Proximal Femur and Total Femur are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three overlapping wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular fashion around the eagle.

Public Health Service

JAN - 9 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Wright Medical Technology, Inc. % Mr. Ehab M. Esmail Director, Regulatory Affairs 5677 Airline Road Arlington, Tennessee 38002

Re: K072367

Trade/Device Name: REPIPHYSIS® Limb Salvage System Regulation Number: 21 CFR 888.3510 Regulation Name: Knee joint femorotibial metal/polymer constrained cemented prosthesis Regulatory Class: II Product Code: KRO, JDI Dated: December 18, 2007 Received: December 28, 2007

Dear Mr. Esmail:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 – Mr. Ehab M. Esmail

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Mullan

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K072367

Device Name:_REPIPHYSIS® Limb Salvage System

Indications For Use:

The REPIPHYSIS Limb Salvage Proximal Femur and REPIPHYSIS Limb Salvage Total Femur are indicated for cemented procedures where radical resection and replacement of the proximal or total femur is required in skeletally immature patients (weighing up to 67 lbs) with osteosarcoma.

REPIPHYSIS Limb Salvage Proximal and Total Femur devices are intended for cemented use only.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

1 of 1

Barbave Buehrn

Division of General, Restorative, and Neurological Devices

**510(k) Number** K072367

---

**Source:** [https://fda-staging.innolitics.com/device/K072367](https://fda-staging.innolitics.com/device/K072367)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).