← Product Code [MQV](/productcode/MQV) · K072265

# PROGENIX DMB PUTTY (K072265)

_Medtronic Sofamor Danek · MQV · Jan 9, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072265

## Device Facts

- **Applicant:** Medtronic Sofamor Danek
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Jan 9, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

PROGENIX™ DBM Putty is intended for use as a bone graft substitute in bony voids or gaps of the skeletal system not intrinsic to the stability of the bony structure (i.e. spine, pelvis and extremities). The voids or gaps may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PROGENIX™ DBM Putty provides a bone void filler that is resorbed/remodeled and is replaced by host bone during the healing process. When used in the extremeties or pelvis, the device is used by itself. When used in the spine, the device must be mixed with autograft bone and used as a bone graft extender.

## Device Story

PROGENIX™ DBM Putty is a bone graft substitute composed of human demineralized bone matrix (DBM), bovine collagen, and sodium alginate in phosphate buffered saline. Provided as a ready-to-use, malleable putty, it is manipulated by surgeons to fill bony voids or gaps. The device is osteoconductive and osteoinductive, facilitating bony ingrowth while resorbing during the healing process. It is not designed to provide mechanical stability. In spinal procedures, it functions as an extender when mixed with autograft; in the pelvis and extremities, it is used as a standalone filler. The device benefits patients by promoting bone healing in non-structural defects.

## Clinical Evidence

No human clinical data provided. Evidence consists of a rabbit spinal fusion study demonstrating effectiveness as a bone graft extender and enhancer, and a validated athymic nude rat assay used to confirm osteoinductive potential of every lot via histologic evidence of osteoblasts, chondroblasts, and/or woven bone.

## Technological Characteristics

Composition: Human demineralized bone matrix, bovine collagen, sodium alginate, phosphate buffered saline. Form: Malleable, ready-to-use putty. Principle: Osteoconductive and osteoinductive bone void filler. Resorption: Remodeled and replaced by host bone. Sterilization: Not specified.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- PROGENIX™ DBM Putty ([K060794](/device/K060794.md))
- Connexus Putty ([K050690](/device/K050690.md))
- Actifuse™ ABX E-Z-fil Putty Bone Graft Substitute ([K071206](/device/K071206.md))
- DBX Demineralized Bone Matrix Putty and Paste ([K040262](/device/K040262.md))
- ALLOMATRIX® Putty ([K041168](/device/K041168.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K072265

# Medtronic Sofamor Danek PROGENIX™ DBM Putty 510(k) Summarv July 2007

-Company: Medtronic Sofamor Danek USA 1800 Pyramid Place JAN - 9 2008 Memphis, TN 38132 Telephone: (901) 396-3133 Fax: (901) 346-9738

Contact: Christine Scifert Director, Regulatory Affairs

- ll. PROGENIX™ DBM Putty Proposed Proprietary Trade Name: Classification Name: Bone Void Filler Product Code: MQV Regulation No .: 888.3045
#### III. Product Description/Purpose of Application

PROGENIX™ DBM Putty contains human demineralized bone matrix (DBM) in a biocompatible carrier. The carrier is a mixture of bovine collagen with a natural polysaccharide (sodium alginate). The components are mixed in phosphate buffered saline to achieve a flowable or moldable consistency.

PROGENIX™ DBM Putty is a single use product intended for use as a bone graft substitute, bone graft extender, and bone void filler in bony voids or gaps of the skeletal system (i.e. spine, pelvis and extremities) not intrinsic to the stability of the bony structure. Additionally, this product is not designed to impart any mechanical strength to the surgical site. PROGENIX™ DBM Putty is provided in ready-to-use malleable forms that may be molded or manipulated by the surgeon into various shapes. This product has been shown to be osteoconductive as well as osteoinductive in an athymic rat assay, allowing for bony ingrowth across the graft site while resorbing at a rate consistent with bony healing.

