K072237 · Siemens Medical Solutions USA, Inc. · LNH · Oct 10, 2007 · Radiology
Device Facts
Record ID
K072237
Device Name
MAGNETOM VERIO
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Oct 10, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MAGNETOM Verio is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and / or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and / or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM Verio may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.
Device Story
MAGNETOM Verio is a 3 Tesla superconducting MR scanner; utilizes Total Imaging Matrix (Tim) technology and Matrix Coils. Inputs include magnetic resonance signals from patient anatomy; system transforms signals into transverse, sagittal, coronal, and oblique cross-sectional images or spectra. Operated by trained clinicians in clinical settings; supports interventional procedures using MR-safe needles and in-room displays. Output images/spectra interpreted by physicians to assist in diagnosis. Benefits include high-field diagnostic imaging capabilities for head, body, and extremities.
Clinical Evidence
Bench testing only. Performance and safety data measured per NEMA, IEC, and ISO standards. Metrics included signal-to-noise ratio, geometric distortion, image uniformity, slice thickness, and spatial resolution. Performance compared directly to MAGNETOM Trio a Tim System to confirm equivalence.
Technological Characteristics
3 Tesla superconducting MR scanner; utilizes Total Imaging Matrix (Tim) technology and Matrix Coils. Software platform: syngo MR B15B (I-class and T-class releases). Conforms to NEMA, IEC, and ISO standards for MR safety (static field, acoustic noise, dB/dt, RF heating, biocompatibility).
Indications for Use
Indicated for patients requiring diagnostic imaging of head, body, or extremities via MRDD. Used for cross-sectional imaging, spectroscopic imaging, and interventional procedures with MR-compatible devices. Contraindications include patients incompatible with MR environments.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K072237
## Section 5: 510(k) Summary
# 5 510(k) Summary
001 1 0 2007
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
#### I. General Information
| Establishment | Siemens Medical Solutions. Inc.<br>51 Valley Stream Parkway<br>Malvern. PA 19355 | |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Registration Number | 2240869 | |
| Manufacturer | Siemens AG. Bereich Med<br>Henkestrasse 127<br>D-91052 Erlangen. Germany | |
| Registration Number | 8010024 | |
| Contact Person | Ms. Judy Campbell<br>Technical Specialist. Regulatory Submissions<br>51 Valley Stream Parkway<br>Malvern. PA 19355<br>Phone: (610) 448-4918<br>Fax: (610) 448-1787 | |
| Device Name | Trade Name: MAGNETOM Verio | |
| | Classification Name:<br>CFR Code:<br>Classification: | Magnetic Resonance Diagnostic<br>Device<br>21 CFR § 892.1000<br>Class II |
#### Performance Standards
None established under Section 514 the Food. Drug. and Cosmetic Act.
{1}------------------------------------------------
## Section 5: 510(k) Summary
### II. Safety and Effectiveness Information Supporting Substantial Equivalence.
### Intended Use
The MAGNETOM Verio is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and / or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and / or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM Verio may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.
#### Device Description
The MAGNETOM Verio comes with I-class and T-class releases based on syngo MR B15B. It is a 3 Tesla MR scanner based on a superconducting magnet. It consists of the same types of hardware (with a modified magnet, gradient coil and RF body coil) that are currently available with the MAGNETOM Trio a Tim system including Matrix Coils and Total Imaging Matrix (Tim) technology.
#### Substantial Equivalence
The system is substantially equivalent to the following cleared medical devices:
| Predicate Device Name | FDA Clearance<br>Number | FDA Clearance Date |
|------------------------------------------|-------------------------|--------------------|
| Siemens 3T MAGNETOM Trio a<br>Tim System | K050200 | Feb 28, 2005 |
| software syngo MR B15 (VB15A) | K062454 | Nov 3, 2006 |
{2}------------------------------------------------
#### General Safety and Effectiveness Concerns:
Operation of the MAGNETOM Verio is substantially equivalent to the commercially available MAGNETOM 3 T Trio a Tim System.
As specified in the FDA guidance document "Guidance for the Submission Of Premarket Notifications for Magnetic Resonance Devices" the following measurements of performance and safety data have been performed following National Electrical Manufacturing Association (NEMA) or equivalent IEC and ISO standards.
#### Safety
Static Field Strength Acoustic Noise dB/dt RF Heating Biocompatibility
#### Performance
Signal-to-Noise Ratio Geometric Distortion Image Uniformity Slice Thickness Spatial Resolution
The MAGNETOM Verio will conform to the measurement of safety parameters to the international IEC, ISO and NEMA standards for safety issues with Magnetic Resonance Imaging Diagnostic Devices.
Furthermore performance measurements have been done on MAGNETOM Verio and MAGNETOM Trio a Tim System to show that the performance of the MAGNETOM Verio is equivalent with respect to the predicate device MAGNETOM Trio a Tim System.
This will assure that the performance of MAGNETOM Verio is substantially equivalent with respect to the currently available MAGNETOM Trio a Tim System.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure, with three wavy lines forming the body and head.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### 007 @ 2007
Ms. Judy Campbell Regulatory Technical Specialist Siemens Medical Solutions USA. Inc. Corporate Headquarters 51 Valley Stream Parkway MALVERN PA 19355
Re: K072237
Trade/Device Name: MAGNETOM Verio Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: August 8, 2007 Received: August 13, 2007
#### Dear Ms. Campbell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1796, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 4 Indications for Use Statement
510(k) Number (if known)
K072231
Device Name: MAGNETOM Verio
Indications for Use:
The MAGNETOM Verio is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and / or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM Verio may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.
(please do not write below this line- continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation
Prescription Use
OR Over-The-Counter Use
Division of Reproductive, Abdomina Radiological Devic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.