CAPNOSTREAM 20 WITH A2 ADAPTIVE AVERAGING SOFTWARE
K072295 · Oridion Capnography, Inc. · DQA · Oct 11, 2007 · Cardiovascular
Device Facts
Record ID
K072295
Device Name
CAPNOSTREAM 20 WITH A2 ADAPTIVE AVERAGING SOFTWARE
Applicant
Oridion Capnography, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Oct 11, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Capnostream20 combined capnograph/pulse oximeter monitor is intended to provide professionally trained health care providers the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric and adult patients in hospital type facilities, intra hospital transport and home environments.
Device Story
Capnostream20 is a bedside monitor integrating miniMediCO2 capnography and pulse oximetry modules. Device inputs include CO2 waveform data and pulse oximetry signals. The miniMediCO2 module utilizes an adaptive averaging (A2) algorithm (software version 2.31) to calculate respiration rate from CO2 waveforms; algorithm filters noise and fluctuations to reduce nuisance alarms while maintaining sensitivity to clinically significant changes. Host device displays parameters and triggers alarms when thresholds are crossed; host performs no modification to module-calculated values. Used by physicians, nurses, and trained healthcare providers in critical care, anesthesiology, intensive care, and neonatal units. Benefits include improved accuracy of respiration rate and reduced false-positive alarms, aiding clinical decision-making regarding patient respiratory status.
Clinical Evidence
Bench testing only. Software validation performed to confirm performance of the adaptive averaging algorithm and substantial equivalence to the predicate. Hazard analysis conducted to evaluate risks associated with the new algorithm.
Technological Characteristics
Two-parameter bedside monitor. Components: miniMediCO2 capnography module and pulse oximetry module. Standards: ISO 21647 (capnography), ISO 9919 (pulse oximetry), IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 60601-1-8, UL 60601-1, ISO 14971. Non-sterile. Software-controlled modules provide parameters to host for display and alarm generation.
Indications for Use
Indicated for neonatal, pediatric, and adult patients requiring continuous, non-invasive monitoring of CO2 concentration, respiration rate, functional arterial oxygen saturation (SpO2), and pulse rate in hospital, transport, or home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "K072295 CONFIDENTIAL". The text is written in a bold, sans-serif font. The word "CONFIDENTIAL" is written in all capital letters and is located below the number. The image is likely a document or record that has been marked as confidential.
# 4. 510(k) Summary of Safety and Effectiveness
This summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92
OCT 1 1 2007
### (This document is not confidential)
#### DATE THIS SUMMARY WAS PREPARED
August 15, 2007
#### SUBMITTER'S NAME AND ESTABLISHMENT ADDRESS:
Oridion Capnography Inc. 160 Gould Street Needham, MA 02494
#### ESTABLISHMENT REGISTRATION NUMBER
3003941644
#### CONTACT PERSON:
Rachel Weissbrod, Director of Regulatory Affairs
Oridion Medical 1987 Ltd.
Har Hotzvim Science Park
POB 45025
91450 Jerusalem, Israel
Telephone: +972-2-589-9115
Fax: +972-2-586-6680
#### DEVICE INFORMATION
Trade Name: Capnostream20 with A2 software Common Name: Two Parameter Bedside Monitor Classification Name: Capnograph/Pulse Oximeter Regulation Number: 868.1400, Carbon Dioxide Analyzer (Classification CCK) 870.2700 Pulse Oximeter (Classification DQA) Device Listing Number: B051971
{1}------------------------------------------------
#### PREDIČATE DEVICE
Capnostream20 with the new miniMediCO2 module adaptive averaging (A3) software is substantially equivalent to the following commercially available device:
| Manufacturer | Device | 510(k)No. | Clearance Date |
|--------------------------|---------------|-----------|----------------|
| Oridion 1987 Medical Ltd | Capnostream20 | K060065 | May 4th, 2006 |
#### DEVICE DESCRIPTION
The Capnostream20 bedside monitor is a two parameter monitor consisting of a miniMediCO2 capnography module and a pulse oximetry module implemented in a host device. The host device displays parameters received from the respective modules and generates alarms when preset alarm thresholds are crossed. The device is classified as CCK Class II according to 21 CFR § 868.1400 - Carbon Dioxide Analyzer.
This device has two modules that are classified as follows:
- 21 CFR 868.1400, Carbon Dioxide Analyzer (Classification CCK) .
- 21 CFR870.2700 Pulse Oximeter (Classification DQA) .
Each module is controlled by dedicated software that is an integral part of the respective module. Each module provides parameters to the host software (the Capnostream20 device software) which then controls the display of the received parameter values and creates alarms when the values cross the preset thresholds. The miniMediCO2 capnography module software presented in this submission includes an adaptive averaging algorithm defined as the A Algorithm for calculating the respiration rate from the CO2 waveform introduced in software version 2.31 of the miniMediCO2 capnography module software. The calculated respiration rate parameter is then provided to the host (the Capnostream20 device software). The host makes no modification to the values received from the module. The host triggers an alarm when the respiration rate high or respiration rate low thresholds have been crossed. The algorithm employed in the respiration rate calculation reduces false positive alarms by filtering out noise and instantaneous fluctuations without missing true alarms that may indicate a clinically significant change to respiration rate. By
{2}------------------------------------------------
employing the adaptive averaging algorithm, the respiration rate accurately reflects the patient's condition and significantly reduces the generation of nuisance alarms by the host.
