K072190 · Mast Biosurgery · FTL · Jan 16, 2008 · General, Plastic Surgery
Device Facts
Record ID
K072190
Device Name
ORTHO-WRAP BIORESORBABLE SHEET
Applicant
Mast Biosurgery
Product Code
FTL · General, Plastic Surgery
Decision Date
Jan 16, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The resorbable Protective Film minimizes tissue attachment to the device in case of direct contact with the tissues. The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is also indicated for reinforcement of soft tissues repaired by suture or suture anchors, during tendon repair surgery, including reinforcement of the flexor or extensor tendons of the fingers, thumb and wrist, rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. MAST Biosurgery Ortho-Wrap is NOT intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the flexor or extensor tendons of the fingers, thumb and wrist, rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair. The MAST Biosurgery Ortho-Wrap reinforces soft tissue and provides a remodelable scaffold that is replaced by the patients own soft tissue. The device is indicated for open and laparoscopic / endoscopic procedures. Laparoscopic / endoscopic procedures are limited to sizes from 0.02mm - 0.2mm in thickness.
Device Story
Ortho-Wrap Bioresorbable Sheet is a resorbable implant manufactured from polylactic acid (PLA). Provided as solid sheets (25mm x 25mm to 500mm x 500mm; 0.02mm to 1.0mm thickness). Device is malleable when heated to ~55°C (e.g., in sterile hot water), allowing 3D conformation to anatomical structures. Used in open or laparoscopic/endoscopic orthopedic surgeries by surgeons to manage/protect tendon injuries and reinforce soft tissue repairs. Can be cut to size and fixated with sutures. Acts as a remodelable scaffold replaced by patient's own tissue; minimizes tissue attachment. Does not provide primary biomechanical strength; used in conjunction with sutures or bone anchors. Benefits include tissue protection and reinforcement during healing.
Clinical Evidence
No clinical data. Safety and efficacy supported by bench testing (viscosity stability under heat, aging tests, mechanical strength testing) and animal studies.
Technological Characteristics
Material: Polylactic acid (PLA). Form: Resorbable, malleable sheet. Dimensions: 25mm x 25mm to 500mm x 500mm; 0.02mm to 1.0mm thickness. Principle: Thermal malleability at ~55°C for anatomical conformation. Sterilization: Not specified. Connectivity: None (implantable).
Indications for Use
Indicated for management and protection of tendon injuries without substantial tissue loss and reinforcement of soft tissues repaired by suture/anchors during tendon repair surgery (e.g., flexor/extensor tendons of fingers, thumb, wrist, rotator cuff, patellar, Achilles, biceps, quadriceps). Not for replacing body structure or providing primary biomechanical strength. Indicated for open and laparoscopic/endoscopic procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
# ADMINISTRATIVE INFORMATION
| Manufacturer Name: | MAST Biosurgery, Inc.<br>6749 Top Gun Street, Suite 108<br>San Diego, CA 92121 | JAN 16 2008 |
|--------------------|----------------------------------------------------------------------------------------------|-------------|
| Official Contact: | Kenneth K. Kleinhenz<br>Regulatory Affairs<br>Telephone (858) 458-0900<br>Fax (858) 458-0994 | |
# DEVICE NAME
| Classification Name: | Surgical Mesh, Polymeric |
|-------------------------|--------------------------------|
| Trade/Proprietary Name: | Ortho-Wrap Bioresorbable Sheet |
# ESTABLISHMENT REGISTRATION NUMBER 3004661493
# DEVICE CLASSIFICATION AND PRODUCT CODE
As shown in 21CFR 878.3300, Surgical Mesh are polymeric screens intended to be implanted to reinforce soft tissues. These devices are classified as Class II. Surgical Mesh have been assigned Product Code FTL and FTM.
### INTENDED USE
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The resorbable Protective Film minimizes tissue attachment to the device in case of direct contact with the tissues.
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is also indicated for reinforcement of soft tissues repaired by suture or suture anchors, during tendon repair surgery, including reinforcement of the flexor or extensor tendons of the fingers, thumb and wrist, rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. MAST Biosurgery Ortho-Wrap is NOT intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the flexor or extensor tendons of the fingers, thumb and wrist, rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the The MAST Biosurgery Ortho-Wrap reinforces soft tissue and provides a tendon repair. remodelable scaffold that is replaced by the patients own soft tissue.
The device is indicated for open and laparoscopic / endoscopic procedures. Laparoscopic / endoscopic procedures are limited to sizes from 0.02mm - 0.2mm in thickness.
Ko72190
12
{1}------------------------------------------------
#### 510(k) Summary Ortho-Wrap Bioresorbable Sheet
K072190 2/3
### DEVICE DESCRIPTION
### Design Characteristics
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is a resorbable implant in sheet form manufactured from polylactic acid (PLA). The Ortho-Wrap Bioresorbable Sheet can be cut with scissors to the desired shape and size. The Ortho-Wrap Bioresorbable Sheet is fully malleable when heated to approximately 55℃ (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The Ortho-Wrap Bioresorbable Sheet can be used either alone or in conjunction with soft tissue fixation devices such as resorbable sutures, which can also serve to fixate the Ortho-Wrap Bioresorbable Sheet and prevent dislocation.
