CANCELLO-PURE BONE WEDGE, 12MM EVANS OR 6MM COTTON

K072327 · Regeneration Technologies, Inc. · MQV · Oct 24, 2007 · Orthopedic

Device Facts

Record IDK072327
Device NameCANCELLO-PURE BONE WEDGE, 12MM EVANS OR 6MM COTTON
ApplicantRegeneration Technologies, Inc.
Product CodeMQV · Orthopedic
Decision DateOct 24, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3045
Device ClassClass 2

Indications for Use

These products are indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. They are indicated to be placed into bony voids or gaps of the skeletal system (e.g., extremities, spine, ilium and/or pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a void filler that remodels into the recipient's skeletal system.

Device Story

CANCELLO-PURE™ Wedge is a bone void filler machined from bovine bone; processed via BioCleanse® Tissue Sterilization Process. Device is preshaped into a wedge for surgeon convenience. Intended for placement into non-structural bony voids or gaps in the skeletal system (extremities, spine, ilium, pelvis) resulting from surgery or trauma. Device acts as a scaffold that remodels into the recipient's skeletal system. Used by surgeons in clinical settings.

Clinical Evidence

No clinical data provided. Evidence consists of animal model remodeling studies and viral inactivation bench testing showing reduction of a viral panel to below detectable limits.

Technological Characteristics

Machined bovine bone; processed via BioCleanse® Tissue Sterilization Process. Wedge form factor. Remodeling bone void filler. No electronic or software components.

Indications for Use

Indicated for patients with bony voids or gaps in the skeletal system (extremities, spine, ilium, pelvis) not intrinsic to structural stability. Applicable to surgically created or traumatic osseous defects.

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Regeneration Technologies, Inc. The logo consists of a black square on the left, followed by the word "Regeneration" in a bold, sans-serif font. Below the word "Regeneration" is the phrase "TECHNOLOGIES, INC." in a smaller, sans-serif font. The letters are all capitalized. 11621 Research Circle Post Office Box 2650 Alachua, FL 32616-2650 USA K072327 Tel 386.418.8888 Toll Free 877.343.6832 Fax 386.418.0342 www.rtix.com # Attachment 4: 510(K) Summary August 17, 2007 Submitted by: Lisa Simpson Regeneration Technologies, Inc. 11621 Research Circle Alachua, FL 32615 Phone: 386-418-8888 x4326 Fax: 386-418-1627 Proprietary Name: CANCELLO-PURE™ Wedge Common Name: Filler, bone void, calcium compound Product Code: MQV, Orthopedics Panel Code Section: 21 CFR 888.3045 ## Substantial Equivalence: The CANCELLO-PURE™ Wedge is substantially equivalent to predicate devices in materials, design, and function. ## Description: The CANCELLO-PURE™ Wedge is machined into a wedge shape from bovine bone processed with the BioCleanse® Tissue Sterilization Process. These products are preshaped for surgeon convenience. ## Intended Use: These products are indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. They are indicated to be placed into bony voids or gaps of the skeletal system (e.g., extremities, spine, ilium and/or pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a void filler that remodels into the recipient's skeletal system. ## Summary of Technological Characteristics: The CANCELLO-PURE™ Wedge has materials, design and function equivalent to predicate devices. The source of bovine bone used in the manufacture of these products is a closed herd located in the U.S.A. The BioCleanse® Tissue Sterilization Process, used in the manufacture of these products, has been shown to render processed bovine bone capable of remodeling comparably to OCT 2 4 2007 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Regeneration Technologies, Inc. The word "Regeneration" is in large, bold, black letters with a trademark symbol in the upper right corner. Below that, in smaller letters, is the text "TECHNOLOGIES, INC." 11621 Research Circle Post Office Box 2650 Alachua, FL 32616-2650 USA Tel 386.418.8888 Toll Free 877.343.6832 Fax 386.418.0342 www.rtix.com allograft in an animal model. A viral inactivation study using a worst-case representation of the BioCleanse® process has shown reduction of a panel of viruses to below detectable limits. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Regeneration Technologies, Inc. % Ms. Lisa Simpson 11621 Research Circle P. O. Box 2650 Alachua, FL 32616-2650 OCT 2 4 2007 Re: K072327 Trade/Device Name: Cancello-Pure™ Wedge Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV Dated: September 20, 2007 Received: September 24, 2007 Dear Ms. Simpson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Lisa Simpson This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark H. Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Attachment 2: Indications for Use 510(k) Number (if known): Device Name: CANCELLO-PURE™ Wedge Indications for Use: These products are indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. They are indicated to be placed into bony voids or gaps of the skeletal system (e.g., extremities, spine, ilium and/or pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a void filler that remodel into the recipient's skeletal system. Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PACE OF NEEDED) Mark A. Milliken (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number**
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...