The purpose of this 510(k) application is to expand the indication for the PROGENIX™ DBM Putty device so that it may be used as a bone graft extender and to include use in spine fusion procedures, in addition to the previously

{1}------------------------------------------------

cleared use as a bone graft substitute and bone void filler in bony voids or gaps of the pelvis, illium and extremities not intrinsic to the stability of the bony structure. The results of the included rabbit study suggest that PROGENIX™ DBM Putty is effective in producing a spinal fusion by radiographic and manual palpation criteria in an extender and enhancer mode.

#### IV. Indications

PROGENIX™DBM Putty is intended for use as a bone graft substitute in bony voids or gaps of the skeletal system not intrinsic to the stability of the bony structure (i.e. spine. pelvis and extremities). The voids or gaps may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PROGENIX™ DBM Putty provides a bone void filler that is resorbed/remodeled and is replaced by host bone during the healing process. When used in the extremeties or pelvis, the device is used by itself. When used in the spine, the device must be mixed with autograft bone and used as a bone graft extender.

#### V. Substantial Equivalence

Documentation is provided which demonstrates PROGENIX™ DBM Putty to be substantially equivalent to previously cleared bone void fillers such as PROGENIX™ DBM Putty (Medtronic Sofamor Danek, K060794, SE 12/18/06), Connexus Putty (K050690, SE 07/07/05), Actifuse™ ABX E-Z-fil Putty Bone Graft Substitute (ApaTech Limited, K071206, SE 5/31/07), DBX Demineralized Bone Matrix Putty and Paste (Musculoskeletal Transplant Foundation, K040262, SE 03/17/05), and ALLOMATRIX® Putty (Wright Medical Technology, K041168, SE 08/02/04).

#### VI. Osteoinductivity Potential

All DBM used in the preparation of PROGENIX™ DBM Putty must induce bone formation when evaluated in a validated athymic nude rat assay. Additionally, every lot of PROGENIX™ DBM PUTTY must also induce bone formation in this assay system prior to being released for use. In both the raw material and final product screening, every lot must show histologic evidence of osteoinduction through the presence of osteoblasts, chondroblasts, and/or woven bone. Osteoinduction assay results using the 

{2}------------------------------------------------

athymic rat assay should not be interpreted to predict clinical performance in human subjects.

### Viral Inactivation VII.

PROGENIX™ DBM Putty is produced from tissue and collagen which undergoes processing steps validated to inactivate a panel of viruses representative of those which are clinically relevant. The cortical bone used to produce the DBM undergoes a proprietary process demonstrated to inactivate viruses. Furthermore, the DBM undergoes additional steps which are also effective at inactivating viruses. The viral inactivation testing demonstrates suitable viral inactivation potential of the processing methods for a wide range of potential human viruses. These processing steps further reduce the risk of viral contamination beyond donor screening and testing.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

JAN - 9 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medtronic Sofamor Danek % Ms. Christine Scifert Associate Regulatory Affairs Specialist 1800 Pyramid Place Memphis, TN 38132

Re: K072265

Trade/Device Name: PROGENIX™ DBM Putty Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV and MBP Dated: December 5, 2007 Received: December 7, 2007

Dear Ms. Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

## Page 2 - Ms. Christine Scifert

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark M. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):

K072265

Device Name: PROGENIX™ DBM Putty

Indications for Use:

PROGENIX™ DBM Putty is intended for use as a bone graft substitute in bony voids or gaps of the skeletal system not intrinsic to the stability of the bony structure (i.e. spine, pelvis and extremities). The voids or gaps may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PROGENIX™ DBM Putty provides a bone void filler that is resorbed/remodeled and is replaced by host bone during the healing process. When used in the extremeties or pelvis, the device is used by itself. When used in the spine, the device must be mixed with autograft bone and used as a bone graft extender.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Oalua Buem

Division of General, Restora and Neurological Devices

**510(k) Number** k072265

---

**Source:** [https://fda-staging.innolitics.com/device/K072265](https://fda-staging.innolitics.com/device/K072265)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