## INTENDED USE
The Capnostream20 combined capnograph/pulse oximeter monitor is intended to provide professionally trained health care providers the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric and adult patients in hospital type facilities, intra hospital transport and home environments.
### COMPARISON TO PREDICATE DEVICE
The Capnostream20 with adaptive averaging (A?) software (miniMediCO2 software version 2.31) is identical to the predicate Capnostream20 (miniMediCO2 software version 2.20) with the exception of the algorithm changes. No changes to the host (Capnostreamon) software were made to support the new module algorithms and no significant hardware changes have been made to the device.
The new device meets the safety and performance standards met by the predicate device.
Test data are provided to validate the performance of the software and its substantial equivalence to the predicate device. The functional features and the intended use of Capnostream20 with adaptive averaging software are substantially equivalent to the predicate device.
A hazard analysis was carried out on the module with the new algorithms. This hazard analysis concluded that any residual risks were judged as acceptable when weighed against the intended benefits of use of the system.
| Attribute | Capnostream20 Bedside<br>Monitor with MiniMediCO2<br>EtCO2 module with adaptive<br>averaging (A2) software<br>(version 2.31) | Predicate Device- Capnostream20 Bedside<br>Monitor K060065 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use | The indications for use are identical<br>to the indications for use in the<br>predicate device | The Capnostream20 combined capnograph/pulse<br>oximeter monitor is intended to provide<br>professionally trained health care providers the<br>continuous, non invasive measurement and |
| | | monitoring of carbon dioxide concentration of the<br>expired and inspired breath and respiration rate,<br>and for the continuous non-invasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2 and pulse rate). It is intended<br>for use with neonatal, pediatric and adult patients<br>in hospitals, hospital type facilities, intra hospital<br>transport and home environments. |
| Target<br>population | It is intended for use with neonatal,<br>pediatric, and adult patients. | It is intended for use with neonatal, pediatric, and<br>adult patients. |
| Design | Identical to MiniMediCO2 module in<br>K060065 with the exception of<br>software version change from 2.20<br>to 2.31 | See K060065 |
| Where Used | It is to be used by physicians,<br>nurses and other trained health<br>care providers in critical care<br>patient settings, such as<br>anesthesiology, intensive care<br>medicine, neonatal Intensive care<br>and other health care areas | It is to be used by physicians, nurses and other<br>trained health care providers in critical care<br>patient settings, such as anesthesiology,<br>intensive care medicine, neonatal Intensive care<br>and other health care areas |
| Performance<br>Standards | ISO21647<br>ISO 9919 | ISO21647<br>ISO 9919 |
| Safety<br>Standards | IEC/EN60601-1<br>IEC/EN60601-1-2(2001)<br>IEC60601-1-8<br>UL60601-1<br>ISO 14971 | IEC/EN60601-1<br>IEC/EN60601-1-2(2001)<br>IEC60601-1-8<br>UL60601-1<br>ISO 14971 |
| Biocompatibility | There are no issues of<br>biocompatibility for this device and<br>no biocompatibility testing was<br>done. | There are no issues of biocompatibility for this<br>device and no biocompatibility testing was done. |
| Sterility | This device does not require<br>sterilization and is shipped marked<br>non-sterile. | This device does not require sterilization and is<br>shipped marked non-sterile. |
{3}------------------------------------------------
## CONCLUSION
Capnostream20 with adaptive averaging software does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed device. Therefore, the device is substantially equivalent to the predicate device with respect to safety effectiveness, and intended use.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.
OCT 1 1 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Oridion Capnography, Incorporated C/O Ms. Rachel Weissbrod Director of Regulatory Affairs Oridion Medical 1987 Limited Har Hotzvim Science Park Post Office Box 45025 91450 Jerusalem ISRAEL
Re: K072295
Trade/Device Name: Capnostream20 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, CCK Dated: September 16, 2007 Received: September 20, 2007
Dear Ms. Weissbrod:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Weissbrod
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sinite y. Michu Dnia
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## 3. Statement of Indications for Use
## TWO PARAMETER CAPNOSTREAM20 MONITOR WITH ADAPTIVE AVERAGING SOFTWARE
## (This document is not confidential)
## Indications for Use
August 15, 2007
# 510(k) Number (if known)」 | し フ 22 9 S
Device Name: Capnostream20
#### Indications For Use:
The Capnostream20 combined capnograph/pulse oximeter monitor is intended to provide professionally trained health care providers the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2 and pulse rate). It is intended for use with neonatal, pediatric and adult patients in hospital type facilities, intra hospital transport and home environments.
Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Snyte Y. Michael Davis
(Division Sign-Off Division of Anesthesiology, General Hospital Infection Control, Dental Devices
KO72295 510(k) Number:
Oridion Medical 1987 Ltd.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.