K072290 · Dima Italia Srl · BYT · May 29, 2008 · Anesthesiology
Device Facts
Record ID
K072290
Device Name
PEGASO V, NEGAVENT RESPIRATOR, MODEL: DA-3 PLUS
Applicant
Dima Italia Srl
Product Code
BYT · Anesthesiology
Decision Date
May 29, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5935
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The New Negavent Ventilator DA-3 Plus Pegaso V is a timed-cycled, pressure ventilator that is intended to support a patient's ventilation by alternately applying and releasing external negative pressure over the diaphragm and upper trunk of the patient. The ventilator works with a cuirass, poncho, or a ventilating chamber (Porta-Lung).
Device Story
Pegaso V is an assisted/controlled non-invasive external negative pressure ventilator. It uses an electric blower as a pressure/flow generator and a mechanical valve controlled by a step-motor to alternate positive and negative pressures over the patient's diaphragm and upper trunk. The device interfaces with a cuirass, poncho, or ventilating chamber. An optional trigger senses spontaneous patient demands, enabling Synchro/Timed, Spontaneous/Timed, and Spontaneous with Plateau modes. The system includes an 'Autoparameters Function' to evaluate respiratory frequency and I/E ratio. Operational parameters (respiratory time, pressure values) are set via a touch keyboard and displayed on an LCD. The device is intended for use in home, hospital, and institutional settings by clinicians or patients. It compensates for leaks cycle-by-cycle. By providing mechanical ventilation similar to spontaneous breathing, it supports patients with respiratory insufficiency.
Clinical Evidence
Bench testing only. Device safety demonstrated through compliance with IEC 60601-1, EN 60601-1-2, EN 60601-1-4, EN ISO 9703-3, EN 794-1, and EN 794-2 standards. No clinical data provided.
Technological Characteristics
External negative pressure ventilator; electric blower-driven; mechanical valve with step-motor and optical sensor; touch keyboard interface with LCD; 110-230Vac power; software-controlled; dimensions/weight: 14.3 lbs; operating temperature 50°F to 122°F; safety standards: EN 60601-1, EN 60601-1-2, EN 60601-1-4.
Indications for Use
Indicated for adult patients requiring ventilatory support via external negative pressure applied to the diaphragm and upper trunk. Compatible with cuirass, poncho, or ventilating chamber (Porta-Lung).
Regulatory Classification
Identification
An external negative pressure ventilator (e.g., iron lung, cuirass) is a device chamber that is intended to support a patient's ventilation by alternately applying and releasing external negative pressure over the diaphragm and upper trunk of the patient.
Predicate Devices
Lifecare Services Inc. NEV-100, Non-invasive Extrathoracic Ventilator (K910947)
Reference Devices
Dima Italia Negavent DA-3 Plus Pegaso Cough
Submission Summary (Full Text)
{0}------------------------------------------------
Owner: Dima Italia Srl Via C. Vighi 29 Bologna, Italy 40133
Lewis Ward Consultant
Prepared 2-13-08
L.W. Ward and Associates, Inc. 4655 Kirkwood Court Boulder, CO 80301 303-530-3279 303-530-4774 Fax
Device Trade Name: Pegaso V
Indications for Use:
The New Negavent Ventilator DA-3 Plus Pegaso V is a timed-cycled, pressure ventilator that is intended to support a patient's ventilation by alternately applying and releasing external negative pressure over the diaphragm and upper trunk of the patilator works with a cuirass, poncho, or a ventilating chamber (Porta-Lung).
Classification Name: Positive Pressure Intermittent Breathing Device
Classification: 868.5935, Product Code BYT
868.5935
Common Name: External Negative Pressure Ventilator
SE Predicate: Lifecare Services Inc. NEV-100, Non-invasive Extrathoracic Ventilator K910947
#### Device Description:
The Pegaso V is an assisted/controlled non-invasive ventilation system. It can work with a cuirass, poncho, or a ventilating chamber (Porta-Lung). An optional Trigger can be installed, sensing the spontaneous demands of the patient and permitting the use in Synchro/Timed, Spontaneous/Timed and Spontaneous with Plateau modes. If the Triggers are installed, the Autoparameters Function is enabled too, permitting the evaluation of the Respiratory Frequency and I/E Ratio of the patient in automatic mode.
The Pegaso V is substantially equivalent to the Lifecare NEV-100. Equivalent technologies are used.
Page 1 of 5
510(k) Summary
K072290
MAY 2 9 2008
{1}------------------------------------------------
The negative pressure ventilation is a mechanical kind of ventilation similar to the human spontaneous ventilation. The Pegaso V is designed around a blower, used as pressure and flow generator, and a mechanical valve, commanding the sign and the air pressure intensity outing to the patient. The blower takes air from the atmosphere, compresses it in order to generate a pressure/depressure controlled by electronic sensors. Leaks are compensated cycle by cycle.
The Inspiratory/Expiratory cycles are determined by the blower rotation and the mechanical valve positioning. This valve is connected to a step-motor, whose position is detected through an optical sensor. The valve lets the compressed flow go toward the patient and the depressed flow toward the atmosphere or, instead, the compressed flow to the atmosphere and the depressed flow toward the patient. The working parameters are displayed on an LCD and controlled through a touch keyboard.
