K072329 · Biotronik, Inc. · DQY · Mar 25, 2008 · Cardiovascular
Device Facts
Record ID
K072329
Device Name
SCOUTPRO 8F, 7F AND SLITTER TOOL ADVANCED
Applicant
Biotronik, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Mar 25, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The ScoutPro CS Lead Introducer System facilitates lead implantation in the left of the heart via the coronary sinus. The ScoutPro accessories are used in conjunction with the ScoutPro CS Lead Introducer System to facilitate lead implantation in the left of the heart via the coronary sinus.
Device Story
ScoutPro family consists of introducer systems and accessories (guiding catheters, dilators, peel-away sheaths, guidewires, hemostatic valves, slitter tools) used for coronary sinus lead placement. System facilitates access to coronary sinus venous system and lead introduction into left heart. Used in clinical settings by physicians. Guiding catheters utilize braided PEBAX tubing; slitter tools facilitate lead size adjustment. Modifications include new shaping process for 7F catheters, additional tubing supplier, and updated slitter tool packaging/manual. Output is physical access/pathway for lead delivery. Benefits include improved procedural access for cardiac lead implantation.
Indicated for patients requiring coronary sinus lead implantation. Used to facilitate lead placement in the left side of the heart via the coronary sinus.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
# Modifications to ScoutPro Accessories
Special 510(k) Premarket Notification
#### 510(K) SUMMARY 1.
Name and Address of Sponsor:
BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035 MAR 2 5 2008
#### Establishment Registration Number:
1028232
#### Device Name:
Proprietary Name: Classification: Classification Name: Product Code:
Proprietary Name: Classification: Classification Name: Product Code:
Proprietary Name: Classification: Classification Name: Product Code:
ScoutPro 8F Class II (21 CFR 870.1250; 870.1310; 870.1330) Wire, Guide, Catheters, Percutaneous DQY, DRE, DQX
K 072329
ScoutPro 7F Class II (21 CFR 870.1250; 870.1310; 870.1330) Wire, Guide, Catheters, Percutaneous DQY, DRE, DQX
ScoutPro Slitter Tool Advanced Class II (21 CFR 870.1250; 870.1310; 870.1330) Wire, Guide, Catheters, Percutaneous DQY, DRE, DQX
#### General Description:
The ScoutPro family of introducer systems and accessories is specifically used for the placement of coronary sinus leads. It is designed to assist with introducing leads into the left side of the heart via the coronary sinus. The system also facilitates access to the coronary sinus venous system as well as probing the coronary sinus. The following ScoutPro accessories are subject to this Special 510(k);
The following ScoutPro 8F accessories were cleared under 510(k) #K060655 on April 4, 2006:
The basic set ScoutPro 8F contains the following components:
- . 1 hemostatic valve
- . 2 quiding catheters BIO1 and BIO2
- 1 dilator for the guiding catheter .
- . 1 peel-away sheath 11F with dilator
- . 1 guidewire
- � 1 needle
- � 1 syringe
- 2 slitter tools 4.9 F and 6.3 F for different lead sizes .
ScoutPro Sheath "Hook" contains the following components:
- . 1 guiding catheter "Hook"
- . 1 dilator for the guiding catheter
{1}------------------------------------------------
ScoutPro Sheath "Multi-Purpose Hook" contains the following components:
- . 1 guiding catheter "Multi-Purpose Hook"
- . 1 dilator for the guiding catheter
ScoutPro Sheath "Amplatz 6.0" contains the following components:
- . 1 guiding catheter "Amplatz 6.0"
- . 1 dilator for the guiding catheter
### The following ScoutPro 7Faccessories were cleared under 510(k) #K060807 on April 24, 2006:
The basic set ScoutPro 7F contains the following components:
- . 1 hemostatic valve
- 2 guiding catheters MPEP and BIO2 .
- . 1 dilator for the guiding catheter
- . 1 peel-away sheath 10F with dilator
- 1 guidewire .
- 1 needle .
- . 1 syringe
- 2 slitter tools 4.9 F and 6.3 F for different lead sizes .
- ScoutPro 7F Sheath "Hook" contains the following components: .
- o 1 guiding catheter "Hook"
- 1 dilator for the guiding catheter o
- ScoutPro 7F Sheath "Multi-Purpose Hook" contains the following components: .
- 1 guiding catheter "Multi-Purpose Hook" o
- o 1 dilator for the guiding catheter
- . ScoutPro 7F Sheath "Amplatz 6.0" contains the following components:
- 1 guiding catheter "Amplatz 6.0" o
- 1 dilator for the guiding catheter o
- . ScoutPro Hemostatic Valve
- ScoutPro Slitter Tool .
.
The following ScoutPro accessory was cleared under 510(k) #K071665 on July 18, 2007:
#### ScoutPro Slitter Tool Advanced
The following ScoutPro accessories are being introduced in this 510(k) and represent further development of the ScoutPro accessories in 510(k) submission #K060807
- . ScoutPro 7F Sheath "Extended Hook" contains the following components:
- 0 1 guiding catheter "Extended Hook"
- 1 dilator for the guiding catheter o
- ScoutPro 7F Sheath "Extended Hook Right" contains the following components: .
- 1 guiding catheter "Extended Hook Right" o
{2}------------------------------------------------
- 1 dilator for the guiding catheter o
The following ScoutPro accessories are being introduced in this 510(k) as separately available ScoutPro accessories, they are unchanged from those submitted in 510(k) submission #K060807 :
- ScoutPro 7F Sheath "MPEP" contains the following components:
- o 1 guiding catheter "MPEP"
- 1 dilator for the quiding catheter o
- ScoutPro 7F Sheath "BIO2" contains the following components: .