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is provided in various shapes such as rectangles, ovals, and circles and will be provided in other shapes and sizes as needed for particular surgical procedures. The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is provided in sheets of 25mm x 25mm to 500mm x 500mm and will be provided in other shapes and sizes as needed for particular surgical procedures. The thickness of the MAST Biosurgery Ortho-Wrap Bioresorbable Sheet ranges from 0.02 mm to 1.0 mm according to the region to be treated. The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is provided in solid sheets. The borders of the sheets may be aligned with holes to attach suture material.
# Material Composition
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is fabricated from polylactic acid (PLA).
### In Vitro Testing
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is intended to be heated in the surgical suite to temperatures above the material's glass transition temperature to facilitate shaping to anatomic structures. Therefore, testing was performed to determine the effect of prolonged heating in saline at 60℃ on inherent viscosity. The testing demonstrates that viscosity stayed within an appropriate range over 120 minutes. The relatively brief exposure anticipated during the surgical preparation of MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is not expected to have a significant effect on its mechanical properties.
Aging testing was performed on MAST Biosurgery Ortho-Wrap Bioresorbable Sheet. Testing demonstrated that the MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is strong enough for the indications for use.
Mechanical testing was performed on the MAST Biosurgery Ortho-Wrap Bioresorbable Sheet which determined the MAST Biosurgery Ortho-Wrap Bioresorbable Sheet to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions.
### In Vivo Testing
Animal studies were conducted to demonstrate safety and efficacy of the MAST Biosurgery Ortho-Wrap Bioresorbable Sheet material. The animal studies demonstrated that the MAST Biosurgery Ortho-Wrap Bioresorbable Sheet materials are safe and efficacious for the indications for use.
{2}------------------------------------------------
### EQUIVALENCE TO MARKETED PRODUCT
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to pre-amendment devices: MAST Biosurgery Surgi-Wrap MAST Tendon Sheet (K063648), S. Jackson Supra FOIL Smooth Nylon foil Sheets (K973379), and Avanta Orthopaedics Tendon Spacer (K964359); Class II medical devices that were cleared for marketing in the United States under K063648, K973379, and K964359 respectively.
#### Indications For Use
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet shares indications for use principles with the predicate devices as all devices are indicated for use in the same orthopedic surgery procedures involving tendons and tendon repair. Furthermore, the MAST Biosurgery Ortho-Wrap Bioresorbable Sheet shares identical indications for use language with each of the predicate devices.
#### Design and Materials
The physical designs of the MAST Biosurgery Ortho-Wrap Bioresorbable Sheet and the MAST Biosurgery Surgi-Wrap MAST Bioresorbable Sheet (K061473), the S. Jackson Supra FOIL Smooth Nylon foil Sheets (K973379), and the Avanta Orthopaedics Tendon Spacer (K964359) predicate devices are substantially equivalent as they are all flexible and malleable implants fabricated from a polymeric material and provided in various sizes that they can be cut to shape intra-operatively and subsequently placed on, under, and around tendon anatomy. The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet and the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet predicate (K063648) share the common design feature of being bioresorbable sheets as they are both made from the identical bioresorbable polylactic acid (PLA) material and they are both fully contourable when heated to approximately 55℃. The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is substantially equivalent to the S. Jackson Supra FOIL Smooth Nylon foil Sheets (K973379) and MAST Biosurgery Surgi-Wrap MAST Bioresorbable Sheet (K061473) predicates as they are all thin, semi-rigid sheets provided in sizes ranging from 25mm x 25mm and 500mm x 500mm with various thicknesses ranging from 0.02mm to 1.0mm. The mechanical characteristics of the MAST Biosurgery Ortho-Wrap Bioresorbable Sheet are substantially equivalent to the predicate devices with respect to mechanical characteristics.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol of medicine and healing, with three figures intertwined.
JAN 16 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAST Biosurgery, Inc. % Mr. Kenneth K. Keinhenz Regulatory Affairs 6749 Top Gun Street, Suite C San Diego, California 92121
Re: K072190
Trade/Device Name: Ortho-Wrap Bioresorbable Sheet Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL Dated: January 3, 2008 Received: January 7, 2008
Dear Mr. Keinhenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Kenneth K. Keinhenz
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number : K072190
#### Device Name: Ortho-Wrap Bioresorbable Sheet
#### Indications for Use:
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The resorbable Protective Film minimizes tissue attachment to the device in case of direct contact with the tissues.
The MAST Biosurgery Ortho-Wrap Bioresorbable Sheet is also indicated for reinforcement of soft tisssues repaired by suture or suture anchors, during tendon repair surgery, including reinforcement of the flexor or extensor tendons of the fingers, thumb and wrist, rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. MAST Biosurgery Ortho-Wrap is NOT intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the flexor or extensor tendons of the fingers, thumb and wrist, rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair. The MAST Biosurgery Ortho-Wrap reinforces soft tissue and provides a remodelable scaffold that is replaced by the patients own soft tissue.
The device is indicated for open and laparoscopic / endoscopic procedures. Laparoscopic / endoscopic procedures are limited to sizes from 0.02mm - 0.2mm in thickness.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**(Division Sign-Off)**
Division of General, Restorative,
and Neurological Devices
Page 1 of 1
510(k) Number /
$$2072190$$
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.