HAPPYNEB II, HAPPYNEB III, SPEEDY, NEBBY PLUS AND MYNEB, MODEL 5
K072163 · 3A Health Care S.R.L. · CAF · Mar 6, 2008 · Anesthesiology
Device Facts
Record ID
K072163
Device Name
HAPPYNEB II, HAPPYNEB III, SPEEDY, NEBBY PLUS AND MYNEB, MODEL 5
Applicant
3A Health Care S.R.L.
Product Code
CAF · Anesthesiology
Decision Date
Mar 6, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The intended use of the Happyneb II, Happyneb III, Speedy, Nebby Plus and the Myneb is to spray liquids in aerosol form into gases that are delivered directly to adult or pediatric patients who have been prescribed inhalation therapy or medication. Each of these nebulizers must be used exclusively with their own nebulizer and mouthpiece. The nebulizers Happyneb II, Happyneb III, Speedy, and the Nebby Plus are intended to be used primarily by patients in the home care market, although they may also be used by trained hospital staff personnel as well. The Myneb model is intended to be used only in home health care. The Happyneb II, Happyneb III, Speedy, Nebby Plus and the Myneb are intended for multiple use, are non-sterile and for use with pharmaceutical products which are under physician prescription.
Device Story
Family of five nebulizer systems (Happyneb II, Happyneb III, Speedy, Nebby Plus, Myneb) designed to convert liquid medication into aerosol for inhalation. Four models (Happyneb II, Happyneb III, Speedy, Nebby Plus) utilize AC power; one model (Myneb) utilizes a DC motor with rechargeable battery and external charger. Systems use Fasterjet or Nebjet nebulizers to deliver aerosolized medication. Intended for home care use by patients, with AC models also suitable for use by trained hospital staff. Devices are non-sterile and intended for multiple use. Operation involves connecting the compressor to the specific nebulizer and mouthpiece to deliver prescribed pharmaceutical products. Clinical benefit is the effective delivery of aerosolized medication to the respiratory tract.
Clinical Evidence
Bench testing only. Performance evaluated per FDA 'Reviewer guidance for nebulizers, metered dose inhalers, spacers and actuators' (1993). Results demonstrate equivalent performance parameters (MMAD, GSD, FPF, PFD) compared to predicate devices. No clinical data required.
Technological Characteristics
Compressor-based nebulizer systems. AC-powered (4 models) or DC-powered with rechargeable battery (1 model). Plastic housings. Uses Fasterjet or Nebjet nebulizers. Non-sterile, multiple-use devices. No software or electronic algorithms.
Indications for Use
Indicated for adult and pediatric patients requiring aerosolized inhalation therapy or medication as prescribed by a physician.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
CONFIDENTIAL
K072163
## 510(k) Summary for the 3A Healthcare nebulizers: Happyneb II, Happyneb III, Speedy, Nebby Plus and the Myneb
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
## 1. General Information
| Submitter: | 3A Health Care S.r.l.<br>via Marziale Cerutti 90/F<br>25017 - Lonato (BS)<br>Italy<br>Simone Abate |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Maureen O'Connell<br>O'Connell Regulatory Consultants, Inc.<br>5 Timber Lane<br>North Reading, MA 01864<br>Telephone:<br>978-207-1245 |
| Consultant: | Guido Bonapace<br>ISENET<br>Via Calindri, 50<br>40068 - S.Lazzaro di Savena (BO)<br>Italy<br>Telephone:<br>+39-051-625 7315<br>Fax:<br>+39-051-8284344<br>Email: gbonapace@alice.it |
| Summary Preparation Date: | July 24, 2007 |
| 2. Names | Device Name: The families of Happyneb III<br>Happyneb III, Speedy, Nebby Plus and<br>Myneb |
| Classification Name: | Nebulizer<br>Regulation number: 868.5630<br>Product Code: CAF |
{1}------------------------------------------------
## 3. Predicate Devices
The 3A Health Care nebulizers Happyneb II, Happyneb III, Speedy, Nebby Plus and the Myneb, with their own nebulizers, Fasterjet and Nebjet, are substantially equivalent to a combination of the following devices:
- Healthdyne, Inc NEBULIZER SYSTEM K922623
- > Invacare - PRO, COMPACT and PORTABLE DESKTOP - K042483
- V Salter Labs - Salter Aire Compressor - K992285
- Medical Industries Sport-Neb (K964078)
for compressors, and
- Salter Labs 8900 nebulizer K870027 >
- > Medic-Aid Sidestream Nebulizer - K991725
- > Pari LC STAR nebulizer - K963924
for nebulizers.
### 4. Device Description
The family of 3A Health Care nebulizers include 5 different devices: 4 are AC powered devices (Happyneb II, Happyneb III, Speedy, Nebby Plus) and one model (Myneb) with a DC motor, a rechargeable battery pack and an external charger/power supplier.
The 4 AC compressors (Happyneb II, Happyneb III, Speedy, and Nebby Plus) have different plastic housings. There are three different types of electrical motors for AC models and one type for the DC model (Myneb). The 5 models of compressors are designed to use two nebulizers, the Fasterjet and Nebjet.
### 5. Indications for Use
The intended use of the Happyneb II, Happyneb III, Speedy, Nebby Plus and the Myneb is to spray liquids in aerosol form into gases that are delivered directly to adult or pediatric patients who have been prescribed inhalation therapy or medication. Each of these nebulizers must be used exclusively with their own nebulizer and mouthpiece.
The nebulizers Happyneb II, Happyneb III, Speedy, and the Nebby Plus are intended to be used primarily by patients in the home care market, although they may also be used by trained hospital staff personnel as well. The Myneb model is intended to be used only in home health care.
The Happyneb II, Happyneb III, Speedy, Nebby Plus and the Myneb are intended for multiple use, are non-sterile and for use with pharmaceutical products which are under physician prescription.
{2}------------------------------------------------
## 6. Performance Data
The non clinical performance bench tests performed by 3A Health Care Srl, have been executed as requested in "Reviewer guidance for nebulizers, metered dose inhalers, spacers and actuators" issued in October 1993, and demonstrate that 3A Health Care nebulizers have the same effectiveness as their predicate devices because they have equivalent performance parameters (MMAD, GSD, FPF and PFD). Therefore, clinical data are not required.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### MAR - 6 2008
3A Health Care S.r.l. C/O Ms. Maureen O'Connell Regulatory Consultant O'Connell Regulatory Consultants, Incorporated 5 Timber Lane North Reading, Massachusetts 01864
Re: K072163
Trade/Device Name: Happyneb II, Happyneb III, Speedy, Nebby Plus and the Myneb Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: January 23, 2008 Received: January 28, 2008
Dear Ms. O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Ms. O'Connell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Snyte Y. Michael Omd.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Happyneb II, Happyneb III, Speedy, Nebby Plus and the Myneb
Indications for Use:
The intended use of the Happyneb II. Happyneb III. Speedy, Nebby Plus and the Myneb is to spray liquids in aerosol form into gases that are delivered directly to adult or pediatric patients who have been prescribed inhalation therapy or medication. Each of these devices must be used exclusively with their own nebulizer and mouthpiece.
The devices Happyneb II, Happyneb III, Speedy, and the Nebby Plus are intended to be used primarily by patients in the home care market, although they may also be used by trained hospital staff personnel as well. The Myneb model is intended to be used only in home health care.
The Happyneb II, Happyneb III. Speedy, Nebby Plus and the Myneb are intended for multiple use, are non-sterile and for use with pharmaceutical products which are under physician prescription.
Prescription Use >< (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mui Muly
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.