- o 1 guiding catheter "BIO2"
- o 1 dilator for the guiding catheter
- . ScoutPro Guidewire
#### Device Modifications:
All ScoutPro 7F guiding catheters will utilize a new shaping process and an additional supplier of the braided tubing, made out of PEBAX. No changes have been made to the chemical or physical composition of the PEBAX, only the supplier of the tubing. The additional supplier for the braided tubing will be Quan Emerted, while the original ScoutPro 7F guiding catheters were supplied by Accelent. Both suppliers may be used in the future and they both receive the PEBAX raw material and manufacture the tubes to the same specifications. As with the material from the original supplier, BIOTRONIK receives the tubes, adds the white tip and grip, and creates the shape.
These changes affect the guiding catheters "MPEP" and "BIO2" that are sold in the ScoutPro 7F kit, as well as the separately sold ScoutPro 7F Sheaths, "Hook," "Multipurpose Hook," and "Amplate 6.0." This Additionally, the "MPEP" and "BIO2" guiding catheters previously approved as part of the ScoutPro 7F kit will now also be available separately. The ScoutPro Guidewire will also be available separately from the ScoutPro 7F kit. No changes were made to the ScoutPro Guidewire.
The usage of the ScoutPro 7F kit as well as all the ScoutPro 7F guiding catheters and guidewire remains unchanged and the product characteristics such as indications for use, contraindications, and functions are identical to the previously cleared ScoutPro accessories in submission #K060807, cleared on April 24, 2006. Therefore, the previously cleared versions will serve as predicate devices for the modified products included in this Special 510(k).
Minor modifications were made to the ScoutPro Slitter Tool including packaging and changes to the Technical Manual. The ScoutPro Slitter Tool is packaged in a single, sterile package instead of double packaging. Additionally, the Technical Manual for the ScoutlPro Slitter Tool was modified to provide more accurate information relating to the use of the different slitter tools for different lead sizes.
All ScoutPro accessories in this submission, including those previously deared under #K060655, #K060807 and #K071665 and those being introduced in this submission, will utilize non-lingual labeling.
#### Predicate Devices:
BIOTRONIK proposes the following delivery system cleared through 510(k) notification as the predicate device for the new ScoutPro 7F guiding catheters and separately sold accessories after the minor modifications described in this Special 510(k):
- . BIOTRONIK's ScoutPro 8F (#K060655, cleared on April 4, 2006)
- . BIOTRONIK's ScoutPro 7F (#K060807, cleared on April 24, 2006)
- BIOTRONIK's ScoutPro Slitter Tool Advanced (#K071665, cleared on July 18, 2007)
All ScoutPro accessories introduced in this submission represent further development of the ScoutPro accessories cleared in Special 510(k) #K060655, #K060807, and #K071665.
{3}------------------------------------------------
#### Indications for Use:
: :
:
The ScoutPro CS Lead Introducer System facilitates lead implantation in the heart via the coronary sinus.
The ScoutPro accessories are used in conjunction with the ScoutPro CS Lead Introducer System to facilitate lead implantation in the left of the heart via the coronary sinus.
| Name and Address of Manufacturer: | BIOTRONIK GmbH & Co. KG<br>Woermannkehre 1,<br>12359 Berlin, Germany<br>011-49-30-689-05-1210 |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Manufacturer's Registration Number: | 9610139 |
| Name and Address of Contract Manufacturer: | BIOTRONIK AG<br>Ackerstraße 6<br>8180 Bülach, Switzerland<br>011-41-44-864-5169 |
| Contract Manufacturer's Registration Number: | 8043892 |
| Contact Person(s) and Phone Number: | Jon Brumbaugh<br>Vice President, Regulatory Affairs and Compliance<br>Phone (888) 345-0374<br>Fax (503) 635-9936 |
jon.brumbaugh@biotronik.com
、
・
:
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle-like symbol with three wing-like shapes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 5 2008
BIOTRONIK, Inc. c/o Mr. Jon Brumbaugh Vice President, Regulatory Affairs and Compliance 6024 Jean Road Lake Oswego, OR 97035
Re: K072329
ScoutPro 8F, ScoutPro 7F, ScoutPro Slitter Tool Advanced Regulation Number: 21 CFR 870.1250; 870.1310; 870.1330 Regulation Name: Wire, Guide, Catheters, Percutaneous Regulatory Class: Class II (two) Product Code: DQY, DRE, DOX Dated: March 11, 2008 Received: March 12, 2008
Dear Mr. Brumbaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Jon Brumbaugh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Difector Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):
5072329
Device Name:
ScoutPro 8F CS Lead Introducer System ScoutPro 7F CS Lead Introducer System ScoutPro Sheath "Hook" ScoutPro Sheath "Multi-Purpose Hook" ScoutPro Sheath "Amplatz 6.0" ScoutPro 7F Sheath "Hook" ScoutPro 7F Sheath "Multi-Purpose Hook" ScoutPro 7F Sheath "Amplatz 6.0" ScoutPro 7F Sheath "MPEP" ScoutPro 7F Sheath "BIO2" ScoutPro 7F Sheath "Extended Hook" ScoutPro 7F Sheath "Extended Hook Right" ScoutPro Guidewire ScoutPro Hemostatic Valve ScoutPro Slitter Tool ScoutPro Slitter Tool Advanced
#### Indications for Use:
The ScoutPro CS Lead Introducer System facilitates lead implantation in the left of the heart via the coronary sinus.
The ScoutPro accessories are used in conjunction with the ScoutPro CS Lead Introducer System to facilitate lead implantation in the left of the heart via the coronary sinus.
Prescription Use > (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number
2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.