The selectable parameters are:
- . Respiratory Time
- Positive Pressure Value .
- Negative Pressure Value .
Device Safety is demonstrated through meeting device safety standards IEC, EN 60601-1-2, EN ISO 9703-3, EN 794-1, and EN 794-2.
{2}------------------------------------------------
# Comparison Table
| Feature | Lifecare NEV-100 | Dima Italia Negavent DA-3<br>Plus Pegaso V |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | An external negative pressure<br>ventilator that is intended to<br>support a patient's ventilation<br>by alternately applying and<br>releasing external pressure over<br>the diaphragm and upper trunk<br>of the patient. The ventilator<br>works with a cuirass, poncho<br>(body suit), or ventilating<br>chamber (Porta-Lung). | A timed-cycled pressure<br>ventilator that is intended to<br>support a patient's ventilation by<br>alternately applying and<br>releasing external negative<br>pressure over the diaphragm and<br>upper trunk of the patient. The<br>ventilator works with a cuirass,<br>poncho (body suit), or ventilating<br>chamber (Porta-Lung). |
| Line | | |
| Voltage | 120-240Vac | 110-230Vac |
| Frequency | 50/60 Hz | 50/60 Hz |
| Power | 500W/600W | 400W |
| Use settings | Home, hospital, institution | Home, hospital, institution |
| Patient Use | Adult | Adult |
| Negative Pressure I | Adjustable -5 to -100 cm H2O | Variable from 0 to -99 cm H2O |
| Positive/Negative Pressure E | Unknown | Variable from +99 to -25 cm<br>H2O |
| Frequency | 4 to 60 bpm | Variable from 5 to 50 bpm |
| I/E Ratio | Adjustable 1:0.5 to 1:29.1 | Variable from 1/0.5 to 1/99<br>(based on present frequency and<br>adjustable to clinician's selected<br>setting) |
| Controls | Menu driven, software<br>controlled | Menu driven, software controlled |
| Inspiratory Time (Ti) | 0.5 to 5.0 seconds | 0.1 to 5.0 seconds |
| Modes | - Controlled ventilation<br>- Assisted controlled<br>ventilation | - Controlled ventilation<br>- Assisted controlled ventilation<br>- Continuous negative<br>- Assisted with Plateau |
| Alarms | - High and Low respiratory<br>pressure<br>- Power Failure<br>- Constant Pressure<br>- Internal failure | - High and Low respiratory<br>pressure<br>- Power Failure |
| Weight | 31 pounds | 14.3 pounds |
| Accessories | Patient Hose<br>Optional proximal pressure line<br>Nasal Cannula<br>Operator's Manual | Power cord (230V,110V)<br>Vacuum Airflow Tube<br>Operator's Manual |
| Electrical Safety | UL, TUV, CSA | EN 60601 |
| Feature | Emerson Cough Assist | Dima Italia Negavent DA-3<br>Plus Pegaso Cough |
| Standards | UL 747Y | EN 60601-1, EN 60601-1-2,<br>EN 60601-1-4 |
| EMC | UL, TUV, FCC Part 15, CSA,<br>Class B | EN 60601-1-2, FCC Part 15,<br>Class B |
| CE Conformity | Not listed | Risk Class IIb<br>93/42 EEC Directive<br>CE 0476 |
| Environmental Temperature | 41° F to 104° F | 50° F to 122° F |
{3}------------------------------------------------
510(k) Summary Page 4 of 5
.
. . . . . .
.
.
:
:
:
{4}------------------------------------------------
## Pegaso V Similarities and Differences
The Dima Italia ventilator is substantially equivalent to the Lifecare NEV-100 Non-invasive Extrathoracic Ventilator.
#### Similarities:
1. The products have equivalent Indications for Use.
2. The fundamental technology is the same. An electric powered blower produces positive and negative pressures. The pressures are alternately applied/released over the diaphragm and upper trunk of the patient.
3. Both devices meet safety evaluations under IEC 60601 standards.
4. Pressures developed are comparable.
- 5. Both units are available in 110V and 220/230V versions.
- 6. Both units are intended for adult patients.
- 7. Use settings for home, hospital, and institution are the same.
- 8. Frequencies of breaths are comparable.
- 9. Both systems are software controlled.
### Differences:
No major differences.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 9 2008
Dima Italia SRL C/O Mr. Lewis Ward Consultant L.W. Ward and Associates, Incorporated 4655 Kirkwood Court Boulder, Colorado 80301
Re: K072290
Trade/Device Name: Pegaso V Regulation Number: 21 CFR 868.5935 Regulation Name: External Negative Pressure Ventilator Regulatory Class: II Product Code: BYT Dated: May 20, 2008 Received: May 23, 2008
Dear Mr. Ward:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Ward
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known):
Device Name: Pegaso V
Indications for Use:
The New Negavent Ventilator DA-3 Plus Pegaso V is a timed-cycled, pressure ventilator that is intended to support a patient's ventilation by alternately applying and releasing external negative pressure over the diaphragm and upper trunk of the patient. The ventilator works with a cuirass, poncho, or ventilating chamber (Porta-Lung).
Prescription Use X (Part 21 CFR 801 Subpart D) Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kiri Te Kanawa
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K072296
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.