# AIRISTAR AIR PURIFICATION SYSTEM, MODEL 500 & 1000 (K072185)

_Airistar Technologies, LLC · FRF · Mar 27, 2008 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K072185

## Device Facts

- **Applicant:** Airistar Technologies, LLC
- **Product Code:** [FRF](/productcode/FRF.md)
- **Decision Date:** Mar 27, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5045
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

Airistar Air Purification Systems (Model 1000) are intended for use in filtering airborne particles from air for medical purposes.

## Device Story

Airistar Air Purification System (Models 500 and 1000) functions as medical air cleaner; removes particulate matter from air. Device utilizes 6-stage filter system; 3 fan speeds. Model 500 includes software-controlled display panel for run times, fan speeds, and filter life indicator; Model 1000 uses manual power switch and control knobs. Used in homes, hospitals, nursing homes, schools, and offices. Operates by drawing air through filtration stages to improve air quality. Benefits patient by reducing airborne particulate exposure. Healthcare providers or users monitor filter life via display (Model 500) or manual settings to maintain air purification performance.

## Clinical Evidence

Bench testing only. Testing confirmed particulate removal efficiency and safety of UV lamp components (zero ozone production). No clinical data provided.

## Technological Characteristics

Medical recirculating air cleaner (21 CFR 880.5045). Features 6-stage filtration system, 3-speed fan. Model 500 includes digital display panel for run-time programming and filter life monitoring. Model 1000 uses manual control knobs. Electrical operation. Filters listed by Underwriters Laboratory.

## Regulatory Identification

A medical recirculating air cleaner is a device used to remove particles from the air for medical purposes. The device may function by electrostatic precipitation or filtration.

## Predicate Devices

- Hepa-Care Air Cleaner, Model HC400F & HC400E-UV ([K984116](/device/K984116.md))
- MOBILE PARTICULATE CONTAMINATION CONTROL SYSTEM, Airex, Inc. ([K023693](/device/K023693.md))
- Advanced Air Cleaner System, Healthway Products ([K012549](/device/K012549.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) SUMMARY

Airistar Technologies, L.L.C. 2330 Ernie Krueger Circle Waukegan, IL 60087 847-775-8018 847-775-8019 (fax) 800-755-8006 (toll-free) info@airistar.com Traditional 510(k) Premarket Submission Contact: Roy Kibbe, President Date prepared: March 12, 2008

K072185

#### Air Filtration Systems - Air Cleaners - Air Purifiers Common Name: Proprietary Name: Airistar Air Purification System (Model 500 & Model 1000)

### Classification:

Regulation # 880.5045; Class 2 Product Code FRF

Description: The Airistar Air Purification Systems - Model 1000 are medical air cleaners designed to help remove particulate matter from the air. These systems are for home use and use in hospitals, nursing homes, schools, offices, etc.

The Model 500 Airistar Air Purification System contains software and a display panel to program runs times and fan speeds, the panel also displays filter life indicator bar. The Model 1000 Airistar Air Purification System is operated via a power switch and control knobs to adjust the fan speed. Both models have 3 fan speeds and contain a 6 Stage filter system.

Intended Use: Airistar Air Purification Systems (Model 1000) are intended for use in filtering airborne particles from air for medical purposes.

Predicate Devices: The Airistar Air Purification System is substantially equivalent to the following predicate devices.

Hepa-Care Air Cleaner, Model HC400F & HC400E-UV by Abatement Technologies, Inc. 510(k) #: K984116

MOBILE PARTICULATE CONTAMINATION CONTROL SYSTEM, Airex, Inc. 510(k) #: K023693

Advanced Air Cleaner System, Healthway Products, 510(k) #: K012549

{1}------------------------------------------------

# Safety and Effectiveness and Summary and Conclusion Regarding Substantial Equivalence:

By definition, a device is substantially equivalent when the device has the same intended use and the Dy delimition, a al characteristics as the predicate device, or has the same intended use and different technological characteristics, but it can be demonstrated that the device is as safe and effective as the predicate device and the new device does not raise different questions regarding safety and effectiveness as compared to the predicate device.

The minor differences between the Airistar Air Purification Systems (Model 500 & Model 1000) and the predicate devices cited do not raise any additional questions regarding safety and effectiveness. The intended use of the Airistar Air Purification Systems (Model 500 & Model 1000) is the same as the intended use of the predicate devices.

The device, as designed, is as safe and effective as the predicate device, and the device is determined to be substantially equivalent to the referenced predicate device.

Bench testing has confirmed the devices' ability to remove particles efficiently. The filters have been tested and listed by Underwriters Laboratory testing showed the UV lamp produced zero ozone.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird symbol, composed of three curved lines, representing the department's commitment to health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the bird symbol. The logo is presented in black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Airistar Technologies LLC C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW Buffalo, Minnesota 55313

MAR 2 7 2008

Re: K072185

Trade/Device Name: Airistar Air Purification Systems - Model 500 & Model 1000 Regulation Number: 21 CFR 880.5045 Regulation Name: Medical Recirculating Air Cleaner Regulatory Class: II Product Code: FRF Dated: March 13, 2008 Received: March 14, 2008

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 -- Mr. Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

K072185 510(k) Number (if known):

Device Name: Airistar Air Purification Systems - Model 500 & Model 1000

Indications For Use: Intended for use in filtering airborne particles from air for medical purposes.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sheila A Murphy
(Division Sign-Off)

Page 1 of 1

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K022185

---

**Source:** [https://fda-staging.innolitics.com/device/K072185](https://fda-staging.innolitics.com/device/K072185)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com